Diurnal Intraocular Pressure Fluctuations with Self-tonometry in Glaucoma Patients and Suspects: A Clinical Trial

被引:32
|
作者
Huang, Jessie [1 ,2 ]
Katalinic, Paula [1 ,2 ]
Kalloniatis, Michael [1 ,2 ]
Hennessy, Michael P. [1 ,3 ]
Zangerl, Barbara [1 ,2 ]
机构
[1] Univ New South Wales Sydney, Ctr Eye Hlth, Sydney, NSW, Australia
[2] Univ New South Wales Sydney, Sch Optometry & Vis Sci, Sydney, NSW, Australia
[3] Prince Wales Hosp, Dept Ophthalmol, Sydney, NSW, Australia
基金
英国医学研究理事会;
关键词
OCULAR HYPERTENSION; REBOUND TONOMETER; RISK-FACTOR; ICARE-ONE; PROGRESSION; 24-HOUR; ACCURACY; TERM; PATTERNS;
D O I
10.1097/OPX.0000000000001172
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
SIGNIFICANCE: This article shows that self-tonometry can provide robust measures of diurnal intraocular pressure (IOP) and also detect changes to IOP in response to treatment within a short period of monitoring. These advances in IOP monitoring may contribute to improved management of glaucoma patients and suspects. PURPOSE: The aim of this study was to prospectively investigate the utility of rebound self-tonometry performed over several weeks for detecting diurnal IOP fluctuations in glaucoma patients and suspects and also initial response to topical treatment in glaucoma patients. METHODS: Forty patients were recruited following glaucoma-specific examination. Subsequent to successful training with the iCare HOME tonometer, patients were instructed to measure IOP, in a sitting position, four times a day over 4 to 6 weeks. Date, time, laterality, and IOP downloaded from the tonometer and clinical examination data, including applanation IOP and corneal thickness, were analyzed. A user satisfaction survey was also administered at study completion. t Test and analysis of variance were used to compare groups and IOP across days. Pearson correlation was used to compare measurements to Goldmann applanation tonometry and IOP measurements from the first day/s to the overall mean IOP. RESULTS: Twenty-seven patients (18 suspects and 9 glaucoma patients) completed data collection. Patients self-measured IOP on 118 (29) occasions for 40 (+/- 7.4) days. Two dominant patterns of fluctuation were revealed: peak IOP upon awakening (n = 11) and at midday (n = 13). Diurnal IOP measured in the first 7 days showed strong correlation to diurnal IOP across the entire study period (r(2) = 0.82, P < .0001). Within 24 hours of treatment commencement (latanoprost 0.005% ophthalmic solution), IOP reduced from 23.9 (+/- 5.2) to 16.1 (+/- 2.6) mmHg. Overall, patients rated the instrument as easy to use, although difficulties with correct alignment were expressed. CONCLUSIONS: Rebound self-tonometry demonstrated utility for measuring diurnal IOP fluctuations in most patients, hence enhancing management of patient with or at risk of developing glaucoma.
引用
收藏
页码:88 / 95
页数:8
相关论文
共 50 条
  • [21] Comparison of intraocular pressure peak and fluctuations among Filipino patients with non-glaucomatous eyes and glaucoma suspects using water drinking test and diurnal intraocular pressure
    Koh, Anelisa
    Verzosa, Cynthia
    INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2021, 14 (11) : 1729 - 1734
  • [22] Large diurnal fluctuations in intraocular pressure are an independent risk factor in patients with glaucoma
    Asrani, S
    Zeimer, R
    Wilensky, J
    Gieser, D
    Vitale, S
    Lindenmuth, K
    JOURNAL OF GLAUCOMA, 2000, 9 (02) : 134 - 142
  • [23] Comparison of intraocular pressure peak and fluctuations among Filipino patients with non-glaucomatous eyes and glaucoma suspects using water drinking test and diurnal intraocular pressure
    Anelisa Koh
    Cynthia Verzosa
    International Journal of Ophthalmology, 2021, (11) : 1729 - 1734
  • [24] Postural and diurnal fluctuations in intraocular pressure across the spectrum of glaucoma
    Gautam, Natasha
    Kaur, Savleen
    Kaushik, Sushmita
    Raj, Srishti
    Pandav, Surinder Singh
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2016, 100 (04) : 537 - 541
  • [25] Relevance of Intraocular Pressure and Tonometry in Glaucoma
    Erb, Carl
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 2011, 228 (02) : 97 - 97
  • [26] Self-tonometry in Glaucoma Management-Past, Present and Future
    Liang, Steve Y. -W.
    Lee, Graham A.
    Shields, David
    SURVEY OF OPHTHALMOLOGY, 2009, 54 (04) : 450 - 462
  • [27] Intraocular pressure fluctuations in glaucoma
    Orzalesi, Nicola
    Fogagnolo, Paolo
    Rossetti, Luca
    ARCHIVES OF OPHTHALMOLOGY, 2008, 126 (05) : 745 - 745
  • [28] Self-tonometry is useful to detect IOP elevations in DALK patients with apparently normal intraocular pressures
    Smedowski, A.
    Tarnawska, D.
    Wylegala, E.
    ACTA OPHTHALMOLOGICA, 2014, 92
  • [29] Comparison of self-measured diurnal intraocular pressure profiles using rebound tonometry between primary angle closure glaucoma and primary open angle glaucoma patients
    Tan, Shaoying
    Baig, Nafees
    Hansapinyo, Linda
    Jhanji, Vishal
    Wei, Shihui
    Tham, Clement C.
    PLOS ONE, 2017, 12 (03):
  • [30] Usability of self-tonometry using "Ocuton S®" for diurnal IOP profiles at home
    Vogt, R
    Duncker, GIW
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 2005, 222 (10) : 814 - 821