The effect of arterial stiffness on cuff-based blood pressure measurement

被引:4
|
作者
Ma, Yuxiang [1 ]
Chen, Ying [2 ,3 ]
Ma, Yinji [1 ]
Feng, Xue [1 ]
机构
[1] Tsinghua Univ, Ctr Flexible Elect Technol, Dept Engn Mech, Beijing 100084, Peoples R China
[2] Inst Flexible Elect Technol THU, Jiaxing 314000, Peoples R China
[3] Qiantang Sci & Technol Innovat Ctr, Hangzhou 310016, Peoples R China
基金
中国国家自然科学基金;
关键词
Cuff-based blood pressure measurement; Arterial collapse; Arterial stiffness; Arteriosclerosis; KOROTKOFF SOUNDS; BRACHIAL-ARTERY; PSEUDOHYPERTENSION;
D O I
10.1016/j.eml.2021.101298
中图分类号
TH [机械、仪表工业];
学科分类号
0802 ;
摘要
Cuff-based blood pressure measurement is the gold standard method for blood pressure monitoring. However, the wall of the sclerotic arteries (stiffen or thicken) generates extra resistance to the cuff pressure in the cuff-based blood pressure measurement, which tends to overestimate the blood pressure. This paper uses a collapse model to study the arterial deformation under the cuff pressure in the cuff-based blood pressure measurement, which gives the effects of arterial stiffness on measured systolic and diastolic blood pressure by cuff-based method, respectively. The results have general utility for future work in improving the accuracy of cuff-based BP measurement. (C) 2021 Published by Elsevier Ltd.
引用
收藏
页数:7
相关论文
共 50 条
  • [21] Pilot study of comparability of a smartphone blood pressure monitoring algorithm to conventional cuff-based blood pressure measurements
    Vischer, A.
    Rosania, J.
    Socrates, T.
    Blaschke, C.
    Proenca, M.
    Bonnier, G.
    Proust, Y. -M.
    Ferrario, D.
    Eckstein, J.
    Lemay, M.
    Burkard, T.
    EUROPEAN HEART JOURNAL, 2021, 42 : 2350 - 2350
  • [22] CUFF-BASED AMBULATORY BLOOD PRESSURE MEASUREMENT CAUSES AROUSAL REACTIONS AND RAISES SYSTOLIC BLOOD PRESSURE AT NIGHT - COMPARISON WITH BEAT-TO-BEAT AND CUFF-LESS METHOD
    Kuechler, G.
    Gehring, J.
    Fiedel, C.
    Schmitt, D.
    Voelker, W.
    Gohler, A.
    JOURNAL OF HYPERTENSION, 2017, 35 : E81 - E82
  • [23] Evaluation of the Vicorder, a novel cuff-based device for the noninvasive estimation of central blood pressure
    Pucci, Giacomo
    Cheriyan, Joseph
    Hubsch, Annette
    Hickson, Stacey S.
    Gajendragadkar, Parag R.
    Watson, Timothy
    O'Sullivan, Michael
    Woodcock-Smith, Jean
    Schillaci, Giuseppe
    Wilkinson, Ian B.
    McEniery, Carmel M.
    JOURNAL OF HYPERTENSION, 2013, 31 (01) : 77 - 85
  • [24] Effect of timing of blood pressure measurement in the assessment of arterial stiffness: The SMART study
    Simons, PCG
    Bots, ML
    Algra, A
    Stijn, A
    Van Teeffelen, P
    Van der Graaf, Y
    ULTRASOUND IN MEDICINE AND BIOLOGY, 1998, 24 (09): : 1285 - 1289
  • [25] Effect of timing of blood pressure measurement in the assessment of arterial stiffness: The SMART study
    Julius Center for Patient Oriented Research, University Hospital Utrecht, Utrecht, Netherlands
    不详
    Ultrasound Med. Biol., 9 (1285-1289):
  • [26] Comparison of invasive and brachial cuff-based noninvasive measurements for the assessment of blood pressure amplification
    Atsushi Nakagomi
    Sho Okada
    Toshihiro Shoji
    Yoshio Kobayashi
    Hypertension Research, 2017, 40 : 237 - 242
  • [27] Comparison of invasive and brachial cuff-based noninvasive measurements for the assessment of blood pressure amplification
    Nakagomi, Atsushi
    Okada, Sho
    Shoji, Toshihiro
    Kobayashi, Yoshio
    HYPERTENSION RESEARCH, 2017, 40 (03) : 237 - 242
  • [28] NOVEL METHOD FOR ASSESSING ARTERIAL STIFFNESS BASED ON OSCILLOMETRIC BLOOD PRESSURE MEASUREMENT
    Zhang, Chi
    Gao, Yongjia
    Long, Taochen
    Du, Yi
    Hou, Xiaoxu
    Li, Deyu
    JOURNAL OF MECHANICS IN MEDICINE AND BIOLOGY, 2018, 18 (07)
  • [29] Comparing blood pressure measurements between a photoplethysmography-based and a standard cuff-based manometry device
    Nachman, Dean
    Gepner, Yftach
    Goldstein, Nir
    Kabakov, Eli
    Ben Ishay, Arik
    Littman, Romi
    Azmon, Yuval
    Jaffe, Eli
    Eisenkraft, Arik
    SCIENTIFIC REPORTS, 2020, 10 (01)
  • [30] VALIDATION ASSESSMENT OF A CUFF-BASED CENTRAL BLOOD PRESSURE DEVICE ACCORDING TO ARTERY SOCIETY RECOMMENDATIONS
    Bui, Tan
    Picone, Dean
    Kosmala, Wojciech
    Kosowski, Michal
    Zachwyc, Jowita
    Pucci, Giacomo
    Schultz, Martin
    Black, Andrew
    Dwyer, Nathan
    Roberts-Thomson, Phillip
    Adams, Heath
    Curcio, Rosa
    Arrivi, Alessio
    Paleri, Sarang
    Sharman, James
    JOURNAL OF HYPERTENSION, 2024, 42 (SUPPL 3)