The effects of set-up errors on dose distribution in radiotherapy treatment for lung cancer

被引:1
|
作者
Cao, X. [1 ]
Liu, M. [1 ]
Zhai, F. [1 ]
Li, N. [1 ]
Bao, C. [1 ]
Liu, Y. [1 ]
Chen, G. [2 ]
机构
[1] Hebei Med Univ, Hosp 3, Dept Radiotherapy & Oncol, Shijiazhuang 050017, Hebei, Peoples R China
[2] Hebei Med Univ, Hosp 3, Dept Resp Med, Shijiazhuang 050017, Hebei, Peoples R China
来源
关键词
Tomography; X-ray computer; cone-beam; set-up error; lung tumor/radiation therapy; dosimetry; GUIDED RADIATION-THERAPY; IMAGE;
D O I
10.18869/acadpub.ijrr.19.3.515
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Our aim was to analyze the effects of set-up errors on dose distribution in radiotherapy treatment for lung cancer by using kilovoltage cone-beam CT (CBCT). Materials and Methods: In this study, we used a Varian IX linear accelerator system to perform CBCT scans of 30 lung cancer patients before radiotherapy. Subsequently, the image was matched with the planned CT, and the left and right (LR), top and bottom (SI), and front and back (AP) directions were set incorrectly. And in the CMS planning system, the center of the plan has been moved to the center of the actual scan. Finally, the dose distribution before the bed-moving is simulated. We want to explore the impact of the planned target volume setting error (PTV), the total tumor volume (GTV), and radiation of normal tissues. Results: The set-up errors of the LR, SI and AP directions were (-0.20 +/- 2.84), (-1.09 +/- 5.40), and (-2.61 +/- 2.08) mm, respectively. The 5mm error accounted for 97.8%, 73% and 92.6% in the three directions. Statistically significant differences were found in the distribution of 95% PTV dose, the average dose of PTV, 95% GTV dose and the average dose of GTV without bed-moving, compared with the original plan. Conclusions: In clinical lung cancer radiotherapy, the commonly used setting error is usually less than 5mm, most of which are along the AP direction. In this study, we found that the setting error is related to the patient's inherent characteristics and can significantly change the radiation treatment dose in the target area.
引用
收藏
页码:515 / 520
页数:6
相关论文
共 50 条
  • [31] ANALYSIS AND MANAGEMENT OF SET-UP UNCERTAINTIES IN RECTAL CANCER RADIOTHERAPY
    Jurado Bruggeman, D.
    Munoz, C.
    RADIOTHERAPY AND ONCOLOGY, 2009, 92 : S178 - S179
  • [32] Set-up error for conformal radiotherapy of prostate cancer - Reply
    Nutting, CM
    RADIOTHERAPY AND ONCOLOGY, 2000, 56 (03) : 355 - 356
  • [33] Evaluation of set-up verification with the analysis of systematic and random errors in radiotherapy - a study of the Great Poland Cancer Centre
    Szczurek, Lukasz
    Skrobala, Agnieszka
    Piotrowski, Tomasz
    Jodda, Agata
    NUKLEONIKA, 2008, 53 (04) : 167 - 171
  • [34] A treatment planning investigation into the dosimetric effects of systematic prostate patient rotational set-up errors
    Cranmer-Sargison, Gavin
    MEDICAL DOSIMETRY, 2008, 33 (03) : 199 - 205
  • [35] The potential impact of the tension of the pelvic muscles on set-up errors in radiotherapy for pelvic malignancies
    Bujko, K
    Czuchraniuk, P
    Zólciak, A
    Kukolowicz, P
    Kepka, L
    Bielik, A
    ACTA ONCOLOGICA, 2004, 43 (08) : 740 - 743
  • [36] Evaluation of set-up errors in head and neck radiotherapy using electronic portal imaging
    Strbac, Bojan
    Jokic, Vesna Spasic
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2013, 29 (05): : 531 - 536
  • [37] The importance of communication skills required to reduce set-up errors in the radiotherapy planning phase
    Merlotti, A.
    Reali, A.
    Russi, E.
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2019, 18 (04) : 395 - 396
  • [38] Organ motion, set-up variation and treatment margins in radical radiotherapy of urinary bladder cancer
    Muren, LP
    Smaaland, R
    Dahl, O
    RADIOTHERAPY AND ONCOLOGY, 2003, 69 (03) : 291 - 304
  • [39] Portal imaging to assess set-up errors, tumor motion and tumor shrinkage during conformal radiotherapy of non-small cell lung cancer
    Erridge, SC
    Seppenwoolde, Y
    Muller, SH
    van Herk, M
    De Jaeger, K
    Belderbos, JSA
    Boersma, LJ
    Lebesque, JV
    RADIOTHERAPY AND ONCOLOGY, 2003, 66 (01) : 75 - 85
  • [40] Quantification and assessment of Day-To-Day interfraction set-up errors in radiotherapy treatment delivery based on EPID detector
    Klimas, Aleksandra
    Janik, Michal
    Balamut, Karolina
    Slosarek, Krzysztof
    Cholewka, Armand
    POLISH JOURNAL OF MEDICAL PHYSICS AND ENGINEERING, 2025, 31 (01): : 1 - 9