The effects of set-up errors on dose distribution in radiotherapy treatment for lung cancer

被引:1
|
作者
Cao, X. [1 ]
Liu, M. [1 ]
Zhai, F. [1 ]
Li, N. [1 ]
Bao, C. [1 ]
Liu, Y. [1 ]
Chen, G. [2 ]
机构
[1] Hebei Med Univ, Hosp 3, Dept Radiotherapy & Oncol, Shijiazhuang 050017, Hebei, Peoples R China
[2] Hebei Med Univ, Hosp 3, Dept Resp Med, Shijiazhuang 050017, Hebei, Peoples R China
来源
关键词
Tomography; X-ray computer; cone-beam; set-up error; lung tumor/radiation therapy; dosimetry; GUIDED RADIATION-THERAPY; IMAGE;
D O I
10.18869/acadpub.ijrr.19.3.515
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background: Our aim was to analyze the effects of set-up errors on dose distribution in radiotherapy treatment for lung cancer by using kilovoltage cone-beam CT (CBCT). Materials and Methods: In this study, we used a Varian IX linear accelerator system to perform CBCT scans of 30 lung cancer patients before radiotherapy. Subsequently, the image was matched with the planned CT, and the left and right (LR), top and bottom (SI), and front and back (AP) directions were set incorrectly. And in the CMS planning system, the center of the plan has been moved to the center of the actual scan. Finally, the dose distribution before the bed-moving is simulated. We want to explore the impact of the planned target volume setting error (PTV), the total tumor volume (GTV), and radiation of normal tissues. Results: The set-up errors of the LR, SI and AP directions were (-0.20 +/- 2.84), (-1.09 +/- 5.40), and (-2.61 +/- 2.08) mm, respectively. The 5mm error accounted for 97.8%, 73% and 92.6% in the three directions. Statistically significant differences were found in the distribution of 95% PTV dose, the average dose of PTV, 95% GTV dose and the average dose of GTV without bed-moving, compared with the original plan. Conclusions: In clinical lung cancer radiotherapy, the commonly used setting error is usually less than 5mm, most of which are along the AP direction. In this study, we found that the setting error is related to the patient's inherent characteristics and can significantly change the radiation treatment dose in the target area.
引用
收藏
页码:515 / 520
页数:6
相关论文
共 50 条
  • [1] INFLUENCE OF DAILY SET-UP ERRORS ON DOSE DISTRIBUTION DURING PELVIS RADIOTHERAPY
    Kasabasic, Mladen
    Rajevac, Vedran
    Jurkovic, Slaven
    Ivkovic, Ana
    Sobat, Hrvoje
    Faj, Dario
    ARHIV ZA HIGIJENU RADA I TOKSIKOLOGIJU-ARCHIVES OF INDUSTRIAL HYGIENE AND TOXICOLOGY, 2011, 62 (03): : 261 - 267
  • [2] Construction of Prediction Model of Radiotherapy Set-Up Errors in Patients with Lung Cancer
    Yang, Fan
    Li, Xinxia
    BIOMED RESEARCH INTERNATIONAL, 2022, 2022
  • [3] Construction of Prediction Model of Radiotherapy Set-Up Errors in Patients with Lung Cancer
    Yang, Fan
    Li, Xinxia
    BIOMED RESEARCH INTERNATIONAL, 2022, 2022
  • [4] Evaluation of the trend of set-up errors during the treatment period using set-up margin in prostate radiotherapy
    Sasaki, Hinako
    Morishita, Takumi
    Irie, Naho
    Kojima, Rena
    Kiriyama, Tetsukazu
    Nakamoto, Akira
    Nishioka, Kunio
    Takahashi, Shotaro
    Tanabe, Yoshinori
    MEDICAL DOSIMETRY, 2024, 49 (04) : 291 - 297
  • [5] The effect of breathing and set-up errors on the cumulative dose to a lung tumor
    Engelsmann, M
    Damen, EMF
    De Jaeger, K
    van Ingen, KM
    Mijnheer, BJ
    RADIOTHERAPY AND ONCOLOGY, 2001, 60 (01) : 95 - 105
  • [6] Adaptive set-up correction (ASC) to account for set-up errors in prostate radiotherapy
    Lauve, A
    Siebers, J
    Crimaldi, A
    Hagan, M
    Keall, P
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (01): : S332 - S332
  • [7] RETRACTED: Construction of Prediction Model of Radiotherapy Set-Up Errors in Patients with Lung Cancer (Retracted Article)
    Yang, Fan
    Li, Xinxia
    BIOMED RESEARCH INTERNATIONAL, 2022, 2022
  • [8] Optimisation of treatment set-up during high dose radiotherapy of the prostate
    Syndikus, I.
    Heaton, A.
    Mayles, W.
    Sage, J.
    RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S195 - S196
  • [9] INFLUENCE OF COMPUTER-SIMULATED SET-UP ERRORS ON DOSE DISTRIBUTIONS IN EXTERNAL RADIOTHERAPY
    ROSENOW, U
    PHYSICS IN MEDICINE AND BIOLOGY, 1977, 22 (01): : 135 - 135
  • [10] An Evaluation of Translational Set-Up Errors to Define An Institutional Protocol for Set-Up Margins in the Treatment of Cancer Cervix Using Image Guided Radiotherapy (IGRT)
    Rani, G.
    Kaur, H.
    Ninan, J.
    Chandrasenan, P.
    Abdussameeh, M. P.
    MEDICAL PHYSICS, 2021, 48 (06)