A Study on Performance Evaluation of Magnetic Dental Attachments

被引:3
|
作者
Kang, Seen-Young [1 ]
Yu, Ji-Min [1 ]
Kim, Hyoung-Sik [1 ]
Park, Ki-Sook [1 ]
Lee, Seung-Youl [1 ]
机构
[1] Natl Inst Food & Drug Safety Evaluat, Med Device Res Div, 187 Osongsaengmyeong 2 Ro, Cheongju 28159, Chungcheongbuk, South Korea
关键词
magnetic dental attachment; ISO; 13017; retentive force; clinical environment; RETENTIVE FORCE; IN-VITRO; STUD ATTACHMENTS; IMPLANT; OVERDENTURES;
D O I
10.4283/JMAG.2019.24.4.733
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
This study measures retentive force as function of the diameter, the cross-head speed, and the number of detachments of dental magnetic attachments in a clinical environment, and analyzes the validity of the international standard method for testing them. In this study, tests 1, 2, and 3 were used to measure the retentive force as a function of the contact area of the magnetic attachment, and tests 2 and 4 as a function of the cross-head speed. Test 2 and 5 compared function of the retentive force as a function of repeated detachments Results showed that the retentive force increases as the sample surface increases, and decreases as the cross head speed increases. Additionally, after 1500 detachment cycles, the retentive force increased. Finally, the international standard test method was validated, because an objective method for testing magnetic attachments in clinical environment could not be found.
引用
收藏
页码:733 / 738
页数:6
相关论文
共 50 条
  • [31] Study on Performance Evaluation System for Transcranial Magnetic Stimulation Devices
    Li, Jiangtao
    Liang, Zheng
    Li, Jianhao
    Zhao, Zhijie
    Shi, Qindong
    2013 35TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2013, : 3582 - 3585
  • [32] Automated placement of dental attachments based on orthodontic pathways
    Lv, Yiheng
    Wei, Guangshun
    Fan, Yeying
    Ma, Long
    Liu, Dongxu
    Zhou, Yuanfeng
    COMPUTER AIDED GEOMETRIC DESIGN, 2024, 111
  • [33] MAGNETIC MACHINE-TOOL ATTACHMENTS
    KONSTANTINOV, OY
    RUSSIAN ENGINEERING JOURNAL, 1975, 55 (07): : 57 - 61
  • [34] THEORY OF MAGNETIC ATTACHMENTS HAVING A MAGNETIC-CIRCUIT
    ZAKHAROV, VA
    SOVIET JOURNAL OF NONDESTRUCTIVE TESTING-USSR, 1978, 14 (03): : 250 - 255
  • [35] Evaluation of polishing systems performance to dental porcelain
    Avieira, R
    Ribeiro, FC
    JOURNAL OF DENTAL RESEARCH, 2000, 79 (05) : 1114 - 1114
  • [36] Correlation of the performance evaluation in two dental course
    Navarro, A
    Garbero, I
    Erbiti, S
    Castillo, V
    Lencina, V
    Delgado
    JOURNAL OF DENTAL RESEARCH, 2000, 79 (05) : 1024 - 1024
  • [37] Evaluation of leakage flux out of a dental magnetic attachment
    Nishida, M.
    Tegawa, Y.
    Kinouchi, Y.
    2007 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-16, 2007, : 3520 - +
  • [38] EVALUATION OF MAGNETIC FILTER PERFORMANCE
    PARKER, MR
    IEEE TRANSACTIONS ON MAGNETICS, 1982, 18 (03) : 822 - 826
  • [39] Corrosion of Dental Magnet Attachments for Removable Prostheses on Teeth and Implants
    Boeckler, Arne F.
    Ehring, Carolin
    Morton, Dean
    Geis-Gerstorfer, Juergen
    Setz, Juergen M.
    JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2009, 18 (04): : 301 - 308
  • [40] Outcomes of resin-bonded attachments for removable dental prostheses
    Garling, Anne
    Krummel, Anna
    Kern, Matthias
    JOURNAL OF PROSTHODONTIC RESEARCH, 2024, 68 (01) : 100 - 104