摘要
Background Increasing numbers of patients with active implantable medical devices (AIMDs) require magnetic resonance (MR) examinations. The manufacturers are continuing to improve the MR compatibility of their AIMDs. To this end, a variety of measurement methods and numerical simulations are used to evaluate the risks associated with magnetic resonance imaging (MRI). Objective In this article, test methods used to investigate interactions between AIMDs with radio frequency fields and time-varying magnetic gradient fields are reviewed. Materials and methods A literature review of known test methods for radio frequency and gradient field exposure of AIMDs with leads, in particular for neurostimulators, cochlear implants, and implanted infusion pumps, is presented. The state of the art and promising methods are discussed. Results ISO/TS 10974 describes the design of high frequency and gradient injection setups to test conductive materials. A large number of sensor designs have been published to measure the induced voltages and currents through radio frequency and gradient fields and for monitoring AIMDs during MR examinations in in vitro tests. Conclusion The test methods should be planned to be as conservative as possible to cover the worst case scenario. However, in vitro measurements and computer simulation are far from being able to cover all possible configurations in their complexity and uniqueness. For safer MR examinations, current research recommends in vivo testing prior to MR, parallel radiofrequency transmission techniques, and new sequences with reduced energy input in the presence of AIMDs.
引用
收藏
页码:40 / 45
页数:6
相关论文
共 50 条
  • [31] Implantable medical devices
    Crickx, B.
    Arrault, X.
    ANNALES DE DERMATOLOGIE ET DE VENEREOLOGIE, 2008, 135 (01): : S66 - S69
  • [32] Novel test field diversity method for demonstrating magnetic resonance imaging safety of active implantable medical devices
    Yao, Aiping
    Zastrow, Earl
    Neufeld, Esra
    Cabanes-Sempere, Maria
    Samaras, Theodoros
    Kuster, Niels
    PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (07):
  • [33] Safety of active implantable devices during MRI examinations:: A finite element analysis of an implantable pump
    Buechler, Philippe
    Simon, Anne
    Burger, Juergen
    Ginggen, Alec
    Crivelli, Rocco
    Tardy, Yanik
    Luechinger, Roger
    Olsen, Sigbjorn
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2007, 54 (04) : 726 - 733
  • [34] Active implantable medical devices. Medico-legal aspects
    Bartsch, C.
    Irnich, W.
    RECHTSMEDIZIN, 2011, 21 (05) : 493 - 502
  • [35] MR-Sicherheitsbewertung von aktiven implantierbaren medizinischen GerätenMR safety assessment of active implanted medical devices. German version
    Sarra Aissani
    Elmar Laistler
    Jacques Felblinger
    Der Radiologe, 2019, 59 (10): : 869 - 874
  • [36] Better assessment and safety improvement for medical devices
    Belghiti, J.
    BULLETIN DE L ACADEMIE NATIONALE DE MEDECINE, 2019, 203 (05): : 255 - 256
  • [37] Distribution of MR Safety Conditions Across Active Implanted Devices
    Panda, A.
    Zhou, Y.
    Fahrenholtz, S.
    Yu, J.
    Rand, E.
    Hines, J.
    Sensakovic, W.
    MEDICAL PHYSICS, 2022, 49 (06) : E177 - E177
  • [38] An Invited Review for the Special 20th Anniversary Issue of MRMS New Insights into MR Safety for Implantable Medical Devices
    Kuroda, Kagayaki
    Yatsushiro, Satoshi
    MAGNETIC RESONANCE IN MEDICAL SCIENCES, 2022, 21 (01) : 110 - 131
  • [39] Implantable RF Medical Devices
    Chow, Eric Y.
    Morris, Milton M.
    Irazoqui, Pedro P.
    IEEE MICROWAVE MAGAZINE, 2013, 14 (04) : 64 - 73
  • [40] Testing implantable medical devices
    Cortner, JM
    EE-EVALUATION ENGINEERING, 2004, 43 (06): : 38 - +