Integrated megavoltage portal imaging with a 1.5 T MRI linac

被引:31
|
作者
Raaymakers, B. W. [1 ]
de Boer, J. C. J. [1 ]
Knox, C. [1 ]
Crijns, S. P. M. [1 ]
Smit, K. [1 ]
Stam, M. K. [1 ]
van den Bosch, M. R. [1 ]
Kok, J. G. M. [1 ]
Lagendijk, J. J. W. [1 ]
机构
[1] Univ Med Ctr Utrecht, Dept Radiotherapy, NL-3584 CX Utrecht, Netherlands
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2011年 / 56卷 / 19期
关键词
RADIOTHERAPY; DEVICE; VERIFICATION; DISTORTIONS; QUALITY; CANCER;
D O I
10.1088/0031-9155/56/19/N01
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this note, the feasibility of complementing our hybrid 1.5 T MRI linac (MRL) with a megavoltage (MV) portal imager is investigated. A standard aSi MV detector panel is added to the system and both qualitative and quantitative performances are determined. Simultaneous MR imaging and transmission imaging can be performed without mutual interference. The MV image quality is compromised by beam transmission and longer isocentre distance; still, the field edges and bony anatomy can be detected at very low dose levels of 0.4 cGy. MV imaging integrated with the MRL provides an independent and well-established position verification tool, a field edge check and a calibration for alignment of the coordinate systems of the MRI and the accelerator. The portal imager can also be a valuable means for benchmarking MRI-guided position verification protocols on a patient-specific basis in the introductory phase.
引用
收藏
页码:N207 / N214
页数:8
相关论文
共 50 条
  • [21] A PHOTODIODE-SCINTILLATION DETECTOR ARRAY FOR MEGAVOLTAGE PORTAL IMAGING
    MORTON, EJ
    SWINDELL, W
    BRITISH JOURNAL OF RADIOLOGY, 1988, 61 (726): : 561 - 561
  • [22] MRI commissioning of 1.5T MR-linac systems - a multi-institutional study
    Tijssen, Rob H. N.
    Philippens, Marielle E. P.
    Paulson, Eric S.
    Glitzner, Markus
    Chugh, Brige
    Wetscherek, Andreas
    Dubec, Michael
    Wang, Jihong
    van der Heide, Uulke A.
    RADIOTHERAPY AND ONCOLOGY, 2019, 132 : 114 - 120
  • [23] Comprehensive MRI Acceptance Testing & Commissioning of a 1.5T MR-Linac: Guidelines and Results
    Tijssen, R. H. N.
    Crijns, S. P. M.
    Bluemink, J. J.
    Hacket, S. S.
    DeVries, J. H. W.
    Kruiskamp, M. J.
    Philippens, M. E. P.
    Lagendijk, J. J. W.
    Raaymakers, B. W.
    RADIOTHERAPY AND ONCOLOGY, 2017, 123 : S130 - S131
  • [24] Repeatability of diffusion-weighted MRI in head and neck cancer at a 1.5 T MR-Linac
    Habrich, J.
    Boeke, S.
    Nachbar, M.
    Nikolaou, K.
    Schick, F.
    Gani, C.
    Zips, D.
    Thorwarth, D.
    RADIOTHERAPY AND ONCOLOGY, 2021, 161 : S147 - S148
  • [25] Time management and hands-on experience with ELEKTA Unity 1.5T MRI-Linac
    Marks, C.
    Stolte, A.
    Thorwarth, D.
    Braun, L. H.
    Boeke, S.
    Wegener, D.
    Boldt, J.
    Ortinau, C.
    Kammler, M.
    Holl-Henkel, B.
    Gani, C.
    Zips, D.
    Nachbar, M.
    Dohm, O.
    Moennich, D.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S608 - S608
  • [26] Dose escalation and hypofractionation for SBRT of lymph node oligometastases on the 1.5T MRI-Linac
    Winkel, D.
    Bol, G.
    Werensteijn-Honingh, A.
    Kiekebosch, I.
    Hes, J.
    Intven, M.
    Eppinga, W.
    Raaymakers, B.
    Jurgenliemk-Schulz, I.
    Kroon, P.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S443 - S444
  • [27] MRI-Guided Adaptive Head and Neck Radiotherapy On a 1.5T MR Linac System
    Aliotta, E.
    Hesse, J.
    Subashi, E.
    Scripes, P.
    Aristophanous, M.
    Mechalakos, J.
    Cervino, L.
    Tyagi, N.
    Zakeri, K.
    Lee, N.
    MEDICAL PHYSICS, 2022, 49 (06) : E761 - E761
  • [28] Optimizing Acquisition Speed and Contrast of Respiratory Correlated 4D-MRI on a 1.5T MRI-Linac
    Stemkens, B.
    Bruijnen, T.
    Van den Berg, C. A. T.
    Lagendijk, J. J. W.
    Tijssen, R. H. N.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S285 - S286
  • [29] Minimizing the need for coil attenuation correction in integrated PET/MRI at 1.5 T using low-density MR-linac receive arrays
    Zijlema, Stefan E.
    Branderhorst, Woutjan
    Bastiaannet, Remco
    Tijssen, Rob H. N.
    Lagendijk, Jan J. W.
    van den Berg, Cornelis A. T.
    PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (20):
  • [30] Imaging single mammalian cells with a 1.5 T clinical MRI scanner
    Foster-Gareau, P
    Heyn, C
    Alejski, A
    Rutt, BK
    MAGNETIC RESONANCE IN MEDICINE, 2003, 49 (05) : 968 - 971