DYNAMIC PIXEL BINNING ALLOWS SPATIAL AND ANGULAR RESOLUTION TRADEOFFS TO IMPROVE IMAGE QUALITY IN X-RAY C-ARM CT

被引:0
|
作者
Steg, Alexander [1 ]
Reichenbach, Marc [1 ]
Soell, Christopher [1 ]
Shi, Lan [1 ]
Maier, Andreas [1 ]
Riess, Christian [1 ]
机构
[1] Friedrich Alexander Univ Erlangen Nuremberg, Dept Comp Sci, Dept Elect Engn, Erlangen, Germany
来源
2016 IEEE 13TH INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING (ISBI) | 2016年
关键词
Pixel Binning; X-ray detector; C-arm reconstruction; Dynamic Pixel Binning; X-ray C-arm CT;
D O I
10.1109/ISBI.2016.7493334
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
When designing an acquisition protocol for a C-arm CT, a tradeoff has to be made between scan time and the amount of acquired data. Increased scan time allows data acquisition with higher spatial and angular resolution. Fixed scan time implies a reduction of either spatial and angular resolution, which may either lead to oversmoothing or to artifacts from angular undersampling. In this work, we investigate smart sensors for pushing the boundary for this bottleneck considerably further. Detector pixels are enhanced by processing units that are able to perform a limited amount of local operations. This allows an online, dynamical adaptation of the pixel binning, depending on whether a local neighborhood is smooth or not. We demonstrate on phantom data that this approach allows to save about 7 5% of the data, obtaining a reconstruction quality comparable to 1 x 1 binning at a data rate comparable to 2 x 2 pixel binning.
引用
收藏
页码:577 / 580
页数:4
相关论文
共 39 条
  • [31] Correlating spatial model of coal quality with coal microstructures detected from x-ray CT image analysis
    Heriawan, Mohamad Nur
    Koike, Katsuaki
    PROCEEDINGS OF THE IAMG '07: GEOMATHEMATICS AND GIS ANALYSIS OF RESOURCES, ENVIRONMENT AND HAZARDS, 2007, : 86 - +
  • [32] Study of 20 Hz high spatial-temporal resolution monochromatic X-ray dynamic micro-CT
    Qiu Z.
    Li K.
    Xie H.
    Du G.
    Deng B.
    Xiao T.
    He Jishu/Nuclear Techniques, 2023, 46 (07):
  • [33] An Analytical Approach for Estimating Angular Change of a C-Arm Fluoroscope Necessary to Provide Discrete X-Ray Fields at the Plane of Patient Skin in Fluoroscopically Guided Interventions
    Zhao, A.
    Wunderle, K.
    MEDICAL PHYSICS, 2017, 44 (06)
  • [34] Regularization Designs for Uniform Spatial Resolution and Noise Properties in Statistical Image Reconstruction for 3-D X-ray CT
    Cho, Jang Hwan
    Fessler, Jeffrey A.
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2015, 34 (02) : 678 - 689
  • [35] A portable test platform for image acquisition and calibration for cone beam computed tomography (CBCT) and region of interest CBCT (ROI-CBCT) on a commercial X-ray C-arm system
    Rangwala, H.
    Chityala, R.
    Rudin, S.
    Hoffmann, K.
    MEDICAL PHYSICS, 2006, 33 (06) : 1997 - 1998
  • [36] 3D cone-beam CT with a twin robotic x-ray system in elbow imaging: comparison of image quality to high-resolution multidetector CT
    Grunz, Jan-Peter
    Weng, Andreas Max
    Kunz, Andreas Steven
    Veyhl-Wichmann, Maike
    Schmitt, Rainer
    Gietzen, Carsten Herbert
    Pennig, Lenhard
    Herz, Stefan
    Erguen, Sueleyman
    Bley, Thorsten Alexander
    Gassenmaier, Tobias
    EUROPEAN RADIOLOGY EXPERIMENTAL, 2020, 4 (01)
  • [37] 3D cone-beam CT with a twin robotic x-ray system in elbow imaging: comparison of image quality to high-resolution multidetector CT
    Jan-Peter Grunz
    Andreas Max Weng
    Andreas Steven Kunz
    Maike Veyhl-Wichmann
    Rainer Schmitt
    Carsten Herbert Gietzen
    Lenhard Pennig
    Stefan Herz
    Süleyman Ergün
    Thorsten Alexander Bley
    Tobias Gassenmaier
    European Radiology Experimental, 4
  • [38] Addition of computer simulated noise to investigate radiation dose and image quality in images with spatial correlation of statistical noise: an example application to X-ray CT of the brain
    Britten, AJ
    Crotty, M
    Kiremidjian, H
    Grundy, A
    Adam, EJ
    BRITISH JOURNAL OF RADIOLOGY, 2004, 77 (916): : 323 - 328
  • [39] Anatomical Analysis of Turgescent and Semi-Dry Resurrection Plants: The Effect of Sample Preparation on the Sample, Resolution, and Image Quality of X-Ray Micro-Computed Tomography (μCT)
    Korte, Nikola
    Porembski, Stefan
    MICROSCOPY RESEARCH AND TECHNIQUE, 2011, 74 (04) : 364 - 369