MR-based truncation and attenuation correction in integrated PET/MR hybrid imaging using HUGE with continuous table motion

被引:32
|
作者
Lindemann, Maike E. [1 ]
Oehmigen, Mark [1 ]
Blumhagen, Jan O. [2 ]
Gratz, Marcel [1 ,3 ]
Quick, Harald H. [1 ,3 ]
机构
[1] Univ Duisburg Essen, Univ Hosp Essen, High Field & Hybrid MR Imaging, Essen, Germany
[2] Siemens Healthcare GmbH, Magnet Resonance, Erlangen, Germany
[3] Univ Duisburg Essen, Erwin L Hahn Inst Magnet Resonance Imaging, Essen, Germany
关键词
attenuation correction; continuous table motion; FOV extension; PET quantification; truncation correction; whole-body PET/MR; WHOLE-BODY PET/MR; EMISSION TOMOGRAPHY/COMPUTED TOMOGRAPHY; QUANTITATIVE-EVALUATION; SYSTEMS; PHANTOM; CANCER;
D O I
10.1002/mp.12449
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The objective of this study was to introduce and evaluate a method for MR-based attenuation and truncation correction in phantom and patient measurements to improve PET quantification in PET/MR hybrid imaging. Methods: The fully MR-based method HUGE (B0 Homogenization using gradient enhancement) provides field-of-view extension in MR imaging, which can be used for truncation correction and improved PET quantification in PET/MR hybrid imaging. The HUGE method in this recent implementation is combined with continuously moving table data acquisition to provide a seamless nontruncated whole-body data set of the outer patient contours to complete the established standard MRbased Dixon-VIBE data for attenuation correction. The method was systematically evaluated in NEMA standard phantom experiments to investigate the impact of HUGE truncation correction on PET quantification. The method was then applied to 24 oncologic patients in whole-body PET/MR hybrid imaging. The impact of MR-based truncation correction with HUGE on PET data was compared to the impact of the established PET-based MLAA algorithm for contour detection. Results: In phantom and in all patient measurements, the standard Dixon-VIBE attenuation correction data show geometric distortions and signal truncations at the edges of the MR imaging field-ofview. In contrast, the Dixon-VIBE-based attenuation correction data additionally extended by applying HUGE shows significantly less distortion and truncations and due to the continuously moving table acquisition robustly provides smooth outer contours of the patient arms. In the investigated patient cases, MLAA frequently showed an overestimation of arm volume and associated artifacts limiting contour detection. When applying HUGE, an average relative increase in SUVmean in patients' lesion of 4.2% and for MLAA of 4.6% were measured, when compared to standard DixonVIBE only. In specific lesions maximal differences in SUVmean up to 13% for HUGE and 14% for MLAA were measured. Quantification in truncated regions showed maximal differences up to 40% for both, MLAA and HUGE. Average differences in those regions in SUVmean for HUGE are 13.3% and 14.6% for MLAA. In a patient with I-124 radiotracer PET-based MLAA contour detection completely failed in this specific case, whereas HUGE as MR-based approach provided accurate truncation correction. Conclusions: The HUGE method for truncation correction combined with continuous table movement extends the lateral MR field-of-view and effectively reduces truncations along the outer contours of the patient's arms in whole-body PET/MR imaging. HUGE as a fully MR-based approach is independent of the choice of radiotracer, thus also offering robust truncation correction in patients that are not injected with Fluordesoxyglucose (FDG) as radiotracer. Therefore, this method improves the standard Dixon MR-based attenuation correction and PET image quantification in whole-body PET/MR imaging applications. (C) 2017 American Association of Physicists in Medicine
引用
收藏
页码:4559 / 4572
页数:14
相关论文
共 50 条
  • [21] MR-based attenuation correction for torso-PET/MR imaging: pitfalls in mapping MR to CT data
    Thomas Beyer
    Markus Weigert
    Harald H. Quick
    Uwe Pietrzyk
    Florian Vogt
    Christoph Palm
    Gerald Antoch
    Stefan P. Müller
    Andreas Bockisch
    European Journal of Nuclear Medicine and Molecular Imaging, 2008, 35 : 1142 - 1146
  • [22] MR-based attenuation correction for torso-PET/MR imaging:: pitfalls in mapping MR to CT data
    Beyer, Thomas
    Weigert, Markus
    Quick, Harald H.
    Pietrzyk, Uwe
    Vogt, Florian
    Palm, Christoph
    Antoch, Gerald
    Mueller, Stefan P.
    Bockisch, Andreas
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2008, 35 (06) : 1142 - 1146
  • [23] Accurate hybrid template-based and MR-based attenuation correction using UTE images for simultaneous PET/MR brain imaging applications
    Baran, Jakub
    Chen, Zhaolin
    Sforazzini, Francesco
    Ferris, Nicholas
    Jamadar, Sharna
    Schmitt, Ben
    Faul, David
    Shah, Nadim Jon
    Cholewa, Marian
    Egan, Gary F.
    BMC MEDICAL IMAGING, 2018, 18
  • [24] The effect of positioning aids on PET quantification following MR-based attenuation correction (AC) in PET/MR imaging
    Mantlik, Frederic
    Hofmann, Matthias
    Kupferschlaeger, Juergen
    Werner, Matthias
    Pichler, Bernd
    Beyer, Thomas
    JOURNAL OF NUCLEAR MEDICINE, 2010, 51
  • [25] MR-based attenuation correction in cardiovascular PET/MR imaging: challenges and practical solutions for cardiorespiratory motion and tissue class segmentation
    Karakatsanis, Nicolas
    Robson, Philip
    Dweck, Marc
    Abgral, Ronan
    Trivieri, Maria
    Sanz, Javier
    Contreras, Johanna
    Narula, Jagat
    Padilla, Maria
    Gidwani, Umesh
    Fuster, Valentin
    Kovacic, Jason
    Fayad, Zahi
    JOURNAL OF NUCLEAR MEDICINE, 2016, 57
  • [26] MR-based Attenuation Correction for Whole-body PET/MR System
    Hu, Z.
    Renisch, S.
    Schweizer, B.
    Blaffert, T.
    Ojha, N.
    Guo, T.
    Tang, J.
    Tung, C.
    Kaste, J.
    Schulz, V.
    Torres, I.
    Shao, L.
    2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 2119 - 2122
  • [27] MR-based attenuation correction for whole-body PET-MR
    Ouyang, Jinsong
    Chun, Se Young
    Brady, Thomas
    El Fakhri, Georges
    JOURNAL OF NUCLEAR MEDICINE, 2010, 51
  • [28] MR-based attenuation correction for whole-body PET-MR
    Ouyang, Jinsong
    Chun, Se Young
    El Fakhri, Georges
    JOURNAL OF NUCLEAR MEDICINE, 2011, 52
  • [29] Quantitative PET/MR imaging of lung cancer in the presence of artifacts in the MR-based attenuation correction maps
    Kuttner, Samuel
    Lassen, Martin Lyngby
    Oen, Silje Kjaernes
    Sundset, Rune
    Beyer, Thomas
    Eikenes, Live
    ACTA RADIOLOGICA, 2020, 61 (01) : 11 - 20
  • [30] MR-Based Motion Correction in Coronary 18F-Fluoride PET Imaging Using Simultaneous PET-MR
    Petibon, Y.
    Wilks, M. Q.
    Ma, C.
    Huang, C.
    Lu, M. T.
    Zanni, M. K.
    Grinspoon, S. K.
    Ouyang, J.
    El Fakhri, G.
    MEDICAL PHYSICS, 2017, 44 (06) : 3139 - 3139