Determination of 99Tc in fresh water using TRU resin by ICP-MS

被引:16
|
作者
Guerin, Nicolas [1 ]
Riopel, Remi [1 ]
Kramer-Tremblay, Sheila [1 ]
de Silva, Nimal [1 ]
Cornett, Jack [1 ]
Dai, Xiongxin [1 ]
机构
[1] CNL, Radiol Protect Res & Instrumentat Branch, Chalk River, ON K0J 1J0, Canada
关键词
Technetium-99; Rhenium; Inductively coupled plasma mass spectrometry; TRU resin; Anion exchange resin; Environmental water samples; ENVIRONMENTAL-SAMPLES; VAPOR-PRESSURES; TECHNETIUM; RHENIUM; SEPARATION; EXTRACTION; HEPTOXIDE; LOBSTERS; TC;
D O I
10.1016/j.aca.2017.08.013
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Technetium-99 (Tc-99) determination at trace level by inductively coupled plasma mass spectrometry (ICP-MS) is challenging because there is no readily available appropriate Tc isotopic tracer. A new method using Re as a recovery tracer to determine Tc-99 in fresh water samples, which does not require any evaporation step, was developed. Tc(VII) and Re(VII) were pre-concentrated on a small anion exchange resin (AER) cartridge from one litre of water sample. They were then efficiently eluted from the AER using a potassium permanganate (KMnO4) solution. After the reduction of KMnO4 in 2 M sulfuric acid solution, the sample was passed through a small TRU resin cartridge. Tc(VII) and Re(VII) retained on the TRU resin were eluted using near boiling water, which can be directly used for the ICP-MS measurement. The results for method optimisation, validation and application were reported. Crown Copyright (C) 2017 Published by Elsevier B.V. All rights reserved.
引用
收藏
页码:114 / 120
页数:7
相关论文
共 50 条
  • [41] Strategy for water analysis using ICP-MS
    Fernández-Turiel, JL
    Llorens, JF
    López-Vera, F
    Gómez-Artola, C
    Morell, I
    Gimeno, D
    FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 2000, 368 (06): : 601 - 606
  • [42] Overcoming ICP-QMS instrumental limitations for 99Tc determination in environmental solid samples using radiochemistry
    Mas, JL
    García-León, M
    Bolívar, JP
    APPLIED RADIATION AND ISOTOPES, 2006, 64 (04) : 502 - 507
  • [43] Strategy for water analysis using ICP-MS
    J. L. Fernández-Turiel
    J. F. Llorens
    F. López-Vera
    C. Gómez-Artola
    I. Morell
    D. Gimeno
    Fresenius' Journal of Analytical Chemistry, 2000, 368 : 601 - 606
  • [44] DETERMINATION OF 99Tc IN ENVIRONMENTAL SAMPLES.
    Univ of Lund, Dep of Radiation Physics
    Rioseco, J.
    Garcia-Leon, M.
    Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 1983, 223 (2-3) : 204 - 207
  • [45] Determination of mercury in potable water by ICP-MS using gold as a stabilising agent
    Allibone, J
    Fatemian, E
    Walker, PJ
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1999, 14 (02) : 235 - 239
  • [46] DETERMINATION OF LOW-LEVEL TC-99 IN ENVIRONMENTAL-SAMPLES BY HIGH-RESOLUTION ICP-MS
    YAMAMOTO, M
    SYARBAINI
    KOFUJI, K
    TSUMURA, A
    KOMURA, K
    UENO, K
    ASSINDER, DJ
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY-ARTICLES, 1995, 197 (01): : 185 - 194
  • [47] Comparison of ICP-MS and ICP-OES in the Determination of Trace Elements in Water
    Cidu, R.
    Atomic Spectroscopy, 17 (04):
  • [48] Determination of selenium in the commercially bottled drinking water by ICP-MS - Determination of trace elements in environmental samples by ICP-MS .4.
    Naohara, J
    Yamashita, E
    JOURNAL OF THE JAPANESE SOCIETY FOR FOOD SCIENCE AND TECHNOLOGY-NIPPON SHOKUHIN KAGAKU KOGAKU KAISHI, 1996, 43 (10): : 1133 - 1137
  • [50] A comparison of ICP-OES and ICP-MS in the determination of elements in lake water
    Baralkiewicz, Danuta
    Gramowska, Hanka
    Hanc, Anetta
    Krzyzaniak, Izabela
    ATOMIC SPECTROSCOPY, 2007, 28 (05) : 164 - 170