Determination of mercury in potable water by ICP-MS using gold as a stabilising agent

被引:129
|
作者
Allibone, J [1 ]
Fatemian, E [1 ]
Walker, PJ [1 ]
机构
[1] Thames Water Util Ltd, Millharbour Lab, Great Eastern Enterprise, London E14 9XP, England
关键词
D O I
10.1039/a806193i
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A method for the routine determination of mercury in potable water samples was developed that uses addition of gold to samples to preserve mercury by amalgamation and subsequent analysis by inductively coupled plasma mass spectrometry (ICP-MS). The method was validated using the guidelines of the NS 30 document and shows compliance with the analytical performance criteria required by the UK's Drinking Water Inspectorate over the range 0-1.2 mu g l(-1) Hg. The prescribed concentration or value (PCV) for mercury is 1.0 mu g l(-1). A limit of detection 0.032 mu g l(-1) and a recovery of 99.0% at the PCV were obtained. Gold was added off-line to both samples and standard solutions. It was also added to the ICP-MS wash water to eliminate any memory effects from mercury in previous samples. The details of gold addition and the operating conditions of the Perkin-Elmer SCIEX Elan 6000 ICP-MS instrument are described. This work forms part of a project to enable the routine determination of all the low-concentration metals in potable water samples to be achieved in a single run using ICP-MS.
引用
收藏
页码:235 / 239
页数:5
相关论文
共 50 条
  • [1] Determination of Mercury, Arsenic, Selenium, and Antimony in potable water by ICP-MS
    Allibone, J
    Fatemian, E
    Walker, PJ
    PLASMA SOURCE MASS SPECTROMETRY: THE NEW MILLENNIUM, 2001, (267): : 216 - 229
  • [2] Use of gold as a routine and long term preservative for mercury in potable water, as determined by ICP-MS
    Fatemian, E
    Allibone, J
    Walker, PJ
    ANALYST, 1999, 124 (08) : 1233 - 1236
  • [3] Comment on "Determination of mercury in potable water by ICP-MS using gold as a stabilising agent" (J.!Allibone, E.!Fatemian and P. J.!Walker, J-Anal. At. Spectrom., 1999, 14, 235)
    Sturman, BT
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2000, 15 (11) : 1512 - 1512
  • [4] Mercury determination with ICP-MS: signal suppression by acids
    L. Jian
    W. Goessler
    K. J. Irgolic
    Fresenius' Journal of Analytical Chemistry, 2000, 366 : 48 - 53
  • [5] The determination of mercury in microwave digests of foodstuffs by ICP-MS
    Entwisle, J
    AMERICAN LABORATORY, 2004, 36 (06) : 11 - +
  • [6] Mercury determination with ICP-MS: signal suppression by acids
    Jian, L
    Goessler, W
    Irgolic, KJ
    FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 2000, 366 (01): : 48 - 53
  • [7] DETERMINATION OF ULTRATRACE URANIUM AND THORIUM IN GOLD BY ICP-MS
    ITO, K
    KUBOYAMA, S
    OGURO, H
    KAGAYA, T
    SHIMAMURA, T
    BUNSEKI KAGAKU, 1988, 37 (01) : 25 - 29
  • [9] Determination of uranium in tap water by ICP-MS
    El Himri, M
    Pastor, A
    de la Guardia, M
    FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 2000, 367 (02): : 151 - 156
  • [10] Determination of uranium in tap water by ICP-MS
    El Himri M.
    Pastor A.
    De La Guardia M.
    Fresenius' Journal of Analytical Chemistry, 2000, 367 (2): : 151 - 156