摘要
An analytical procedure has been proposed for the determination of precise uranium isotope ratios in a thin uranium layer on a biological surface by laser ablation inductively coupled plasma sector field mass spectrometry (LA-ICP-MS). A cooled laser ablation chamber using a Peltier element was developed in order to analyze element distribution in thin cross-sections of frozen tissues with a lateral resolution in the mu m range. In order to study the figures of merit of LA-ICP-MS with the cooled laser ablation chamber, one drop (20 mu L U concentration 200 ng mL(-1)), each of the certified isotope reference materials NIST U350 and NIST U930, the uranium isotopic standard CCLU-500 and also a drop of uranium with a natural isotopic pattern was deposited and analyzed on the biological surface (flower leaf). The precision and accuracy of isotope ratio measurements are significantly improved using cooled laser ablation chamber in comparison to non-cooled chamber. The precision of the measurements of isotope ratios in the range of 2.0-1.6% for U-234/U-238, 1.3-0.4% for U-235/U-238 and 2.1-1.0% for U-236/U-238 in selected uranium isotopic standards reference material were determined by microlocal analysis (diameter of laser ablation crater: 15, 25 and 50 mu m) using LA-ICP-MS with a cooled laser ablation chamber. The accuracy of U-234/U-238, U-235/U-238 and U-236/U-238 isotope ratio measurements varied in the range of 4.2-1.1%, 2.4-0.5% and 4.8-1.1%, respectively, and were dependent on the diameter of the laser beam used. (c) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:297 / 302
页数:6
相关论文
共 50 条
  • [1] Determination of Uranium Isotope Ratios in Uranium Particles by Laser Ablation Multiple Collector Inductively Coupled Plasma Mass Spectrometry
    Wang Wei
    Li Zhi-Ming
    Xu Jiang
    Zhou Guo-Qing
    Shen Xiao-Pan
    Zhai Li-Hua
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2015, 43 (05) : 703 - 708
  • [2] Multielement analysis of petroleum samples by laser ablation double focusing sector field inductively coupled plasma mass spectrometry (LA-ICP MS)
    Vorapalawut, Nopparat
    Pohl, Pawel
    Bouyssiere, Brice
    Shiowatana, Juwadee
    Lobinski, Ryszard
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2011, 26 (03) : 618 - 622
  • [3] Determination of mercury in sediment samples by laser ablation inductively coupled plasma mass spectrometry
    Tao, GH
    Fujikawa, Y
    Mitsui, M
    Yamada, R
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2002, 17 (05) : 560 - 562
  • [4] Assessment of laser ablation and sector field inductively coupled plasma mass spectrometry for elemental analysis of solid samples
    Shuttleworth, S
    Kremser, DT
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1998, 13 (08) : 697 - 699
  • [5] Elemental Fractionation Studies of Biological Samples Using Laser Ablation Inductively Coupled Plasma Mass Spectrometry
    Li Qing
    Zhang Guo-Xia
    Chen Yi-Rui
    Wang Zheng
    Ding Chuan-Xian
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2017, 45 (06) : 868 - 873
  • [6] Application of laser ablation inductively coupled plasma mass spectrometry for the isotopic analysis of single uranium particles
    Varga, Zsolt
    ANALYTICA CHIMICA ACTA, 2008, 625 (01) : 1 - 7
  • [7] Isotopic uranium determination by inductively coupled plasma atomic emission spectrometry using conventional and laser ablation sample introduction
    Goodall, PS
    Johnson, SG
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1996, 11 (01) : 57 - 60
  • [9] In situ determination of trace elements in melt inclusions using laser ablation inductively coupled plasma sector field mass spectrometry
    Zhang, Le
    Ren, Zhong-Yuan
    Xia, Xiao-Ping
    Yang, Qing
    Hong, Lu-Bing
    Wu, Dan
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2019, 33 (04) : 361 - 370
  • [10] Quantitative Determination of Elements in Plant Samples by Laser Ablation Inductively Coupled Plasma Mass Spectrometry
    Wang Qi
    Zhang Wen
    Wang Li-yun
    Liu Yong-sheng
    Hu Sheng-hong
    Hu Zhao-chu
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2011, 31 (12) : 3379 - 3383