Present limitations of CVD diamond detectors for IMRT applications

被引:24
|
作者
De Angelis, C. [1 ]
Casati, M. [2 ]
Bruzzi, M. [3 ]
Onori, S. [1 ]
Bucciolini, M. [2 ]
机构
[1] Ist Super Sanita, INFN, Dipartimento Tecnol & Salute, I-00161 Rome, Italy
[2] Univ Florence, INFN, Dipartimento Fisiopatol, I-50121 Florence, Italy
[3] Univ Florence, INFN, Dipartimento Engn, I-50121 Florence, Italy
关键词
CVD diamond detector; solid-state dosimetry; IMRT dosimetry;
D O I
10.1016/j.nima.2007.08.186
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The aim of the work was to test the suitability of chemical vapor deposited (CVD) diamond detectors for dosimetry in IMRT fields. We used in-house CVD detectors prepared with state-of-the-art polycrystalline diamond films (Element Six Ltd., UK). The parameters considered were time stability, dynamic response, dose-rate dependence and energy dependence. Output factors and TPR were measured in conventional photon fields and dose measurements were performed in IMRT fields using the step-and-shoot technique. Results prove that CVD diamond detectors are suitable for dosimetry in conventional treatments, but they still do not fit the IMRT dosimetry requirements, mainly because of their slow dynamic response. In particular, the slow dynamics affects linearity at low Monitor Units and renders it impossible to follow the sharp transients of IMRT fields. Time stability and dose-rate dependence as well must be improved to reduce their influence on dose assessment. (c) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:195 / 203
页数:9
相关论文
共 50 条
  • [22] First applications of CVD-diamond detectors in heavy-ion experiments
    Berdermann, E
    Blasche, K
    Moritz, P
    Stelzer, H
    Voss, B
    Zeytouni, F
    NUCLEAR PHYSICS B-PROCEEDINGS SUPPLEMENTS, 1999, 78 : 533 - 539
  • [23] CVD diamond pixel detectors for LHC experiments
    Wedenig, R
    Adam, W
    Bauer, C
    Berdermann, E
    Bergonzo, P
    Bogani, F
    Borchi, E
    Brambilla, A
    Bruzzi, M
    Colledani, C
    Conway, J
    Dabrowski, W
    Delpierre, P
    Deneuville, A
    Dulinski, W
    van Eijk, B
    Fallou, A
    Fizzotti, F
    Foulon, F
    Friedl, M
    Gan, KK
    Gheeraert, E
    Grigoriev, E
    Hallewell, G
    Hall-Wilton, R
    Han, S
    Hartjes, F
    Hrubec, J
    Husson, D
    Kagan, H
    Kania, D
    Kaplon, J
    Karl, C
    Kass, R
    Knöpfle, KT
    Krammer, M
    Logiudice, A
    Lu, R
    Manfredi, PF
    Manfredotti, C
    Marshall, RD
    Meier, D
    Mishina, M
    Oh, A
    Pan, LS
    Palmieri, VG
    Pernicka, M
    Peitz, A
    Pirollo, S
    Polesello, P
    NUCLEAR PHYSICS B-PROCEEDINGS SUPPLEMENTS, 1999, 78 : 497 - 504
  • [24] Radiation hardness studies of CVD diamond detectors
    Bauer, C
    Baumann, I
    Colledani, C
    Conway, J
    Delpierre, P
    Djama, F
    Dulinski, W
    Fallou, A
    Gan, KK
    Gilmore, RS
    Grigoriev, E
    Hallewell, G
    Han, S
    Hessing, T
    Honschied, K
    Hrubec, J
    Husson, D
    Kagan, H
    Kania, D
    Kass, R
    Kinnison, W
    Knopfle, KT
    Krammer, M
    Llewellyn, TJ
    Manfredi, PF
    Pan, LS
    Pernegger, H
    Pernicka, M
    Plano, R
    Re, V
    Roe, S
    Rudge, A
    Schaeffer, M
    Schnetzer, S
    Somalwar, S
    Speziali, V
    Stone, R
    Tapper, RJ
    Tesarek, R
    Trischuk, W
    Turchetta, R
    Thomson, GB
    Wagner, R
    Weilhammer, P
    White, C
    Ziock, H
    Zoeller, M
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1995, 367 (1-3): : 207 - 211
  • [25] A study of the thermoluminescent properties of CVD diamond detectors
    Marczewska, B
    Bilski, P
    Nesladek, M
    Olko, P
    Rebisz, M
    Waligórski, MPR
    PHYSICA STATUS SOLIDI A-APPLIED RESEARCH, 2002, 193 (03): : 470 - 475
  • [26] Response of CVD diamond detectors to alpha radiation
    Souw, EK
    Meilunas, RJ
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1997, 400 (01): : 69 - 86
  • [27] Development of CVD diamond detectors for clinical dosimetry
    Piliero, M. A.
    Hugtenburg, R. P.
    Ryde, S. J. S.
    Oliver, K.
    RADIATION PHYSICS AND CHEMISTRY, 2014, 104 : 10 - 14
  • [28] CVD diamond detectors for radiation pulse characterisation
    Foulon, F
    Bergonzo, P
    Jany, C
    Gicquel, A
    Pochet, T
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 380 (1-2): : 42 - 45
  • [29] Fusion neutron diagnostics with CVD diamond detectors
    Weiss, Christina
    Griesmayer, Erich
    FUSION ENGINEERING AND DESIGN, 2024, 203
  • [30] Reproducibility of CVD Diamond Detectors for Radiotherapy Dosimetry
    Betzel, G.
    Lansley, S.
    Mckay, D.
    Meyer, J.
    MEDICAL PHYSICS, 2012, 39 (06) : 3739 - 3739