Development of an on-line automated sample clean-up method and liquid chromatography-tandem mass spectrometry analysis: application in an in vitro proteolytic assay

被引:10
|
作者
Drexler, D
Barlow, DJ
Falk, P
Cantone, J
Hernandez, D
Ranasinghe, A
Sanders, M
Warrack, B
McPhee, F
机构
[1] Bristol Myers Squibb Co, Pharmaceut Res Inst, Pharmaceut Candidate Optimizat Discovery Analyt, Wallingford, CT 06492 USA
[2] Bristol Myers Squibb Co, Pharmaceut Res Inst, Virol Discovery, Wallingford, CT 06492 USA
[3] Bristol Myers Squibb Co, Pharmaceut Res Inst, Pharmaceut Candidate Optimizat Discovery Analyt, Princeton, NJ 08543 USA
关键词
automation; on-line sample clean-up; ion exchange and peptide microtraps; liquid chromatography-tandem mass spectrometry (LC-MS; MS); proteolytic assay;
D O I
10.1007/s00216-005-0263-7
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Fluorescence detection has been a method of choice in industry for screening assays, including identification of enzyme inhibitors, owing to its high-throughput capabilities, excellent reproducibility, and sensitivity. Occasionally, inhibitors are identified that challenge the fluorescence assay limit, necessitating the development of more sensitive detection methods to assess these compounds. For data mining purposes, however, original assay conditions may be required. A direct method transfer to highly sensitive and specific LC-MS-based methods has not always been possible due to the presence of MS-incompatible neutral detergents and non-volatile salts in the assay matrix. Utilizing an in vitro proteolytic screening assay for the serine protease hepatitis C virus (HCV) nonstructural (NS) 3 protease as a test case, we report the development of an automated sample clean-up procedure implemented on-line with liquid chromatography-tandem mass spectrometry (LC-MS/MS) analysis to complement fluorescence detection. Ion exchange and peptide microtraps were employed to remove MS-incompatible assay matrix components. Three protease inhibitors were used to validate the MS/MS method. Comparable potencies were achieved for these compounds when assessed by fluorescence and MS/MS detection. Furthermore, four-fold less enzyme could be utilized when employing the MS/MS method compared to fluorescence detection. The longer analysis time, however, resulted in reduced sample capacity. The potency of our designed HCV NS3 protease inhibitors are thus routinely evaluated using a continuous fluorescence-based assay. Only pertinent inhibitors approaching the fluorescence assay sensitivity limit are subsequently analyzed further by LC-MS/MS. This methodology allows us to maintain a database and to compare results independent of the detection method. Despite the relatively slow sample turnaround time of this LC-MS approach, the versatility of the automated on-line clean-up procedure and sample analysis can be applied to assays containing reagents which were historically considered to be MS incompatible.
引用
收藏
页码:1145 / 1154
页数:10
相关论文
共 50 条
  • [11] Analysis of chloramphenicol in honey by on-line pretreatment liquid chromatography-tandem mass spectrometry
    Shin-ichi Kawano
    Hong-Yuan Hao
    Yuki Hashi
    Jin-Ming Lin
    Chinese Chemical Letters, 2015, 26 (01) : 36 - 38
  • [12] Simultaneous separation, clean-up and analysis of musty odorous compounds in wines by on-line coupling of a pervaporation unit to gas chromatography-tandem mass spectrometry
    Gómez-Ariza, JL
    García-Barrera, T
    Lorenzo, E
    ANALYTICA CHIMICA ACTA, 2004, 516 (1-2) : 165 - 170
  • [13] Analysis of sulfonamide residues in bovine liver by liquid chromatography-tandem mass spectrometry without chemical extraction or clean-up procedures
    Hoff, Rodrigo
    ANALYTICAL BIOCHEMISTRY, 2020, 611
  • [14] Use of gel permeation chromatography for clean-up in the analysis of coccidiostats in eggs by liquid chromatography–tandem mass spectrometry
    J. Chico
    A. Rúbies
    F. Centrich
    R. Companyó
    M. D. Prat
    M. Granados
    Analytical and Bioanalytical Chemistry, 2013, 405 : 4777 - 4786
  • [15] Lack of specificity for the analysis of raltegravir using online sample clean-up liquid chromatography-electrospray tandem mass spectrometry
    Jourdil, Jean Francois
    Bartoli, Mireille
    Stanke-Labesque, Francoise
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2009, 877 (29): : 3734 - 3738
  • [16] Determination of Hexabromocyclododecanes in Fish Using Modified QuEChERS Method with Efficient Extract Clean-Up Prior to Liquid Chromatography-Tandem Mass Spectrometry
    Oksova, Linda
    Tolgyessy, Peter
    SEPARATIONS, 2020, 7 (03)
  • [17] Effective clean-up and ultra high-performance liquid chromatography-tandem mass spectrometry for isoflavone determination in legumes
    Vila-Donat, Pilar
    Caprioli, Giovanni
    Maggi, Filippo
    Ricciutelli, Massimo
    Torregiani, Elisabetta
    Vittori, Sauro
    Sagratini, Gianni
    FOOD CHEMISTRY, 2015, 174 : 487 - 494
  • [18] Confirmatory analysis of stanozolol metabolites in bovine, pig and sheep urines using an optimized clean-up and liquid chromatography-tandem mass spectrometry
    Toelgyesi, Adam
    Sharma, Virender K.
    Fekete, Jeno
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2014, 88 : 45 - 52
  • [19] Determination of Three Plant Neurotransmitters by On-line Liquid Chromatography-Tandem Mass Spectrometry
    Cao Jin
    Miao Lan
    Liu Jian-Xun
    Qi Xue-Yong
    Murch, Susan J.
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2011, 39 (03) : 367 - 371
  • [20] Micro-Sample Analysis of Capsaicin by Liquid Chromatography-Tandem Mass Spectrometry Assay and Application to Oral Absorption Study
    Choi, Y.
    Shin, B. S.
    Kim, M. G.
    Shin, S. W.
    Choo, H. W.
    Kim, T. H.
    Jang, H. C.
    Yoo, S. D.
    ACTA CHROMATOGRAPHICA, 2013, 25 (02) : 363 - 376