Electron dosimetry based on the absorbed dose to water concept:: A comparison of the AAPM TG-51 and DIN 6800-2 protocols

被引:10
|
作者
Dohm, OS
Christ, G
Nüsslin, F
Schüle, E
Bruggmoser, G
机构
[1] Univ Tubingen, Klin Radioonkol, Abt Med Phys, D-72076 Tubingen, Germany
[2] PTW Freiburg, D-79115 Freiburg, Germany
[3] Univ Freiburg Klinikum, Abt Strahlentherapie, D-79106 Freiburg, Germany
关键词
electron dosimetry; absorbed dose to water; Roos chamber; perturbation correction;
D O I
10.1118/1.1414008
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The dosimetry protocols DIN 6800-2 and AAPM TG-51, both based on the absorbed dose to water concept, are compared in their theoretical background and in their application to electron dosimetry. The agreement and disagreement in correction factors and energy parameters used in both protocols will be shown and discussed. Measurements with three different types of ionization chambers were performed and evaluated according to both protocols. As a result the perturbation correction factor P-wall(60Co) for the Roos chamber was determined to 1.024 +/- 0.5%. (C) 2001 American Association of Physicists in Medicine.
引用
收藏
页码:2258 / 2264
页数:7
相关论文
共 37 条
  • [31] Determination of absorbed dose to water in reference conditions for radiotherapy kilovoltage x-rays between 10 and 300 kV: a comparison of the data in the IAEA, IPEMB, DIN and NCS dosimetry protocols
    Peixoto, JGP
    Andreo, P
    PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (03): : 563 - 575
  • [32] Comparison of Absorbed Dose Determination Using the Updated IAEA TRS398 CoP with TRS398 CoP and AAPM TG51 Addendum Protocol
    diMayorca, M.
    Tavakoli, M.
    Wadi-Ramahi, S.
    Andreo, P.
    Huq, M.
    MEDICAL PHYSICS, 2022, 49 (06) : E425 - E425
  • [33] Comparison of the absorbed dose in clinical electron beams, determined following the recommendations of the NCS report 5, IAEA TRS398 and TG51
    Nafaa, L
    Palmans, H
    de Patoul, N
    Denis, J
    Tomsej, M
    Vynckier, S
    MEDICAL PHYSICS, 2002, 29 (06) : 1200 - 1200
  • [34] Comparing calibration methods of electron beams using plane-parallel chambers with absorbed-dose to water based protocols
    Stewart, KJ
    Seuntjens, JP
    MEDICAL PHYSICS, 2002, 29 (03) : 284 - 289
  • [35] The IPEM code of practice for electron dosimetry for radiotherapy beams of initial energy from 4 to 25 MeV based on an absorbed dose to water calibration
    Thwaites, DI
    DuSautoy, AR
    Jordan, T
    McEwen, MR
    Nisbet, A
    Nahum, AE
    Pitchford, WG
    PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (18): : 2929 - 2970
  • [36] Measurement of absorbed dose to water around an electronic brachytherapy source. Comparison of two dosimetry systems: lithium formate EPR dosimeters and radiochromic EBT2 film
    Adolfsson, Emelie
    White, Shane
    Landry, Guillaume
    Lund, Eva
    Gustafsson, Hakan
    Verhaegen, Frank
    Reniers, Brigitte
    Tedgren, Asa Carlsson
    Carlsson, Gudrun Alm
    PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (09): : 3869 - 3882
  • [37] Comparison between absorbed dose to water standards established by water calorimetry at the LNE-LNHB and by application of international air-kerma based protocols for kilovoltage medium energy x-rays
    Perichon, N.
    Rapp, B.
    Denoziere, M.
    Daures, J.
    Ostrowsky, A.
    Bordy, J-M
    PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (09): : 2787 - 2806