Biomechanical Loading Comparison between Titanium and Bioactive Resorbable Screw Systems for Fixation of Intracapsular Condylar Head Fractures

被引:7
|
作者
Sukegawa, Shintaro [1 ,2 ]
Yamamoto, Norio [2 ]
Nakano, Keisuke [3 ]
Takabatake, Kiyofumi [3 ]
Kawai, Hotaka [3 ]
Kanno, Takahiro [4 ]
Nagatsuka, Hitoshi [3 ]
Furuki, Yoshihiko [1 ]
机构
[1] Kagawa Prefectural Cent Hosp, Dept Oral & Maxillofacial Surg, 1-2-1 Asahi Machi, Takamatsu, Kagawa 7608557, Japan
[2] Kagawa Prefectural Cent Hosp, Dept Orthopaed Surg, Takamatsu, Kagawa 600065, Japan
[3] Okayama Univ, Dept Oral Pathol & Med, Grad Sch Med Dent & Pharmaceut Sci, Okayama 7008530, Japan
[4] Shimane Univ, Dept Oral & Maxillofacial Surg, Fac Med, Matsue, Shimane 6930021, Japan
关键词
titanium screw; bioactive resorbable screw; biomechanical loading evaluation; fracture fixation; intracapsular condylar head fracture; POLY-L-LACTIDE; HYDROXYAPATITE HA PARTICLES; INTERNAL-FIXATION; FORGED COMPOSITES; BIORESORBABLE DEVICES; MANDIBULAR CONDYLE; SURGICAL-TREATMENT; OPEN REDUCTION; OSTEOSYNTHESIS; PLATE;
D O I
10.3390/ma13143153
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Osteosynthesis resorbable materials made of uncalcined and unsintered hydroxyapatite (u-HA) particles, poly-L-lactide (PLLA), are bioresorbable, and these materials have feasible bioactive/osteoconductive capacities. However, their strength and stability for fixation in mandibular condylar head fractures remain unclear. This in vitro study aimed to assess the biomechanical strength of u-HA/PLLA screws after the internal fixation of condylar head fractures. To evaluate their biomechanical behavior, 32 hemimandible replicas were divided into eight groups, each consisting of single-screw and double-screw fixations with titanium or u-HA/PLLA screws. A linear load was applied as vertical and horizontal load to each group to simulate the muscular forces in condylar head fractures. Samples were examined for 0.5, 1, 2, and 3-mm displacement loads. Two screws were needed for stable fixation of the mandibular condylar head fracture during biomechanical evaluation. After screw fixation for condylar head fractures, the titanium screws model was slightly more resistant to vertical and horizontal movement with a load for a small displacement than the u-HA/PLLA screws model. There was no statistically significant difference with load for large displacements. The u-HA/PLLA screw has a low mechanical resistance under small displacement loading compared with titanium within the limits of the mandibular head fracture model study.
引用
收藏
页数:12
相关论文
共 50 条
  • [31] Biomechanical comparison of stainless steel and titanium nails for fixation of simulated femoral fractures
    Mahar, AT
    Lee, SS
    Lalonde, FD
    Impelluso, T
    Newton, PO
    JOURNAL OF PEDIATRIC ORTHOPAEDICS, 2004, 24 (06) : 638 - 641
  • [32] Evaluation of internal fixation techniques for condylar head fractures: A finite element analysis and comparison
    Bu, Lingtong
    Chen, Qin
    Huang, Kai
    Zhao, Xiaowen
    Zheng, Jisi
    Qiu, Yating
    Yang, Chi
    ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY, 2022, 133 (05): : E96 - E104
  • [33] Comparison of resorbable plates and titanium plates for fixation stability of combined mandibular symphysis and angle fractures
    Lim, Ho-Yong
    Jung, Chang-Hwa
    Kim, Seong-Yong
    Cho, Jin-Yong
    Ryu, Jae-Young
    Kim, Hyeon-Min
    JOURNAL OF THE KOREAN ASSOCIATION OF ORAL AND MAXILLOFACIAL SURGEONS, 2014, 40 (06) : 285 - 290
  • [34] A biomechanical analysis of the stability of titanium bone fixation systems in proximal phalangeal fractures
    Prevel, CD
    Katona, T
    Eppley, BL
    Moore, K
    McCarty, M
    Ge, J
    ANNALS OF PLASTIC SURGERY, 1996, 37 (05) : 473 - 481
  • [35] Fixation of fractures of the condylar head of the mandible with a new magnesium-alloy biodegradable cannulated headless bone screw
    Leonhardt, H.
    Franke, A.
    McLeod, N. M. H.
    Lauer, G.
    Nowak, A.
    BRITISH JOURNAL OF ORAL & MAXILLOFACIAL SURGERY, 2017, 55 (06): : 623 - 625
  • [36] Biomechanical Comparison of a Novel 3-Screw Fixation vs Conventional 2-Screw Fixation of Calcaneal Tuberosity Avulsion Fractures
    Hong, Choon Chiet
    Tan, Jun-Hao
    Ramruttun, Amit Kumarsing
    Pearce, Christopher Jon
    FOOT & ANKLE INTERNATIONAL, 2022, 43 (12) : 1562 - 1568
  • [37] Biomechanical comparison between various screw fixation angles for Latarjet procedure: a cadaveric biomechanical study
    Hsu, Kai-Lan
    Yeh, Ming-Long
    Kuan, Fa-Chuan
    Hong, Chih-Kai
    Chuang, Hao-Chun
    Wang, Wei-Ming
    Su, Wei-Ren
    JOURNAL OF SHOULDER AND ELBOW SURGERY, 2022, 31 (09) : 1947 - 1956
  • [38] Comparison of monoplanar and polyaxial screw fixation systems in percutaneous intermediate fixation for thoracolumbar fractures
    Liangliang Huang
    Chengjie Xiong
    Zhongyi Guo
    Qiuyu Yu
    Feng Xu
    Hui Kang
    BMC Musculoskeletal Disorders, 23
  • [39] Comparison of monoplanar and polyaxial screw fixation systems in percutaneous intermediate fixation for thoracolumbar fractures
    Huang, Liangliang
    Xiong, Chengjie
    Guo, Zhongyi
    Yu, Qiuyu
    Xu, Feng
    Kang, Hui
    BMC MUSCULOSKELETAL DISORDERS, 2022, 23 (01)
  • [40] Biomechanical comparison of five cannulated screw fixation strategies for young vertical femoral neck fractures
    Jiang, Dajun
    Zhan, Shi
    Wang, Lei
    Shi, Lewis L.
    Ling, Ming
    Hu, Hai
    Jia, Weitao
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2021, 39 (08) : 1669 - 1680