Validation of a novel respiratory rate monitor based on exhaled humidity

被引:17
|
作者
Niesters, M. [1 ]
Mahajan, R. [2 ]
Olofsen, E. [1 ]
Boom, M. [1 ]
Garcia del Valle, S. [3 ]
Aarts, L. [1 ]
Dahan, A. [1 ]
机构
[1] Leiden Univ Med Ctr, Dept Anesthesiol, NL-2300 RC Leiden, Netherlands
[2] Univ Nottingham, Queens Med Ctr, Div Anaesthesia & Intens Care, Nottingham NG7 2UH, England
[3] Hosp Univ Fdn Alcorcon, Area Anestesia & Cuidados Crit, Madrid, Spain
关键词
complications; monitoring; postoperative care; recovery room; respiratory depression; respiratory rate; OXYGEN; PREVENTION; AGREEMENT; ACCURACY; CARE;
D O I
10.1093/bja/aes275
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Postoperative monitoring of ventilation is largely restricted to the measurement of haemoglobin-oxygen saturation (Sp(o2)) and respiratory rate (RR) derived from the ECG. Sp(o2) measurement is inadequate when used with supplemental oxygen and ECG-derived RR is subject to artifacts. A new monitor measures RR by quantifying the humidity of exhaled air (respiR8). The accuracy of the system was tested using a breathing simulator. In healthy volunteers, the respiR8 monitor was compared with two other methods of measuring RR: capnometry and counting of thoracic breathing movements. The ability of the monitor to track changes in RR resulting from the infusion of 2.5 g kg(1) fentanyl was assessed and compared with RR measured from a validated flow measurement system. The RR in 50 postoperative patients measured with the respiR8 was compared with that derived from the ECG. RR values were compared by population-based BlandAltman analyses. The respiR8 monitor was accurate in the range required in clinical practice. There was a close agreement between RR from respiR8, capnometry, and manual counting of respiratory movements without bias (limits of agreement 1 bpm). The respiR8 monitor was well able to accurately track RR changes from fentanyl. In postoperative patients, RR from respiR8 and ECG had a bias of 1.7 (5.7) bpm due to greater RR values observed from the ECG due to artifacts. The respiR8 gives an accurate measurement of RR and is useful in postoperative care.
引用
收藏
页码:981 / 989
页数:9
相关论文
共 50 条
  • [41] Multichannel esophageal signals to monitor respiratory rate in preterm infants
    Buergin, Corine
    Simmen, Patrizia
    Gupta, Nishant
    Suter, Lilian
    Kreuzer, Samuel
    Haeberlin, Andreas
    Schulzke, Sven M.
    Trachsel, Daniel
    Niederhauser, Thomas
    Jost, Kerstin
    PEDIATRIC RESEARCH, 2022, 91 (03) : 572 - 580
  • [42] Standard pulse oximeters can be used to monitor respiratory rate
    Leonard, P
    Beattie, TF
    Addison, PS
    Watson, JN
    EMERGENCY MEDICINE JOURNAL, 2003, 20 (06) : 524 - 525
  • [43] Development of the 'BreathEasy' contactless portable respiratory rate monitor (CPRM)
    Daw, Will
    Kingshott, Ruth
    Scott, Alison
    Saatchi, Reza
    Elphick, Heather
    EUROPEAN RESPIRATORY JOURNAL, 2015, 46
  • [44] Validation of a new disposable respiratory rate counter
    Townsend-Rose, CR
    Breakell, A
    Warburton, CJ
    Calverley, PMA
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 1999, 159 (03) : A832 - A832
  • [45] CLINICAL VALIDATION OF A NOVEL CUFFLESSS BLOOD PRESSURE MONITOR
    Boubouchairopoulou, N.
    Kollias, A.
    Lagou, S.
    Anestis, P.
    Stergiou, G. S.
    JOURNAL OF HYPERTENSION, 2016, 34 : E163 - E163
  • [46] Validation of an activity monitor during sleep in patients with chronic respiratory disorders
    Gloeckl, Rainer
    Damisch, Thomas
    Prinzen, Jos
    van Lummel, Rob
    Pengel, Evert
    Schoenheit-Kenn, Ursula
    Kenn, Klaus
    RESPIRATORY MEDICINE, 2015, 109 (02) : 286 - 288
  • [47] Implementation of a continuous respiratory rate monitor intended for low-resource settings: a novel device pilot study
    Amin, A. A.
    Mtenthaonga, P.
    Coyle, J.
    Mjumira, R.
    Mangwiro, L.
    Oden, M.
    Richards-Kortum, R.
    Dube, Q.
    Kawaza, K.
    Patil, M.
    AMERICAN JOURNAL OF THE MEDICAL SCIENCES, 2025, 369 : S285 - S286
  • [48] VALIDATION OF THE DATEX DELTATRAC METABOLIC MONITOR BY SIMULATION OF RESPIRATORY AND METABOLIC PARAMETERS
    WEYLAND, W
    WEYLAND, A
    LIPPMANN, M
    BRAUN, U
    ANAESTHESIST, 1988, 37 (11): : 716 - 716
  • [49] A breath sampling system assessing the influence of respiratory rate on exhaled breath composition
    Lomonaco, T.
    Salvo, P.
    Ghimenti, S.
    Biagini, D.
    Bellagambi, F.
    Fuoco, R.
    Di Francesco, F.
    2015 37TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2015, : 7618 - 7621
  • [50] Microcontroller Based Heart Rate Monitor
    Fezari, Mohamed
    Bousbia-Salah, Mounir
    Bedda, Mouldi
    INTERNATIONAL ARAB JOURNAL OF INFORMATION TECHNOLOGY, 2008, 5 (04) : 409 - 413