Comparison of Balloon Catheter Ablation Technologies for Pulmonary Vein Isolation: The Laser Versus Cryo Study

被引:41
|
作者
Bordignon, Stefano [1 ]
Chun, K. R. Julian [1 ]
Gunawardene, Melanie [1 ]
Fuernkranz, Alexander [1 ]
Urban, Verena [1 ]
Schulte-Hahn, Britta [1 ]
Nowak, Bernd [1 ]
Schmidt, Boris [1 ]
机构
[1] Cardioangiol Ctr Bethanien, D-60431 Frankfurt, Germany
关键词
atrial fibrillation; catheter ablation; cryoballoon; laser; phrenic nerve; PAROXYSMAL ATRIAL-FIBRILLATION; ANTIARRHYTHMIC-DRUG THERAPY; BIG CRYOBALLOON TECHNIQUE; LUMEN MAPPING CATHETER; TRIAL; EFFICACY; FEASIBILITY; MANAGEMENT; INSIGHTS; SYSTEM;
D O I
10.1111/jce.12192
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Laser- Versus CryoBalloon for PVI. Introduction: Balloon catheters have been developed to facilitate pulmonary vein isolation (PVI) in patients with paroxysmal atrial fibrillation (PAF). We sought to compare the safety and efficacy of the cryoballoon (CB) and the laserballoon (LB) in a pilot study. Methods and Results: One hundred and forty patients with drug-refractory PAF were prospectively allocated in a 1: 1 fashion to undergo a PVI procedure with the 28 mm CB or the LB and were followed for 12 months using 3-day Holter ECG recording. The primary efficacy endpoint was a documented AF recurrence >= 30 seconds between 90 and 365 days after the index ablation. In total, 269 of 270 PVs (99.6%) and 270 of 273 PVs (98.9%) were acutely isolated in the CB and LB group, respectively. Mean procedural time was 136 +/- 30 minutes for the CB group and 144 +/- 33 minutes for the LB group (P = 0.13). Mean fluoroscopy time was longer in the CB group (21 +/- 9 minutes vs 15 +/- 6 minutes; P < 0.001). During 12 months follow-up, 37% of patients in the CB group and 27% in the LB group experienced an AF recurrence (P = 0.18). Phrenic nerve palsies occurred in 5.7% (CB) and 4.2% (LB) of patients, respectively. Conclusion: Balloon catheters are a viable option to safely perform a PVI procedure in patients with drug-refractory PAF. Ninety-nine percent of PVs may be acutely isolated with a single balloon catheter. The AF free survival rate after a single ablation procedure was not statistically different between groups.
引用
收藏
页码:987 / 994
页数:8
相关论文
共 50 条
  • [41] Durability of pulmonary vein isolation by various kind of ablation catheter
    Kobori, A.
    Sasaki, Y.
    Matsumoto, Y.
    Ishizu, K.
    Nakashima, M.
    Ishibashi, K.
    Murai, R.
    Oota, M.
    Kim, K.
    Yamane, T.
    Ehara, N.
    Kinoshita, M.
    Kaji, S.
    Furukawa, Y.
    EUROPEAN HEART JOURNAL, 2016, 37 : 632 - 632
  • [42] Pulmonary Vein Isolation with a Novel Direct Visualization Ablation Catheter
    McElderry, Tom
    Oley, Leslie
    Malchano, Zaxh
    Wylie, Bryan
    Yamada, Takumi
    Kay, G. N.
    CIRCULATION, 2010, 122 (21)
  • [43] EXTENT AND LEVEL OF PULMONARY VEIN ISOLATION WITH THE LASER BALLOON ABLATION IN PATIENTS WITH PAROXYSMAL ATRIAL FIBRILLATION
    Tondo, C.
    Fassini, G.
    Dello Russo, A.
    Casella, M.
    Al-Mohani, G.
    Tundo, F.
    Riva, S.
    Moltrasio, M.
    Barrett, C.
    Danik, S.
    Ruskin, J.
    Mansour, M.
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2011, 22 : S110 - S110
  • [44] Pulmonary vein isolation using a compliant endoscopic laser balloon ablation system in a swine model
    Edward P. Gerstenfeld
    John Michele
    Journal of Interventional Cardiac Electrophysiology, 2010, 29 : 1 - 9
  • [45] Pulmonary vein isolation using a compliant endoscopic laser balloon ablation system in a swine model
    Gerstenfeld, Edward P.
    Michele, John
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2010, 29 (01) : 1 - 9
  • [46] Comparison of Radiofrequency Ablation vs. Visually-Guided Laser Balloon Ablation in Achieving Chronic Pulmonary Vein Isolation in a Porcine Model
    Dukkipati, Srinivas
    d'Avila, Andre
    Neuzil, Petr
    Reddy, Vivek Y.
    CIRCULATION, 2009, 120 (18) : S706 - S706
  • [47] Initial Experience With a New Ablation Catheter Using Laser Energy for Electrical Pulmonary Vein Isolation
    Osca, Joaquin
    Cano, Oscar
    Andres, Ana
    Alonso, Pau
    Jose Sancho-Tello, Maria
    Olague, Jose
    REVISTA ESPANOLA DE CARDIOLOGIA, 2014, 67 (12): : 1061 - 1062
  • [48] Atrial Fibrillation Catheter Ablation Versus Surgical Ablation Treatment After a Failed Initial Pulmonary Vein Isolation Procedure: A Randomized Study
    Pokushalov, Evgeny
    Romanov, Alexander
    Elesin, Dmitry
    Bogachev-Prokonhiev, Alexander
    Losik, Denis
    Bairamova, Sevda
    Karaskov, Alexander
    Mittal, Suneet
    Steinberg, Jonathan S.
    CIRCULATION, 2013, 128 (22)
  • [49] Pulmonary vein isolation with the Pulmonary Vein Ablation Catheter versus conventional point-by-point ablation using the CARTO system: a large, comparative, single-centre study
    De Greef, Y.
    Segers, V. F. M.
    Schwagten, B.
    De Keulenaer, G. W.
    Stockman, D.
    ACTA CARDIOLOGICA, 2011, 66 (05) : 682 - 682
  • [50] Development of a new HotBalloonTM ablation catheter equipped with a balloon surface temperature monitoring sensor for pulmonary vein isolation
    Tsukamoto K.
    Tanahashi A.
    Harada H.
    Takaoka M.
    Sohara H.
    Journal of Medical Engineering and Technology, 2022, 46 (08): : 687 - 692