Comparison of Balloon Catheter Ablation Technologies for Pulmonary Vein Isolation: The Laser Versus Cryo Study

被引:41
|
作者
Bordignon, Stefano [1 ]
Chun, K. R. Julian [1 ]
Gunawardene, Melanie [1 ]
Fuernkranz, Alexander [1 ]
Urban, Verena [1 ]
Schulte-Hahn, Britta [1 ]
Nowak, Bernd [1 ]
Schmidt, Boris [1 ]
机构
[1] Cardioangiol Ctr Bethanien, D-60431 Frankfurt, Germany
关键词
atrial fibrillation; catheter ablation; cryoballoon; laser; phrenic nerve; PAROXYSMAL ATRIAL-FIBRILLATION; ANTIARRHYTHMIC-DRUG THERAPY; BIG CRYOBALLOON TECHNIQUE; LUMEN MAPPING CATHETER; TRIAL; EFFICACY; FEASIBILITY; MANAGEMENT; INSIGHTS; SYSTEM;
D O I
10.1111/jce.12192
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Laser- Versus CryoBalloon for PVI. Introduction: Balloon catheters have been developed to facilitate pulmonary vein isolation (PVI) in patients with paroxysmal atrial fibrillation (PAF). We sought to compare the safety and efficacy of the cryoballoon (CB) and the laserballoon (LB) in a pilot study. Methods and Results: One hundred and forty patients with drug-refractory PAF were prospectively allocated in a 1: 1 fashion to undergo a PVI procedure with the 28 mm CB or the LB and were followed for 12 months using 3-day Holter ECG recording. The primary efficacy endpoint was a documented AF recurrence >= 30 seconds between 90 and 365 days after the index ablation. In total, 269 of 270 PVs (99.6%) and 270 of 273 PVs (98.9%) were acutely isolated in the CB and LB group, respectively. Mean procedural time was 136 +/- 30 minutes for the CB group and 144 +/- 33 minutes for the LB group (P = 0.13). Mean fluoroscopy time was longer in the CB group (21 +/- 9 minutes vs 15 +/- 6 minutes; P < 0.001). During 12 months follow-up, 37% of patients in the CB group and 27% in the LB group experienced an AF recurrence (P = 0.18). Phrenic nerve palsies occurred in 5.7% (CB) and 4.2% (LB) of patients, respectively. Conclusion: Balloon catheters are a viable option to safely perform a PVI procedure in patients with drug-refractory PAF. Ninety-nine percent of PVs may be acutely isolated with a single balloon catheter. The AF free survival rate after a single ablation procedure was not statistically different between groups.
引用
收藏
页码:987 / 994
页数:8
相关论文
共 50 条
  • [1] Pulmonary Vein Isolation Using Laser Balloon as Compared to Standard Radiofrequency Catheter Ablation
    Koopman, Pieter
    Vanhentenrijk, Simon
    Schurmans, Joris
    Dilling, Dagmara
    Vijgen, Johan
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2017, 28 (05) : 600 - 601
  • [2] Pulmonary vein isolation using laser balloon as compared to standard radiofrequency catheter ablation
    Koopman, Pieter
    Vanhentenrijk, Simon
    Schurmans, Joris
    Dining, Dagmara
    Vijgen, Johan
    ACTA CARDIOLOGICA, 2017, 72 (01) : 89 - 90
  • [3] Pulmonary Vein Isolation with a Novel Size-Adjustable Cryo-Balloon Catheter: A Tailored Ablation Protocol
    Teumer, Yannick
    Hilgarth, Franziska
    Katov, Lyuboslav
    Melnic, Rima
    Rottbauer, Wolfgang
    Bothner, Carlo
    Weinmann, Karolina
    JOURNAL OF CLINICAL MEDICINE, 2024, 13 (08)
  • [4] Efficacy of pulmonary vein isolation with a novel hot balloon ablation catheter
    Evonich, Rudolph F.
    Nori, David M.
    Haines, David E.
    JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2012, 34 (01) : 29 - 36
  • [5] Efficacy of pulmonary vein isolation with a novel hot balloon ablation catheter
    Rudolph F. Evonich
    David M. Nori
    David E. Haines
    Journal of Interventional Cardiac Electrophysiology, 2012, 34 : 29 - 36
  • [6] Wide area circumferential ablation for pulmonary vein isolation using radiofrequency versus laser balloon ablation
    Skeete, Jamario
    Sharma, Parikshit S.
    Kenigsberg, David
    Pietrasik, Grzegorz
    Osman, Ahmed F.
    Ravi, Venkatesh
    Du-Fay-de-Lavallaz, Jeanne M.
    Post, Zoe
    Wasserlauf, Jeremiah
    Larsen, Timothy R.
    Krishnan, Kousik
    Trohman, Richard
    Huang, Henry D.
    JOURNAL OF ARRHYTHMIA, 2022, 38 (03) : 336 - 345
  • [7] Initial clinical experience with a balloon laser ablation catheter for pulmonary vein isolation in patients with atrial fibrillation
    Reddy, VY
    Neuzil, P
    Themistoclakis, S
    Bonso, A
    Rossillo, A
    Raviele, A
    Saliba, W
    Schwelkert, R
    Melsky, J
    Olsen, M
    Ruskin, JN
    Natale, A
    CIRCULATION, 2005, 112 (17) : U551 - U552
  • [8] Early results of the Heliostar™ RF balloon ablation catheter for pulmonary vein isolation
    Breitenstein, A.
    Kanthasamy, V.
    Hofer, D.
    Schilling, R. J.
    Hunter, R. J.
    Finlay, M.
    SWISS MEDICAL WEEKLY, 2022, 152 : 51S - 51S
  • [9] Mapping Technologies for Catheter Ablation of Atrial Fibrillation Beyond Pulmonary Vein Isolation
    La Rosa, Giulio
    Quintanilla, Jorge G.
    Salgado, Ricardo
    Jose Gonzalez-Ferrer, Juan
    Canadas-Godoy, Victoria
    Perez-Villacastin, Julian
    Perez-Castellano, Nicasio
    Jalife, Jose
    Filgueiras-Rama, David
    EUROPEAN CARDIOLOGY REVIEW, 2021, 16
  • [10] Pulmonary vein isolation by laser balloon ablation: is energy delivered too deep into the vein?
    Arceluz, M. R.
    Delgado, R.
    Ortega, M.
    Montes De Oca, R.
    Merino, L.
    EUROPEAN HEART JOURNAL, 2016, 37 : 1084 - 1084