Reassessment of the Access Testosterone chemiluminescence assay and comparison with LC-MS method

被引:18
|
作者
Dittadi, Ruggero [1 ]
Matteucci, Mara [2 ]
Meneghetti, Elisa [1 ]
Ndreu, Rudina [3 ]
机构
[1] ULSS 3 Serenissima, Osped Angelo, UOC Lab Anal, Lab Med Dept, Venice, Italy
[2] Eureka Srl, Lab Div, Chiaravalle, AN, Italy
[3] CNR Clin Physiol Inst, Pisa, Italy
关键词
cross-reaction; DxI platform; immunoassay method; liquid chromatography-mass spectrometry; LoQ; testosterone; IMMUNOASSAYS; MS/MS; SERUM;
D O I
10.1002/jcla.22286
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Aim of the studyTo reassess the imprecision and Limit of Quantitation, to evaluate the cross-reaction with dehydroepiandrosterone-sulfate (DHEAS), the accuracy toward liquid chromatography-mass spectrometry (LC-MS) and the reference interval of the Access Testosterone method, performed by DxI immunoassay platform (Beckman Coulter). Material and MethodsImprecision was evaluated testing six pool samples assayed in 20 different run using two reagents lots. The cross-reaction with DHEAS was studied both by a displacement curve and by spiking DHEAS standard in two serum samples with known amount of testosterone. The comparison with LC-MS was evaluated by Passing-Bablock analysis in 21 routine serum samples and 19 control samples from an External Quality Assurance (EQA) scheme. The reference interval was verified by an indirect estimation on 2445 male and 2838 female outpatients. ResultsThe imprecision study showed a coefficient of variation (CV) between 2.7% and 34.7% for serum pools from 16.3 and 0.27nmol/L. The value of Limit of Quantitation at 20% CV was 0.53nmol/L. The DHEAS showed a cross-reaction of 0.0074%. A comparison with LC-MS showed a trend toward a slight underestimation of immunoassay vs LC-MS (Passing-Bablock equations: DxI=-0.24+0.906 LCMS in serum samples and DxI=-0.299+0.981 LCMS in EQA samples). The verification of reference interval showed a 2.5th-97.5th percentile distribution of 6.6-24.3nmol/L for male over 14years and <0.5-2.78nmol/L for female subjects, in accord with the reference intervals reported by the manufacturer. ConclusionsThe Access Testosterone method could be considered an adequately reliable tool for the testosterone measurement.
引用
收藏
页数:5
相关论文
共 50 条
  • [21] Determination of immunosuppressants by LC-MS/MS: Method comparison and clinical impact
    Dery, M.
    Guerette, D.
    CLINICAL BIOCHEMISTRY, 2012, 45 (13-14) : 1105 - 1105
  • [22] Total testosterone quantitative measurement in serum by LC-MS/MS
    Wang, Yuesong
    Gay, Gabrielle D.
    Botelho, Julianne Cook
    Caudill, Samuel P.
    Vesper, Hubert W.
    CLINICA CHIMICA ACTA, 2014, 436 : 263 - 267
  • [23] Relationship between Serum Testosterone and Fracture Risk in Men: A Comparison of RIA and LC-MS/MS
    Tran, Thach S.
    Center, Jacqueline R.
    Seibel, Markus J.
    Eisman, John A.
    Kushnir, Mark M.
    Rockwood, Alan L.
    Nguyen, Tuan V.
    CLINICAL CHEMISTRY, 2015, 61 (09) : 1182 - 1190
  • [24] Comparison of four clinically validated testosterone LC-MS/MS assays: Harmonization is an attainable goal
    French, Deborah
    Drees, Julia
    Stone, Judith A.
    Holmes, Daniel T.
    van der Gugten, J. Grace
    CLINICAL MASS SPECTROMETRY, 2019, 11 : 12 - 20
  • [25] LC-MS/MS methods for the measurement of androstenedione and testosterone in serum
    Erdogan, E.
    Kushnir, M. M.
    Yue, B.
    Blamires, T. L.
    Roberts, W. L.
    Meikle, A. W.
    Rockwood, A. L.
    CLINICAL CHEMISTRY, 2009, 55 (06) : A184 - A184
  • [26] Comparison of the ADVIA centaur® cyclosporine assay with tandem mass spectrometry (LC-MS/MS)
    Binder, L.
    Streit, F.
    Engelmayer, J.
    Goetze, S.
    Luthe, H.
    Budde, K.
    Beckebaum, S.
    Doll, S.
    Oellerich, M.
    CLINICAL CHEMISTRY, 2007, 53 (06) : A88 - A88
  • [27] Substance p assay in dialysate by LC-MS/MS
    Hache, G.
    Halaoui, R.
    Solgadi, A.
    Gulard, B.
    Chaminade, P.
    Coudore, F.
    Gardier, A.
    FUNDAMENTAL & CLINICAL PHARMACOLOGY, 2009, 23 : 75 - 75
  • [28] Multiplexed LC-MS/MS Assay for Urine Albumin
    Beasley-Green, Ashley
    Burris, Nijah M.
    Bunk, David M.
    Phinney, Karen W.
    JOURNAL OF PROTEOME RESEARCH, 2014, 13 (09) : 3930 - 3939
  • [29] Method development and validation of testosterone in muscle by using LC-MS/MS in positive ESI mode
    Dahiya, Meenakshi
    Dubey, Nidhi
    Singh, Prabha
    Singh, G. N.
    INDIAN JOURNAL OF CHEMISTRY SECTION B-ORGANIC CHEMISTRY INCLUDING MEDICINAL CHEMISTRY, 2013, 52 (04): : 573 - 578
  • [30] Quantification of nivolumab in human plasma by LC-MS/HRMS and LC-MS/MS, comparison with ELISA
    Millet, Aurelien
    Khoudour, Nihel
    Bros, Pauline
    Lebert, Dorothee
    Picard, Guillaume
    Machon, Christelle
    Goldwasser, Francois
    Blanchet, Benoit
    Guitton, Jerome
    TALANTA, 2021, 224