Enhancing performance of a miniaturized surface plasmon resonance sensor in the reflectance detection mode using a waveguide-coupled bimetallic chip

被引:9
|
作者
Lee, Yeon Kyung [1 ]
Jang, Dae Ho [2 ]
Lee, Kyeong-Seok [3 ]
Kim, Won Mok [3 ]
Sohn, Young-Soo [1 ]
机构
[1] Catholic Univ Daegu, Dept Biomed Engn, Gyongsan 712702, Gyeongbuk, South Korea
[2] Sungkyunkwan Univ, Bio Lab, KDmedia Inc, Suwon 440746, South Korea
[3] Korea Inst Sci & Technol, Elect Mat Res Ctr, Seoul 136791, South Korea
来源
基金
新加坡国家研究基金会;
关键词
Waveguide-coupled bimetallic chip; Miniaturized SPR sensor; Intensity measurement; Low molecular weight biomolecule; OPTICAL BIOSENSORS; HIGH-SENSITIVITY; FIXED-ANGLE; SPR SENSORS; FILM; ENHANCEMENT; RESOLUTION;
D O I
10.1186/1556-276X-8-344
中图分类号
TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
The characteristics of a waveguide-coupled bimetallic (WcBiM) chip in a miniaturized surface plasmon resonance (SPR) sensor and its detection capability for a low molecular weight biomolecule were investigated. The configuration of the WcBiM chip was gold (Au)/waveguide (ZnS-SiO2)/silver (Ag). In the intensity measurement mode, the sensitivity could be improved by reducing the full width at half maximum (FWHM) of the reflectance curve. The FWHM of the WcBiM chip is narrower than that of the Au chip, which suggests that the slope of the reflectance curve for the WcBiM chip is steeper. In order to generate enhanced resolution, the reflectance should be monitored at the specific angle where the slope is the steepest in the reflectance curve. For the detection of biotin that is a low molecular weight biomolecule, streptavidin was formed on the SPR sensor chip surface. The response of the SPR to biotin at various concentrations was then acquired. The sensitivities of the WcBiM chip and the Au chip were 0.0052%/(ng/ml) and 0.0021%/(ng/ml), respectively. The limit of detection of the biotin concentration for both the WcBiM and Au chips was calculated. The values were 2.87 ng/ml for the WcBiM chip and 16.63 ng/ml for the Au chip. Enhancement of the sensitivity in the intensity detection mode was achieved using the WcBiM chip compared with the Au chip. Therefore, sufficient sensitivity for the detection of a disease-related biomarker is attainable with the WcBiM chip in the intensity measurement mode using a miniaturized SPR sensor.
引用
收藏
页码:1 / 8
页数:8
相关论文
共 50 条
  • [11] A Waveguide-Coupled Surface Plasmon Resonance Sensor Using an Au-MgF2-Au Structure
    Pengfei Zhang
    Le Liu
    Yonghong He
    Xiaoxia Chen
    Kaijie Ma
    Dong Wei
    Plasmonics, 2019, 14 : 187 - 195
  • [12] Resolution Enhancement in Surface Plasmon Resonance Sensor Based on Waveguide Coupled Mode by Combining a Bimetallic Approach
    Lee, Kyeong-Seok
    Son, Ju Myeong
    Jeong, Dae-Yong
    Lee, Taek Sung
    Kim, Won Mok
    SENSORS, 2010, 10 (12) : 11390 - 11399
  • [13] A Waveguide-Coupled Surface Plasmon Resonance Sensor Using an Au-MgF2-Au Structure
    Zhang, Pengfei
    Liu, Le
    He, Yonghong
    Chen, Xiaoxia
    Ma, Kaijie
    Wei, Dong
    PLASMONICS, 2019, 14 (01) : 187 - 195
  • [14] High-resolution surface plasmon resonance sensor with Fano resonance in waveguide-coupled multilayer structures
    Zheng, Gaige
    Cong, Jiawei
    Xu, Linhua
    Wang, Jicheng
    APPLIED PHYSICS EXPRESS, 2017, 10 (04)
  • [15] Tunable Fano resonances and improved sensitivity in waveguide-coupled surface plasmon resonance sensors with a bimetallic layer
    Liu, Kai
    Yang, Heming
    Mao, Huibing
    Song, Changsheng
    Wang, Jiqing
    JOURNAL OF PHYSICS D-APPLIED PHYSICS, 2019, 52 (41)
  • [16] Fano resonance and plasmon-induced transparency in waveguide-coupled surface plasmon resonance sensors
    Hayashi, Shinji
    Nesterenko, Dmitry V.
    Sekkat, Zouheir
    APPLIED PHYSICS EXPRESS, 2015, 8 (02) : 022201
  • [17] A High Sensitivity Surface Plasmon Resonance D-Shaped Fiber Sensor Based on a Waveguide-Coupled Bimetallic Structure: Modeling and Optimization
    Rezaei, Nasim
    Yahaghi, Alireza
    IEEE SENSORS JOURNAL, 2014, 14 (10) : 3611 - 3615
  • [18] Waveguide-coupled Spectral Surface Plasmon Resonance Sensor for Real-time Monitoring of Biochemical Reaction at Surface
    Chen Fang
    Qi Zhi-Mei
    Xia Shan-Hong
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2009, 37 (04) : 568 - 572
  • [19] MicroRNA Biosensing Using Telecommunication Wavelength-Interrogated Waveguide-Coupled Surface Plasmon Resonance
    Hsu, Shih-Hsiang
    Lin, Jih-Hui
    Tsai, Ding-Zhang
    Tsai, Hong-Sin
    Jian, Zhi-Hao
    Liu, Hsin-Fu
    IEEE SENSORS JOURNAL, 2016, 16 (09) : 2890 - 2891
  • [20] Angular Piecewise Modal Analysis for Waveguide-Coupled Surface Plasmon Resonance Structure
    Bera, Mahua
    Ray, Mina
    JOURNAL OF LIGHTWAVE TECHNOLOGY, 2014, 32 (18) : 3199 - 3205