ACCURACY OF END-TIDAL CO2 CAPNOMETERS IN POST-CARDIAC SURGERY PATIENTS DURING CONTROLLED MECHANICAL VENTILATION

被引:19
|
作者
Heines, Serge J. H. [1 ]
Strauch, Ulrich [1 ]
Roekaerts, Paul M. H. J. [1 ]
Winkens, Bjorn [2 ]
Bergmans, Dennis C. J. J. [1 ]
机构
[1] Maastricht Univ, Dept Intens Care, Med Ctr, NL-6202 AZ Maastricht, Netherlands
[2] Maastricht Univ, Dept Methodol & Stat, NL-6202 AZ Maastricht, Netherlands
来源
JOURNAL OF EMERGENCY MEDICINE | 2013年 / 45卷 / 01期
关键词
end-tidal carbon dioxide; capnography; capnometer; monitoring; carbon dioxide; blood gas analysis; CARBON-DIOXIDE; ARTERIAL; TRAUMA; CAPNOGRAPHY; CARE;
D O I
10.1016/j.jemermed.2012.11.019
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Background: The determination of end-tidal carbon dioxide (etCO(2)) is very helpful in cardiac resuscitation for confirmation and monitoring of endotracheal tube placement and as an indicator of return of circulation and effectiveness of chest compressions. There is now also widespread use of capnometry on-site at emergency and trauma fields. Objective: We studied the accuracy and correlation of three capnometers (EMMA, Medtronic, and Evita) with partial pressure of arterial CO2 (PaCO2) measurements. Methods: The three capnometers were placed in-line in the ventilator tubing of the patient. Forty sedated and mechanically ventilated post-cardiac surgery patients were studied. Twenty consecutive etCO(2) values were collected simultaneously from all three monitors while drawing an arterial blood sample. Paired sample t-test and Pearson correlation were used to compare the capnometers and their correlation with PaCO2. Results: The correlation of etCO(2) measurements between all three capnometers was good (Emma vs. Evita: 0.874, Emma vs. Medtronic: 0.949, Evita vs. Medtronic: 0.878). The correlation of PaCO2 with the Evita is the lowest (0.671) as compared to the EMMA (0.693) and the Medtronic (0.727). The lowest dispersion of the difference between etCO(2) and PaCO2 was seen in EMMA (3.30), the highest in Evita (3.98). Conclusions: A good correlation between etCO(2) and PaCO2 was shown with the three capnometers in the present study. However, etCO(2) measurements were not valid to estimate PaCO2 in these patients. Therefore, capnometry cannot be used to replace serial blood gas analyses completely, but may be a good cardiopulmonary trend monitor and alerting system in catastrophic events. (c) 2013 Elsevier Inc.
引用
收藏
页码:130 / 134
页数:5
相关论文
共 50 条
  • [21] Observations of End-Tidal Co2 and Invasive Cardiac Output Measurements During Mechanical CPR in a Porcine Model of Cardiac Arrest
    Di Maio, Rebecca
    Crawford, Paul
    Hartley, Helen
    Brody, David
    Farley, Leighton
    Anderson, John
    Adgey, Jennifer
    CIRCULATION, 2010, 122 (21)
  • [22] An evaluation of end-tidal CO2 change following alterations in ventilation frequency
    Jensen, M. C.
    Lozano, S.
    Gottlieb, D.
    Guttmann, J.
    Moeller, K.
    4TH EUROPEAN CONFERENCE OF THE INTERNATIONAL FEDERATION FOR MEDICAL AND BIOLOGICAL ENGINEERING, 2009, 22 (1-3): : 1113 - 1116
  • [23] End-Tidal Co2 In Patients With Pulmonary Arterial Hypertension
    Truong, L.
    Lee, I.
    Hugh, A.
    Ganesh, S.
    Balekian, A. A.
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2016, 193
  • [24] End-tidal CO2 after CO2 embolism
    Jose Eduardo Tanus-Santos
    Canadian Journal of Anaesthesia, 1998, 45 : 913 - 914
  • [25] Laparoscopic colon surgery:: unreliability of end-tidal CO2 monitoring
    Klopfenstein, C. E.
    Schiffer, E.
    Pastor, C. M.
    Beaussier, M.
    Francis, K.
    Soravia, C.
    Herrmann, F. R.
    ACTA ANAESTHESIOLOGICA SCANDINAVICA, 2008, 52 (05) : 700 - 707
  • [26] End-tidal CO2 after CO2 embolism
    Tanus-Santos, JE
    CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 1998, 45 (09): : 913 - 914
  • [27] END-TIDAL CO2 MEASUREMENT IN THE DETECTION OF ESOPHAGEAL INTUBATION DURING CARDIAC-ARREST
    SAYAH, AJ
    PEACOCK, WF
    OVERTON, DT
    ANNALS OF EMERGENCY MEDICINE, 1990, 19 (08) : 857 - 860
  • [28] Abnormal end-tidal CO2 waveforms
    Jonathan L. Benumof
    Canadian Journal of Anaesthesia, 2003, 50
  • [29] Nasal end-tidal CO2 monitoring
    Weeks, DB
    ANESTHESIOLOGY, 1998, 88 (02) : 547 - 547
  • [30] Abnormal end-tidal CO2 waveforms
    Benumof, JL
    CANADIAN JOURNAL OF ANAESTHESIA-JOURNAL CANADIEN D ANESTHESIE, 2003, 50 (07): : 754 - 754