Effect of Multileaf Collimator Leaf Position Error Determined by Picket Fence Test on Gamma Index Value in Patient-Specific Quality Assurance of Volumetric-Modulated Arc Therapy Plans

被引:3
|
作者
Ceylan, Cemile [1 ,2 ]
Inal, Serpil Yondem [3 ]
Senol, Elif [3 ]
Yilmaz, Berrin [1 ]
Sahin, Sevim [4 ]
机构
[1] Istanbul Oncol Hosp, Radiat Oncol Dept, Istanbul, Turkey
[2] Yeditepe Univ, Hlth Sci Inst, Istanbul, Turkey
[3] Mem Bahcelievler Hosp, Radiat Oncol Dept, Istanbul, Turkey
[4] Fenerbahce Univ, Med Imaging Dept, Istanbul, Turkey
关键词
mlc error; patient-specific quality assurance; gamma-index evaluation; DYNAMIC IMRT; SLIDING WINDOW; MLC; ACCURACY; SENSITIVITY; PERFORMANCE; DELIVERY; CONSEQUENCES;
D O I
10.7759/cureus.12684
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aim The correlation between the MLC QA (IBA Dosimetry, Germany) results of the picket fence test created with intentional errors and the patient's quality assurance (QA) evaluation was investigated to assess the impact of multileaf collimator (MLC) positioning error on patient QA. Materials and methods The picket fence, including error-free and intentional MLC errors, defined in Bank In, Bank Out, and Bank Both were analyzed using MLC QA. The QA of 15 plans consisting of stereotactic radiosurgery (SRS), stereotactic body radiotherapy (SBRT), and conventionally fractionated volumetric-modulated arc therapy (VMAT) acquired with electronic portal imaging devices (EPID) was evaluated in the presence of error-free and MLC errors. The QA of plans were analyzed with 2%/2 mm and 3%/3 mm criteria. Results The passing rates of the picket fence test were 97%, 92%, 91%, and 87% for error-free and intentional errors. The criterion of 3%/3 mm wasn't able to detect an MLC error for either SRS/SBRT or conventionally fractionated VMAT. The criterion of 2%/2mm was more sensitive to detect MLC error for the conventionally fractionated VMAT than SRS/SBRT. While only two of SBRT plans had <90%, four of conventionally fractionated VMAT plans had a <90% passing rate. Conclusion We found that the systematic MLC positioning errors defined with picket fence have a smaller but measurable impact on SRS/SBRT than the VMAT plan for a conventionally fractionated and relatively complex plan such as head and neck and endometrium cases.
引用
收藏
页数:13
相关论文
共 50 条
  • [41] Machine Log File and Calibration Errors-based Patient-specific Quality Assurance (QA) for Volumetric Modulated Arc Therapy (VMAT)
    Ma, Yangguang
    Mou, Xuanqin
    Beeraka, Narasimha M.
    Guo, Yuexin
    Liu, Junqi
    Dai, Jianrong
    Fan, Ruitai
    CURRENT PHARMACEUTICAL DESIGN, 2023, 29 (34) : 2738 - 2751
  • [42] Correlation analysis between 2D and 3D patient-specific quality assurance for volumetric modulated arc therapy
    Sharma, S.
    Sharma, D.
    Subramani, V.
    Gopishankar, N.
    Bhaskar, S.
    Pathy, S.
    Kumar, P.
    Thulkar, S.
    Chander, S.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2023, 21 (04): : 685 - 691
  • [43] Deep learning for patient-specific quality assurance of volumetric modulated arc therapy: Prediction accuracy and cost-sensitive classification performance
    Li, Guangjun
    Duan, Lian
    Xie, Lizhang
    Hu, Ting
    Wei, Weige
    Bai, Long
    Xiao, Qing
    Liu, Wenjie
    Zhang, Lei
    Bai, Sen
    Yi, Zhang
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2024, 125
  • [44] Statistical process control and process capability analysis for non-normal volumetric modulated arc therapy patient-specific quality assurance processes
    Xiao, Qing
    Bai, Sen
    Li, Guangjun
    Yang, Kaixuan
    Bai, Long
    Li, Zhibin
    Chen, Li
    Xian, Lixun
    Hu, Zhenyao
    Zhong, Renming
    MEDICAL PHYSICS, 2020, 47 (10) : 4694 - 4702
  • [45] Improvement Using Planomics Features on Prediction and Classification of Patient-Specific Quality Assurance Using Head and Neck Volumetric Modulated Arc Therapy Plan
    Li, Bing
    Chen, Junying
    Guo, Wei
    Mao, Ronghu
    Zheng, Xiaoli
    Cheng, Xiuyan
    Cui, Tiantian
    Lou, Zhaoyang
    Wang, Ting
    Li, Dingjie
    Tao, Hongyan
    Lei, Hongchang
    Ge, Hong
    FRONTIERS IN NEUROSCIENCE, 2021, 15
  • [46] Patient-specific quality control for intensity-modulated radiation therapy and volumetric-modulated arc therapy using electronic portal imaging device and two-dimensional ion chamber array
    Kausar, Abu
    Mani, Karthick R.
    Azhari, Hasin A.
    Zakaria, Golam A.
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2019, 18 (01) : 26 - 31
  • [47] Effects of high-resolution measurements between different multi-row detectors on volumetric modulated arc therapy patient-specific quality assurance
    Kunii, Y.
    Tanabe, Y.
    Higashi, A.
    Nakamoto, A.
    Nishioka, K.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2023, 21 (03): : 413 - 420
  • [48] Setting the Tolerance and Action Limit for Patient-specific Quality Assurance of Craniospinal Irradiation Volumetric Modulated Arc Therapy: Based on AAPM TG-218 Report
    Raj, Sathiya
    Saminathan, Sathiyan
    Prakasam, Kiruthika
    JOURNAL OF MEDICAL PHYSICS, 2025, 50 (01) : 155 - 159
  • [49] Preliminary study on monitoring patient-specific volumetric modulated arc therapy quality assurance process with statistical process control methodology on the basis of TG-218 report
    Xiao, Qing
    Li, Guangjun
    Li, Zhibin
    Zhang, Xiangbin
    Dai, Guyu
    Bai, Sen
    Shengwu Yixue Gongchengxue Zazhi/Journal of Biomedical Engineering, 2020, 37 (05): : 842 - 847
  • [50] Comparison of Gafchromic EBT2 and EBT3 for patient-specific quality assurance: Cranial stereotactic radiosurgery using volumetric modulated arc therapy with multiple noncoplanar arcs
    Fiandra, Christian
    Fusella, Marco
    Giglioli, Francesca Romana
    Filippi, Andrea Riccardo
    Mantovani, Cristina
    Ricardi, Umberto
    Ragona, Riccardo
    MEDICAL PHYSICS, 2013, 40 (08)