Stent and leaflet stresses across generations of balloon-expandable transcatheter aortic valves

被引:12
|
作者
Xuan, Yue [1 ,2 ]
Dvir, Danny [3 ]
Wang, Zhongjie [1 ,2 ]
Ye, Jian [4 ]
Guccione, Julius M. [1 ,2 ]
Ge, Liang [1 ,2 ]
Tseng, Elaine E. [1 ,2 ]
机构
[1] Univ Calif San Francisco, Div Cardiothorac Surg, 500 Parnassus Ave,Suite 405W,Box 0118, San Francisco, CA 94143 USA
[2] San Francisco VA Med Ctr, 500 Parnassus Ave,Suite 405W,Box 0118, San Francisco, CA 94143 USA
[3] Univ Washington, Div Cardiol, Seattle, WA 98195 USA
[4] St Pauls Hosp, Dept Surg, Vancouver, BC, Canada
基金
美国国家卫生研究院;
关键词
Transcatheter heart valve; Balloon expandable; Computational simulation; Durability; REPLACEMENT; DURABILITY;
D O I
10.1093/icvts/ivaa037
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVES: Transcatheter aortic valve replacement (TAVR) is established therapy for severe aortic stenosis patients with intermediate-, high- and prohibitive-risk for surgery. A significant challenge when expanding TAVR to low-risk and younger patients is the unknown long-term durability. High leaflet stresses have been associated with surgical bioprosthetic valve degeneration. In this study, we examined the impact of changes in valve design across 3 generations of same-sized TAVR devices on stent and leaflet stresses. METHODS: The 26-mm Edwards SAPIEN, 23, 26 and 29mm SAPIEN XT (XT) and 26mm SAPIEN 3 (S3) (n = 1 each) underwent micro-computed tomography (micro-CT) scanning. Dynamic finite element computational simulations of 23-26mm SAPIEN, 23-29mm XT and 23-29mm S3 were performed with physiological loading and micro-CT or scaled geometries. RESULTS: Peak stresses were concentrated in the commissure area and along the bottom of the suture, representing areas most likely to develop structural valve degeneration across TAVR generations. Latest-generation S3 showed greatest 99th percentile principal stress on commissural leaflets for 26 and 29 mm, and increased stresses over XT for 23 mm. Percentage of higher stress areas within the leaflets steadily increased across generations, 3.8%, 3.9% and 5.7%, respectively, for 26mm SAPIEN, XT and S3 with similar trend for 29-mm valves. CONCLUSIONS: Using computational simulations based on high-fidelity modelling of balloon-expandable TAVRs, our study demonstrated that maximum stress areas existed in similar leaflet locations across SAPIEN generations, while the latest model S3 had the highest magnitude for both 26- and 29-mm valves. S3 also had the largest area of higher stresses than other generations, which would be prone to degeneration. Our study coupled with future long-term clinical outcomes >10 years will provide insight on biomechanics of TAVR degeneration.
引用
收藏
页码:879 / 886
页数:8
相关论文
共 50 条
  • [1] Impact of Balloon-Expandable Transcatheter Aortic Valve Size on Stent and Leaflet Stresses
    Krishnan, Kapil
    Dvir, Danny
    Ye, Jian
    Ge, Liang
    Tseng, Elaine
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2015, 66 (15) : B270 - B270
  • [2] Pathology of balloon-expandable transcatheter aortic valves
    Yahagi, Kazuyuki
    Ladich, Elena
    Kutys, Robert
    Mori, Hiroyoshi
    Svensson, Lars G.
    Mack, Michael J.
    Herrmann, Howard C.
    Smith, Craig R.
    Leon, Martin B.
    Virmani, Renu
    Finn, Aloke V.
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2017, 90 (06) : 1048 - 1057
  • [3] Stent and leaflet stresses in 26-mm, third-generation, balloon-expandable transcatheter aortic valve
    Xuan, Yue
    Dvir, Danny
    Wang, Zhongjie
    Mizoguchi, Trek
    Ye, Jian
    Guccione, Julius M.
    Ge, Liang
    Tseng, Elaine E.
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2019, 157 (02): : 528 - 536
  • [4] Stent and Leaflet Stresses in 29-mm Second-Generation Balloon-Expandable Transcatheter Aortic Valve
    Xuan, Yue
    Krishnan, Kapil
    Ye, Jian
    Dvir, Danny
    Guccione, Julius M.
    Ge, Liang
    Tseng, Elaine E.
    ANNALS OF THORACIC SURGERY, 2017, 104 (03): : 773 - 781
  • [5] Stent and leaflet stresses in a 26-mm first-generation balloon-expandable transcatheter aortic valve
    Xuan, Yue
    Krishnan, Kapil
    Ye, Jian
    Dvir, Danny
    Guccione, Julius M.
    Ge, Liang
    Tseng, Elaine E.
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2017, 153 (05): : 1065 - 1073
  • [6] Association of Prosthesis Deformation of Balloon-Expandable Transcatheter Aortic Valves With Subclinical Leaflet Thrombosis
    Pawar, Shubhadarshini
    Patel, Vivek
    Nagasaka, Takashi
    Basantes, Jair
    Patel, Dhairya
    Chakravarty, Tarun
    Koren, Ofir
    Nakamura, Mamoo
    Cheng, Wen
    Jilaihawi, Hasanian
    Gupta, Aakriti
    Makkar, Raj
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2023, 82 (17) : B236 - B236
  • [7] Early hypo-attenuated leaflet thickening in balloon-expandable transcatheter aortic heart valves
    Pache, Gregor
    Schoechlin, Simon
    Blanke, Philipp
    Dorfs, Stephan
    Jander, Nikolaus
    Arepalli, Chesnal D.
    Gick, Michael
    Buettner, Heinz-Joachim
    Leipsic, Jonathon
    Langer, Mathias
    Neumann, Franz-Josef
    Ruile, Philipp
    EUROPEAN HEART JOURNAL, 2016, 37 (28) : 2263 - 2271
  • [8] Transcatheter Valve-In-Valve Implantation for Failed Balloon-Expandable Transcatheter Aortic Valves
    Toggweiler, Stefan
    Wood, David A.
    Rodes-Cabau, Josep
    Kapadia, Samir
    Willson, Alexander B.
    Ye, Jian
    Cheung, Anson
    Leipsic, Jonathon
    Binder, Ronald K.
    Gurvitch, Ronen
    Freeman, Melanie
    Thompson, Christopher R.
    Svensson, Lars G.
    Dumont, Eric
    Tuzcu, E. Murat
    Webb, John G.
    JACC-CARDIOVASCULAR INTERVENTIONS, 2012, 5 (05) : 571 - 577
  • [9] Prospective Study of TMVR Using Balloon-Expandable Aortic Transcatheter Valves in MAC
    Guerrero, Mayra
    Wang, Dee Dee
    Eleid, Mackram F.
    Pursnani, Amit
    Salinger, Michael
    Russell, Hyde M.
    Kodali, Susheel K.
    George, Isaac
    Bapat, Vinayak N.
    Dangas, George D.
    Tang, Gilbert H. L.
    Inglesis, Ignacio
    Meduri, Christopher U.
    Palacios, Igor
    Reisman, Mark
    Whisenant, Brian K.
    Jermihov, Anastasia
    Kaptzan, Tatiana
    Lewis, Bradley R.
    Tommaso, Carl
    Krause, Philip
    Thaden, Jeremy
    Oh, Jae K.
    Douglas, Pamela S.
    Hahn, Rebecca T.
    Leon, Martin B.
    Rihal, Charanjit S.
    Feldman, Ted
    O'Neill, William W.
    JACC-CARDIOVASCULAR INTERVENTIONS, 2021, 14 (08) : 830 - 845
  • [10] Balloon aortic valvuloplasty is not required for safe and effective transfemoral implantation of balloon-expandable transcatheter aortic valves
    Aggarwal, S. K.
    Delahunty, N.
    Reinthaler, M.
    Cheang, M. H.
    Ako, E.
    Yap, J.
    Ozkor, M.
    Mullen, M. J.
    EUROPEAN HEART JOURNAL, 2015, 36 : 270 - 270