On-Line Measurement of Blood Urea Based in Optical Sensing by Employ in Hemodialysis Treatment

被引:0
|
作者
Adolfo Martinez, Gustavo [1 ]
机构
[1] Mil Hosp Ctr, Mexico City, DF, Mexico
关键词
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The present paper shows the design and development the On-line measure system of the blood urea concentration during dialysis procedures, using non-invasive technical by employ the optical sensing in the visible range. Previous results obtained in Military Hospital Centre in Mexico displayed experimental evidence of a clear correlation between blood urea contents and optical absorbance in the red region obtained through the arterial line in hemodialysis machine. Our hypothesis considers that the cause of this dependence is the volume change of the red cells with the urea concentration due to the osmotic pressure and to the mechanisms of regulation of the cellular volume. The reduction of the volume of the cells would lead to a decrease of the optical absorbance. Measurements were first performed during experimental dialysis procedures, clinical measurements were carried out on a group of patients during hemodialysis procedures in Mexico City. The most relevant interfering parameter was the water extraction along the dialysis treatment, given that changes in the blood fluid and arterial pressure (Hypotention or Hypertention), thus affecting the measurement principle.
引用
收藏
页码:651 / +
页数:3
相关论文
共 50 条
  • [31] On-line measurement of mechanical, optical properties and roughness parameters
    Hartman, H
    Casajus, A
    Richter, U
    REVISTA DE METALURGIA, 2005, : 74 - 82
  • [32] On-line dynamic measurement of blood viscosity, hematocrit and change of blood volume
    尹志勇
    周继红
    蔡绍皙
    王正国
    中华创伤杂志(英文版), 2000, (02) : 38 - 42
  • [33] On-line dynamic measurement of blood viscosity, hematocrit and change of blood volume
    尹志勇
    周继红
    蔡绍皙
    王正国
    ChineseJournalofTraumatology, 2000, (02)
  • [34] On-line measurement of the thickness and optical quality of float glass with a sensor based on a diffractive element
    Räsänen, Jari
    Peiponen, Kai-Erik
    Applied Optics, 2001, 40 (28): : 5034 - 5039
  • [35] On-line measurement of the thickness and optical quality of float glass with a sensor based on a diffractive element
    Räsänen, J
    Peiponen, KE
    APPLIED OPTICS, 2001, 40 (28) : 5034 - 5039
  • [36] On-line near-infrared spectrometer to monitor urea removal in real time during hemodialysis
    Cho, David S.
    Olesberg, Jonathon T.
    Flanigan, Michael J.
    Arnold, Mark A.
    APPLIED SPECTROSCOPY, 2008, 62 (08) : 866 - 872
  • [37] Dialysis dose (Kt/V) and clearance variation sensitivity using measurement of ultraviolet-absorbance (on-line), blood urea, dialysate urea and ionic dialysance
    Uhlin, Fredrik
    Fridolin, Ivo
    Magnusson, Martin
    Lindberg, Lars-Goran
    NEPHROLOGY DIALYSIS TRANSPLANTATION, 2006, 21 (08) : 2225 - 2231
  • [38] Robotic Deburring Based on On-line Burr Measurement
    Liao, L.
    Xi, F.
    Engin, S.
    BURRS - ANALYSIS, CONTROL AND REMOVAL: PROCEEDINGS OF THE CIRP INTERNATIONAL CONFERENCE ON BURRS, 2009, 2010, : 213 - +
  • [39] On-line Monitoring Method of Pole and Tower Tilt Based on Remote Sensing Satellite Optical Image
    Gao X.
    Li L.
    Liu W.
    Zhongguo Dianji Gongcheng Xuebao/Proceedings of the Chinese Society of Electrical Engineering, 2022, 42 (16): : 5971 - 5978
  • [40] Optical on-line measurement of the thickness distribution of blow molding parisons
    Swan, PL
    Kamal, MR
    GarciaRejon, A
    Cielo, P
    POLYMER ENGINEERING AND SCIENCE, 1996, 36 (07): : 985 - 992