A diamond detector in the dosimetry of high-energy electron and photon beams

被引:73
|
作者
Laub, WU [1 ]
Kaulich, TW [1 ]
Nüsslin, F [1 ]
机构
[1] Univ Tubingen, Dept Med Phys, D-72076 Tubingen, Germany
来源
PHYSICS IN MEDICINE AND BIOLOGY | 1999年 / 44卷 / 09期
关键词
D O I
10.1088/0031-9155/44/9/306
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A diamond detector type 60003 (PTW Freiburg) was examined for the purpose of dosimetry with 4-20 MeV electron beams and 4-25 MV photon beams. Results were compared with those obtained by using a Markus chamber for electron beams and an ionization chamber for photon beams. Dose distributions were measured in a water phantom with the detector connected to a Unidos electrometer (PTW Freiburg). After a pre-irradiation of about 5 Gy the diamond detector shows a stability in response which is better than that of an ionization chamber. The current of the diamond detector was measured under variation of photon beam dose rate between 0.1 and 7 Gy min(-1). Different FSDs were chosen. Furthermore the pulse repetition frequency and the depth of the detector were changed. The electron beam dose rate was varied between 0.23 and 4.6 Gy min(-1) by changing the pulse-repetition frequency. The response shows no energy dependence within the covered photon-beam energy range. Between 4 MeV and 18 MeV electron beam energy it shows only a small energy dependence of about 2%, as expected from theory. For smaller electron energies the response increases significantly and an influence of the contact material used for the diamond detector can be surmised. A slight sublinearity of the current and dose rare was found. Detector current and dose rate are related by the expression i proportional to (D) over dot(Delta), where i is the detector current, (D) over dot is the dose rate and Delta is a correction factor of approximately 0.963. Depth-dose curves of photon beams, measured with the diamond detector, show a slight overestimation compared with measurements with the ionization chamber. This overestimation is compensated for by the above correction term. The superior spatial resolution of the diamond detector leads to minor deviations between depth-dose curves of electron beams measured with a Markus chamber and a diamond detector.
引用
收藏
页码:2183 / 2192
页数:10
相关论文
共 50 条
  • [21] An on-site dosimetry audit for high-energy electron beams
    de Prez, Leon
    Heukelom, Stan
    Jansen, Bartel
    Jansen, Wim
    van de Kamer, Jeroen
    van Klink, Wenze
    Kok, Elfried
    Perik, Thijs
    de Pooter, Jacco
    Wittkamper, Frits
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2018, 5 : 44 - 51
  • [22] Relative dosimetry of high-energy electron beams with radiographic film
    Torres, J
    Guerrero, R
    Gonzalez, A
    Angulo, E
    Palacios, M
    Buade, M
    Almansa, J
    Iborra, M
    RADIOTHERAPY AND ONCOLOGY, 2005, 76 : S203 - S203
  • [23] A FORMULATION FOR HIGH-ENERGY PHOTON AND ELECTRON-BEAM DOSIMETRY
    SHIRAGAI, A
    PHYSICS IN MEDICINE AND BIOLOGY, 1991, 36 (05): : 633 - 642
  • [24] The dose rate dependence of synthetic diamond detectors in the relative dosimetry of high-energy electron therapy beams
    Ade, N.
    Nam, T. L.
    Derry, T. E.
    Mhlanga, S. H.
    RADIATION PHYSICS AND CHEMISTRY, 2014, 98 : 155 - 162
  • [25] Dosimetry of high-energy photon beams based on standards of absorbed dose to water
    McDonald, JC
    HEALTH PHYSICS, 2002, 82 (05): : 751 - 752
  • [26] LEAD-POLYSTYRENE TRANSITION ZONE DOSIMETRY IN HIGH-ENERGY PHOTON BEAMS
    MIJNHEER, BJ
    RICE, RK
    CHIN, LM
    RADIOTHERAPY AND ONCOLOGY, 1988, 11 (04) : 379 - 386
  • [27] Liquid scintillator for 2D dosimetry for high-energy photon beams
    Poenisch, Falk
    Archambault, Louis
    Briere, Tina Marie
    Sahoo, Narayan
    Mohan, Radhe
    Beddar, Sam
    Gillin, Michael T.
    MEDICAL PHYSICS, 2009, 36 (05) : 1478 - 1485
  • [28] ABSORBED DOSE BEAM QUALITY FACTORS FOR THE DOSIMETRY OF HIGH-ENERGY PHOTON BEAMS
    ANDREO, P
    PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (12): : 2189 - 2211
  • [29] CORRECTION FACTORS FOR FRICKE DOSIMETRY IN HIGH-ENERGY ELECTRON-BEAMS
    MA, CM
    NAHUM, AE
    PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (03): : 423 - 438
  • [30] AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams
    Almond, PR
    Biggs, PJ
    Coursey, BM
    Hanson, WF
    Huq, MS
    Nath, R
    Rogers, DWO
    MEDICAL PHYSICS, 1999, 26 (09) : 1847 - 1870