A portable blood plasma clot micro-elastometry device based on resonant acoustic spectroscopy

被引:2
|
作者
Krebs, C. R. [1 ]
Li, Ling [2 ]
Wolberg, Alisa S. [3 ]
Oldenburg, Amy L. [1 ,2 ]
机构
[1] Univ N Carolina, Dept Biomed Engn, Chapel Hill, NC 27599 USA
[2] Univ N Carolina, Dept Phys & Astron, Chapel Hill, NC 27599 USA
[3] Univ N Carolina, Dept Pathol & Lab Med, Chapel Hill, NC 27599 USA
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2015年 / 86卷 / 07期
基金
美国国家卫生研究院;
关键词
PLATELET CONTRACTILE-FORCE; MECHANICAL-PROPERTIES; THROMBIN GENERATION; HEPARIN CONCENTRATION; FIBRIN STRUCTURE; ELASTIC-MODULUS; MANAGEMENT;
D O I
10.1063/1.4926543
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Abnormal blood clot stiffness is an important indicator of coagulation disorders arising from a variety of cardiovascular diseases and drug treatments. Here, we present a portable instrument for elastometry of microliter volume blood samples based upon the principle of resonant acoustic spectroscopy, where a sample of well-defined dimensions exhibits a fundamental longitudinal resonance mode proportional to the square root of the Young's modulus. In contrast to commercial thromboelastography, the resonant acoustic method offers improved repeatability and accuracy due to the high signal-to-noise ratio of the resonant vibration. We review the measurement principles and the design of a magnetically actuated microbead force transducer applying between 23 pN and 6.7 nN, providing a wide dynamic range of elastic moduli (3 Pa-27 kPa) appropriate for measurement of clot elastic modulus (CEM). An automated and portable device, the CEMport, is introduced and implemented using a 2 nm resolution displacement sensor with demonstrated accuracy and precision of 3% and 2%, respectively, of CEM in biogels. Importantly, the small strains (<0.13%) and low strain rates (<1/s) employed by the CEMport maintain a linear stress-to-strain relationship which provides a perturbative measurement of the Young's modulus. Measurements of blood plasma CEM versus heparin concentration show that CEMport is sensitive to heparin levels below 0.050 U/ml, which suggests future applications in sensing heparin levels of post-surgical cardiopulmonary bypass patients. The portability, high accuracy, and high precision of this device enable new clinical and animal studies for associating CEM with blood coagulation disorders, potentially leading to improved diagnostics and therapeutic monitoring. (C) 2015 AIP Publishing LLC.
引用
收藏
页数:11
相关论文
共 50 条
  • [41] Design of a Pendulum Based Micro-Thrust Measuring Device Application to Plasma Microthrusters
    Rustom, Alexios
    Al Raii, Khalil
    Metni, Najib
    Antar, Ghassan
    Habchi, Charbel
    2019 FOURTH INTERNATIONAL CONFERENCE ON ADVANCES IN COMPUTATIONAL TOOLS FOR ENGINEERING APPLICATIONS (ACTEA), 2019,
  • [42] Evaluation of biological effects of visual stimulation based on a portable measuring device for blood pressure and heart rate
    Sugita, N
    Yoshizawa, M
    Tanaka, A
    Masuda, T
    Abel, K
    Yambe, T
    Nitta, S
    SICE 2003 ANNUAL CONFERENCE, VOLS 1-3, 2003, : 2030 - 2033
  • [43] ppb-Level detection of benzene diluted in water with portable device based on bubbling extraction and UV spectroscopy
    Camou, S.
    Shimizu, A.
    Horiuchi, T.
    Haga, T.
    SENSORS AND ACTUATORS B-CHEMICAL, 2008, 132 (02): : 601 - 607
  • [44] A single-use cartridge based on an acoustic sensor for a portable device for rapid assessment of the functional state of the hemostasis system
    Zhukov, E. L.
    Slizevich, D. S.
    Gubarev, F. A.
    Nagornyi, O. A.
    Zyryanov, S. V.
    BIOMEDICAL ENGINEERING, 2024, 58 (02) : 102 - 105
  • [45] Portable, Constriction-Expansion Blood Plasma Separation and Polymerization-Based Malaria Detection
    Shatova, Tatyana A.
    Lathwal, Shefali
    Engle, Marissa R.
    Sikes, Hadley D.
    Jensen, Klays F.
    ANALYTICAL CHEMISTRY, 2016, 88 (15) : 7627 - 7632
  • [46] A cost-effective voice coil motor-based portable micro-indentation device for in situ testing
    Peng, Guangjian
    Xu, Fenglei
    Chen, Jianfeng
    Hu, Yahao
    Wang, Huadong
    Zhang, Taihua
    MEASUREMENT, 2020, 165 (165)
  • [47] Photo-acoustic spectroscopy revealing resonant absorption of self-assembled GaAs-based nanowires
    Leahu, Grigore
    Petronijevic, Emilija
    Belardini, Alessandro
    Centini, Marco
    Voti, Roberto Li
    Hakkarainen, Teemu
    Koivusalo, Eero
    Guina, Mircea
    Sibilia, Concita
    SCIENTIFIC REPORTS, 2017, 7
  • [48] Photo-acoustic spectroscopy revealing resonant absorption of self-assembled GaAs-based nanowires
    Grigore Leahu
    Emilija Petronijevic
    Alessandro Belardini
    Marco Centini
    Roberto Li Voti
    Teemu Hakkarainen
    Eero Koivusalo
    Mircea Guina
    Concita Sibilia
    Scientific Reports, 7
  • [49] Development of NIR Spectroscopy Based Non-invasive Blood Glucose Measuring Device
    Asekar, Megha Shyam
    Sonawane, D. N.
    2018 4TH INTERNATIONAL CONFERENCE FOR CONVERGENCE IN TECHNOLOGY (I2CT), 2018,
  • [50] A continuous flow micro filtration device for plasma/blood separation using submicron vertical pillar gap structures
    Kang, Tae Goo
    Yoon, Yong-Jin
    Ji, Hongmiao
    Lim, Pei Yi
    Chen, Yu
    JOURNAL OF MICROMECHANICS AND MICROENGINEERING, 2014, 24 (08)