A portable blood plasma clot micro-elastometry device based on resonant acoustic spectroscopy

被引:2
|
作者
Krebs, C. R. [1 ]
Li, Ling [2 ]
Wolberg, Alisa S. [3 ]
Oldenburg, Amy L. [1 ,2 ]
机构
[1] Univ N Carolina, Dept Biomed Engn, Chapel Hill, NC 27599 USA
[2] Univ N Carolina, Dept Phys & Astron, Chapel Hill, NC 27599 USA
[3] Univ N Carolina, Dept Pathol & Lab Med, Chapel Hill, NC 27599 USA
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2015年 / 86卷 / 07期
基金
美国国家卫生研究院;
关键词
PLATELET CONTRACTILE-FORCE; MECHANICAL-PROPERTIES; THROMBIN GENERATION; HEPARIN CONCENTRATION; FIBRIN STRUCTURE; ELASTIC-MODULUS; MANAGEMENT;
D O I
10.1063/1.4926543
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Abnormal blood clot stiffness is an important indicator of coagulation disorders arising from a variety of cardiovascular diseases and drug treatments. Here, we present a portable instrument for elastometry of microliter volume blood samples based upon the principle of resonant acoustic spectroscopy, where a sample of well-defined dimensions exhibits a fundamental longitudinal resonance mode proportional to the square root of the Young's modulus. In contrast to commercial thromboelastography, the resonant acoustic method offers improved repeatability and accuracy due to the high signal-to-noise ratio of the resonant vibration. We review the measurement principles and the design of a magnetically actuated microbead force transducer applying between 23 pN and 6.7 nN, providing a wide dynamic range of elastic moduli (3 Pa-27 kPa) appropriate for measurement of clot elastic modulus (CEM). An automated and portable device, the CEMport, is introduced and implemented using a 2 nm resolution displacement sensor with demonstrated accuracy and precision of 3% and 2%, respectively, of CEM in biogels. Importantly, the small strains (<0.13%) and low strain rates (<1/s) employed by the CEMport maintain a linear stress-to-strain relationship which provides a perturbative measurement of the Young's modulus. Measurements of blood plasma CEM versus heparin concentration show that CEMport is sensitive to heparin levels below 0.050 U/ml, which suggests future applications in sensing heparin levels of post-surgical cardiopulmonary bypass patients. The portability, high accuracy, and high precision of this device enable new clinical and animal studies for associating CEM with blood coagulation disorders, potentially leading to improved diagnostics and therapeutic monitoring. (C) 2015 AIP Publishing LLC.
引用
收藏
页数:11
相关论文
共 50 条
  • [21] Resonant Absorption in GaAs-Based Nanowires by Means of Photo-Acoustic Spectroscopy
    Petronijevic, E.
    Leahu, G.
    Belardini, A.
    Centini, M.
    Voti, R. Li
    Hakkarainen, T.
    Koivusalo, E.
    Guina, M.
    Sibilia, C.
    INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2018, 39 (03)
  • [22] EVALUATION OF A PLASMA SEPARATOR MICRO BLOOD COLLECTION DEVICE CONTAINING LITHIUM HEPARIN AS ANTICOAGULANT
    HAMMOND, KB
    HARE, LA
    SCHLINDWEIN, P
    WISEMAN, J
    INCE, G
    NARAYANAN, S
    CLINICAL CHEMISTRY, 1983, 29 (06) : 1220 - 1220
  • [23] A portable microfluidic paper-based analytical device for blood detection and typing assay
    Ansari, Niha
    Trambadiya, Nehal
    Lodha, Anand
    Menon, S. K.
    AUSTRALIAN JOURNAL OF FORENSIC SCIENCES, 2021, 53 (04) : 407 - 418
  • [24] Rapid isolation and detection of erythropoietin in blood plasma by magnetic core gold nanoparticles and portable Raman spectroscopy
    Agoston, Roland
    Izake, Emad L.
    Sivanesan, Arumugam
    Lott, William B.
    Sillence, Martin
    Steel, Rohan
    NANOMEDICINE-NANOTECHNOLOGY BIOLOGY AND MEDICINE, 2016, 12 (03) : 633 - 641
  • [25] Micro-electromechanical film bulk acoustic sensor for plasma and whole blood coagulation monitoring
    Chen, Da
    Song, Shuren
    Ma, Jilong
    Zhang, Zhen
    Wang, Peng
    Liu, Weihui
    Guo, Qiuquan
    BIOSENSORS & BIOELECTRONICS, 2017, 91 : 465 - 471
  • [26] Validation study toward measuring the mechanical properties of blood clots using resonant acoustic spectroscopy with optical vibrometry
    Wu, Gongting
    Wolberg, Alisa S.
    Oldenburg, Amy L.
    ADVANCED BIOMEDICAL AND CLINICAL DIAGNOSTIC SYSTEMS X, 2012, 8214
  • [27] Research on qualitative identifying device of portable sulfur smoked Chinese medicine based on micro fiber spectrometer
    Wang Yongqing
    Liu Yufei
    Chen Wenjun
    Wang Shuonan
    Liu Jing
    Li Duo
    PROCEEDINGS OF THE FIFTH INTERNATIONAL SYMPOSIUM ON TEST AUTOMATION & INSTRUMENTATION, VOLS 1 AND 2, 2014, : 215 - 220
  • [28] Research on Gas Sensing System Using Photoacoustic Spectroscopy Based on a Hemispherical Acoustic Resonant Cavity
    Wang, Gao-Xuan
    Ding, Zhi-Hua
    Gao, Xiao-Ming
    Guang Pu Xue Yu Guang Pu Fen Xi/Spectroscopy and Spectral Analysis, 2025, 45 (03): : 608 - 615
  • [29] MEASUREMENT OF RESONANT AND METASTABLE DENSITIES IN A LOW PRESSURE, MICROWAVE DRIVEN, MICRO ARGON PLASMA BY OPTICAL EMISSION SPECTROSCOPY
    Hermanns, Patrick
    Hillebrand, Bastian
    Fiebrandt, Marcel
    Awakowicz, Peter
    2017 IEEE INTERNATIONAL CONFERENCE ON PLASMA SCIENCE (ICOPS), 2017,
  • [30] DEVELOPMENT OF PORTABLE DEVICE BASED ON COLD ATMOSPHERIC PLASMA FOR SKIN DISINFECTION AND TREATMENT OF CHRONIC WOUNDS
    Pavlichenko, R. O.
    Shypilov, S. A.
    Klosovskij, A. V.
    Vorontsov, P. M.
    Gnidenko, Yu. P.
    Taran, A. V.
    Misiruk, I. O.
    Gubarev, S. P.
    Opaleva, G. P.
    Zolototrubova, M. I.
    Dushyk, L. M.
    Cherkova, N. V.
    Chechel'nitskj, O. G.
    PROBLEMS OF ATOMIC SCIENCE AND TECHNOLOGY, 2025, (01): : 109 - 112