Bioimpedance Measurement of Knee Injuries Using Bipolar Electrode Configuration

被引:6
|
作者
Ye, Xuanjie [1 ]
Wu, Lexi [2 ]
Mao, Kaining [1 ]
Feng, Yiwei [1 ]
Li, Jiajun [1 ]
Ning, Lei [3 ]
Chen, Jie [1 ]
机构
[1] Univ Alberta, Dept Elect & Comp Engn, Edmonton, AB T6G 2R3, Canada
[2] Zhejiang Univ, Sch Med, Hangzhou 310030, Zhejiang, Peoples R China
[3] Zhejiang Univ, Sir Run Run Shaw Hosp, Dept Orthopaed Surg, Med Coll, Hangzhou 310058, Zhejiang, Peoples R China
关键词
Bioimpedance; Electrodes; Injuries; Knee; Current measurement; Biomedical measurement; Impedance; bipolar electrode configuration; knee injuries; machine learning; portable device; BIOELECTRICAL-IMPEDANCE; ELECTRICAL BIOIMPEDANCE; DRY ELECTRODES; EDEMA; BODY; INFLAMMATION; LYMPHEDEMA; MANAGEMENT; DIAGNOSIS; SEVERITY;
D O I
10.1109/TBCAS.2022.3200355
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Currently, there is no suitable solution for the point-of-care diagnosis of knee injuries. A potential portable and low-cost technique for accessing and monitoring knee injuries is bioimpedance measurement. This study validated the feasibility of the bipolar electrode configuration for knee bioimpedance measurement with two electrodes placed on a fixed pair of knee acupuncture locations called Xiyan. Then, the study collected 76 valid samples to investigate the relationship between bioimpedance and knee injuries, among whom 39 patients have unilateral knee injuries, and 37 individuals have healthy knees. The self-contrast results indicated that knee injuries caused a reduction of bioimpedance of the knee by about 5% on average, which was detectable at around 100 kHz (p approximate to 0.001). Furthermore, the results analyzed by principal component analysis and support vector machines show that the detection sensitivity can reach 87.18% using the leave-one-out cross-validation. We also proposed a low-cost and portable bioimpedance measurement device that meets the needs for measuring knee joint bioimpedance.
引用
收藏
页码:962 / 971
页数:10
相关论文
共 50 条
  • [1] A Robust Electrode Configuration for Bioimpedance Measurement of Respiration
    Wang, Hong-bin
    Yen, Chen-wen
    Liang, Jing-tao
    Wang, Qian
    Liu, Guan-zheng
    Song, Rong
    JOURNAL OF HEALTHCARE ENGINEERING, 2014, 5 (03) : 313 - 327
  • [2] Alternate Electrode Positions for the Measurement of Hand Volumes Using Bioimpedance Spectroscopy
    Edwick, Dale O.
    Hince, Dana A.
    Rawlins, Jeremy M.
    Wood, Fiona M.
    Edgar, Dale W.
    LYMPHATIC RESEARCH AND BIOLOGY, 2020, 18 (06) : 560 - 571
  • [3] Four electrode embedded bioimpedance measurement system
    Margo, Cedric
    Katrib, Juliano
    Nadi, Mustapha
    Rouane, Amar
    2013 IEEE FAIBLE TENSION FAIBLE CONSOMMATION (FTFC), 2013,
  • [4] BIOIMPEDANCE ACTIVE ELECTRODE FOR IN-VIVO MEASUREMENT
    RECORD, PM
    ROSELL, X
    RIGAUD, B
    RIU, PJCI
    MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 1994, 32 (06) : 683 - 685
  • [5] Research on Electrode Impedance Characteristic in Electrical Bioimpedance Measurement
    Wang Yan
    Gong WeiYan
    Zhao Shu
    Sha Hong
    Ren ChaoShi
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 2 - DIAGNOSTIC IMAGING, 2009, 25 : 507 - 509
  • [6] Modelling Approach for Assessment of Electrode Configuration and Placement in Bioimpedance Measurements of Skin Irritation
    Koppa, S.
    Savolainen, V.
    Hyttinen, J.
    5TH EUROPEAN CONFERENCE OF THE INTERNATIONAL FEDERATION FOR MEDICAL AND BIOLOGICAL ENGINEERING, PTS 1 AND 2, 2012, 37 : 1238 - 1241
  • [7] A BIPOLAR ELECTRODE FOR KIDNEY PULSE MEASUREMENT
    KATZ, YJ
    JOURNAL OF UROLOGY, 1962, 88 (05): : 713 - &
  • [8] Closed bipolar electrochemistry in a four-electrode configuration
    Gamero-Quijano, Alonso
    Molina-Osorio, Andres F.
    Peljo, Pekka
    Scanlon, Micheal D.
    PHYSICAL CHEMISTRY CHEMICAL PHYSICS, 2019, 21 (19) : 9627 - 9640
  • [9] Electrode errors in bioimpedance measurement systems for long-term applications
    Lozano, A
    PROCEEDINGS OF THE 1997 16TH SOUTHERN BIOMEDICAL ENGINEERING CONFERENCE, 1997, : 3 - 6
  • [10] Electrode Placement Strategies for the Measurement of Radial Artery Bioimpedance: Simulations and Experiments
    Pesti, K.
    Metshein, M.
    Annus, P.
    Koiv, H.
    Min, M.