Dosimetric effects of rotational offsets in stereotactic body radiation therapy (SBRT) for lung cancer

被引:13
|
作者
Yang, Yun [1 ]
Catalano, Suzanne [1 ]
Kelsey, Chris R. [1 ]
Yoo, David S. [1 ]
Yin, Fang-Fang [1 ]
Cai, Jing [1 ]
机构
[1] Duke Univ, Med Ctr, Dept Radiat Oncol, Durham, NC 27710 USA
关键词
Lung SBRT; Rotation; 6D correction; IGRT; RADIOTHERAPY; ERRORS; MARGINS; PATIENT; SETUP;
D O I
10.1016/j.meddos.2013.11.002
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
To quantitatively evaluate dosimetric effects of rotational offsets in stereotactic body radiation therapy (SBRT) for lung cancer. Overall, 11 lung SBRT patients (8 female and 3 male; mean age: 75.0 years) with medially located tumors were included. Treatment plans with simulated rotational offsets of 1 degrees, 3 degrees, and 5 degrees in roll, yaw, and pitch were generated and compared with the original plans. Both clockwise and counterclockwise rotations were investigated. The following dosimetric metrics were quantitatively evaluated: planning target volume coverage (PTV V-100%), max PTV dose (PTV D-max), percentage prescription dose to 0.35 cc of cord (cord D-0.35 cc), percentage prescription dose to 0.35 cc and 5 cc of esophagus (esophagus D-0.35 cc and D-5 cc), and volume of the lungs receiving at least 20 Gy (lung V-20). Statistical significance was tested using Wilcoxon signed rank test at the significance level of 0.05. Overall, small differences were found in all dosimetric matrices at all rotational offsets: 95.6% of differences were < 1% or < 1 Gy. Of all rotational offsets, largest change in PTV V-100%, PTV D-max, cord D-0.35 cc, esophagus D-0.35 cc, esophagus D-5 cc, and lung V-20 was - 8.36%, - 6.06%, 11.96%, 8.66%, 6.02%, and - 0.69%, respectively. No significant correlation was found between any dosimetric change and tumor-to-cord/esophagus distances (R-2 range: 0 to 0.44). Larger dosimetric changes and intersubject variations were observed at larger rotational offsets. Small dosimetric differences were found owing to rotational offsets up to 5 degrees in lung SBRT for medially located tumors. Larger intersubject variations were observed at larger rotational offsets. (C) 2014 American Association of Medical Dosimetrists.
引用
收藏
页码:117 / 121
页数:5
相关论文
共 50 条
  • [1] Stereotactic body radiation therapy (SBRT) and respiratory gating in lung cancer: dosimetric and radiobiological considerations
    Herman, Tania De La Fuente
    Vlachaki, Maria T.
    Herman, Terence S.
    Hibbitts, Kerry
    Stoner, Julie A.
    Ahmad, Salahuddin
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2010, 11 (01): : 158 - 169
  • [2] Dosimetric Impact of Rotational Setup Error in Stereotactic Body Frame Radiation Therapy (SBRT)
    Cao, M.
    Cardenes, H.
    Fakiris, A.
    Desrosiers, C.
    Slessinger, E.
    Langer, M.
    Das, I.
    MEDICAL PHYSICS, 2009, 36 (06) : 2815 - +
  • [3] Dosimetric Evaluation of Lung Stereotactic Body Radiation Therapy (SBRT) Treatment Plans
    Giaddui, T.
    Wang, B.
    Hollander, A.
    Micaily, B.
    Miyamoto, C.
    Li, S.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [4] A Dosimetric Comparison of Two Treatment Setups for Lung Stereotactic Body Radiation Therapy (SBRT) Patients
    Lewis, D.
    Wang, X.
    Gao, S.
    Nitsch, P.
    Pollard, J.
    Sadagopan, R.
    Balter, P.
    MEDICAL PHYSICS, 2015, 42 (06) : 3336 - 3336
  • [5] A review on the technical and dosimetric aspects of stereotactic body radiation therapy (SBRT)
    Saw, Cheng B.
    Bao, Shanglian
    Li, Sicong
    JOURNAL OF RADIATION ONCOLOGY, 2012, 1 (04) : 317 - 322
  • [6] A study for the dosimetric evaluation of rotational setup error for lung stereotactic body radiation therapy
    Lee, Minsik
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2018, 17 (04) : 422 - 426
  • [7] Stereotactic body radiation therapy (SBRT) for lung metastases
    Okunieff, Paul
    Petersen, Anncatrine L.
    Philip, Abraham
    Milano, Michael T.
    Katz, Alan W.
    Boros, Laszlo
    Schell, Michael C.
    ACTA ONCOLOGICA, 2006, 45 (07) : 808 - 817
  • [8] Stereotactic body radiation therapy (SBRT) for lung oligometastasis
    Larrea, L.
    Lopez-Munoz, E.
    Antonini, P.
    Bea, J.
    Banos, M. C.
    Garcia, M. A.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S50 - S51
  • [9] Stereotactic body radiation therapy (SBRT) for lung oligometastasis
    Van den Begin, R.
    Engels, B.
    Gevaert, T.
    Storme, G.
    De Ridder, M.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S50 - S50
  • [10] Stereotactic Body Radiation Therapy (SBRT) for recurrent lung cancer following prior radiation
    Wurzer, J.