Biomechanical evaluation of stand-alone lumbar polyether-ether-ketone interbody cage with integrated screws

被引:38
|
作者
Kornblum, Martin B. [1 ]
Turner, Alexander W. L. [2 ]
Cornwall, G. Bryan [2 ]
Zatushevsky, Michael A. [2 ]
Phillips, Frank M. [3 ]
机构
[1] Mendelson Kornblum Orthopaed & Spine Surg, Warren, MI 48093 USA
[2] NuVasive Inc, San Diego, CA 92121 USA
[3] Rush Univ, Med Ctr, Dept Orthoped Surg, Midwest Orthopaed, Chicago, IL 60612 USA
来源
SPINE JOURNAL | 2013年 / 13卷 / 01期
关键词
Biomechanics; Lumbar; ALIF; Interbody fusion; Fixation; PEDICLE SCREWS; VASCULAR INJURY; IN-VITRO; INVITED SUBMISSION; PERIPHERAL-NERVES; SPINAL-FUSION; MARCH; 2005; FIXATION; PLATE; STABILITY;
D O I
10.1016/j.spinee.2012.11.013
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
BACKGROUND CONTEXT: Stand-alone interbody cages with integrated screws potentially provide a biomechanically stable solution for anterior lumbar interbody fusion (ALIF) that alleviates the need for additional exposure for supplemental fixation, thereby reducing the chance of additional complications and morbidity. PURPOSE: To compare the stability of a stand-alone anterior interbody fusion system with integrated fixation screws against traditional supplemental fixation methods and to evaluate the difference between three and four fixation screws in the stand-alone cage. STUDY DESIGN: In vitro cadaveric biomechanical study. METHODS: Eight cadaveric lumbar spines (L2-sacrum) were tested using a flexibility protocol consisting of three cycles to +/- 7.5 Nm in flexion-extension, lateral bending, and axial rotation. The conditions evaluated were intact spine; polyether-ether-ketone cage (zero integrated screws) at L4-L5; cage (zero screws) + bilateral pedicle screws (PS); cage (three screws); cage (four screws); cage (zero screws) + anterior plate; and cage (three screws) + spinous process plate. Motion at the index level was assessed using an optoelectronic system. RESULTS: The cage without integrated screws reduced the motion in flexion-extension and lateral bending (p < .001) compared with that in the intact spine. In axial rotation, mean range of motion (ROM) was 8% greater than in intact spine (p > .962). The addition of three integrated screws reduced ROM significantly compared with the cage without screws in all motion planes (p < .001). A fourth screw had no statistically significant effect on the ROM, although there was a trend toward less motion with four screws compared with three. In flexion-extension, the cage with three integrated screws and the spinous process plate was the most rigid condition. There was no significant difference from the bilateral PS (p = .537); however, this was more rigid than all other conditions (p < .024). The most stable condition in lateral bending and axial rotation was the cage with bilateral PS. In lateral bending, the cage (three or four screws) was not significantly different from the cage with anterior plate or the cage (three screws) with spinous process plate fixation; however, only the latter condition was statistically comparable with bilateral PS. In axial rotation, there were no significant differences between the conditions that included integrated screws or supplemental fixation (p > .081). CONCLUSIONS: Biomechanical testing revealed that the stand-alone cage with integrated screws provides more immediate stability than a cage alone and provides equivalent stability to ALIF constructs with supplemental fixation in lateral bending and axial rotation. Additional flexion-extension rigidity of the anterior cage maybe realized by the addition of a spinous process plate that was found to be as stable as supplemental bilateral PS. (C) 2013 Elsevier Inc. All rights reserved.
引用
收藏
页码:77 / 84
页数:8
相关论文
共 50 条
  • [1] Failure of a polyether-ether-ketone expandable interbody cage following transforaminal lumbar interbody fusion
    Stein, Isaac C.
    Than, Khoi D.
    Chen, Kevin S.
    Wang, Anthony C.
    Park, Paul
    EUROPEAN SPINE JOURNAL, 2015, 24 : S555 - S559
  • [2] Failure of a polyether-ether-ketone expandable interbody cage following transforaminal lumbar interbody fusion
    Isaac C. Stein
    Khoi D. Than
    Kevin S. Chen
    Anthony C. Wang
    Paul Park
    European Spine Journal, 2015, 24 : 555 - 559
  • [3] Biomechanical comparison of anterior lumbar interbody fusion: stand-alone interbody cage versus interbody cage with pedicle screw fixation - a finite element analysis
    Kyung-Chul Choi
    Kyeong-Sik Ryu
    Sang-Ho Lee
    Yeong Hyeon Kim
    Sung Jae Lee
    Chun-Kun Park
    BMC Musculoskeletal Disorders, 14
  • [4] Biomechanical comparison of anterior lumbar interbody fusion: stand-alone interbody cage versus interbody cage with pedicle screw fixation - a finite element analysis
    Choi, Kyung-Chul
    Ryu, Kyeong-Sik
    Lee, Sang-Ho
    Kim, Yeong Hyeon
    Lee, Sung Jae
    Park, Chun-Kun
    BMC MUSCULOSKELETAL DISORDERS, 2013, 14
  • [5] Evaluation of a Lumbar Intervertebral Spacer With Integrated Screws as a Stand-alone Fixation Device
    Beaubien, Brian P.
    Freeman, Andrew L.
    Turner, Joseph L.
    Castro, Carlos A.
    Armstrong, Will D.
    Waugh, Lindsey G.
    Dryer, Randall F.
    JOURNAL OF SPINAL DISORDERS & TECHNIQUES, 2010, 23 (05): : 351 - 358
  • [6] Biomechanical Comparison of Multilevel Stand-Alone Lumbar Lateral Interbody Fusion With Posterior Pedicle Screws: An In Vitro Study
    Mok, James M.
    Lin, Ye
    Tafur, Julio Castillo
    Diaz, Roberto Leonardo
    Amirouche, Farid
    NEUROSPINE, 2023, 20 (02) : 478 - 486
  • [7] Impact of cage position on biomechanical performance of stand-alone lateral lumbar interbody fusion: a finite element analysis
    Nan, Chong
    Ma, Zhanbei
    Liu, Yuxiu
    Ma, Liang
    Li, Jiaqi
    Zhang, Wei
    BMC MUSCULOSKELETAL DISORDERS, 2022, 23 (01)
  • [8] Impact of cage position on biomechanical performance of stand-alone lateral lumbar interbody fusion: a finite element analysis
    Chong Nan
    Zhanbei Ma
    Yuxiu Liu
    Liang Ma
    Jiaqi Li
    Wei Zhang
    BMC Musculoskeletal Disorders, 23
  • [9] Stand-alone lumbar cage subsidence: A biomechanical sensitivity study of cage design and placement
    Calvo-Echenique, Andrea
    Cegonino, Jose
    Chueca, Raul
    Perez-del Palomar, Amaya
    COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2018, 162 : 211 - 219
  • [10] Biomechanical stability of five stand-alone anterior lumbar interbody fusion constructs
    Tsantrizos, A
    Andreou, A
    Aebi, M
    Steffen, T
    EUROPEAN SPINE JOURNAL, 2000, 9 (01) : 14 - 22