Corneal Power Measurement Obtained by Fourier-Domain Optical Coherence Tomography: Repeatability, Reproducibility, and Comparison With Scheimpflug and Automated Keratometry Measurements

被引:22
|
作者
Wang, Qinmei [1 ,2 ,3 ]
Hua, Yanjun [1 ,2 ,3 ]
Savini, Giacomo [4 ]
Chen, Hao [1 ,2 ,3 ]
Bao, Fangjun [1 ,2 ]
Lin, Shishi [1 ,2 ]
Lu, Weicong [1 ,2 ]
Huang, Jinhai [1 ,2 ,3 ]
机构
[1] Wenzhou Med Univ, Sch Ophthalmol & Optometry, Wenzhou 325027, Zhejiang, Peoples R China
[2] Wenzhou Med Univ, Hosp Eye, Wenzhou 325027, Zhejiang, Peoples R China
[3] Minist Hlth PR China, Key Lab Vis Sci, Wenzhou, Zhejiang, Peoples R China
[4] GB Bietti Fdn IRCCS, Rome, Italy
关键词
corneal power; Scheimpflug; optical coherence tomography; precision; agreement; INTRAOCULAR-LENS POWER; LASER REFRACTIVE SURGERY; STATISTICAL-METHODS; PENTACAM; AGREEMENT; CAMERA; SYSTEM;
D O I
10.1097/ICO.0000000000000564
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose:To assess the repeatability and reproducibility of corneal power values obtained by a Fourier-domain optical coherence tomography (FD-OCT) system (RTVue) and to compare them with the values obtained by a Scheimpflug camera system (Pentacam HR) and by automated keratometry (IOL Master).Methods:Thirty-two eyes from 32 healthy subjects were included in this prospective study. Two experienced observers measured each eye 3 consecutive times with the Pentacam, IOLMaster, and RTVue centered on either the pupil or corneal vertex. The conventional keratometry equivalent (CKE) and anterior (K-a), posterior (K-p), and net (K-n) corneal power values were determined.Results:The corneal power values obtained by the RTVue showed high repeatability (all intraclass correlation coefficient >0.96) and reproducibility (coefficient of variation <1.0%). Pupil-centered FD-OCT performed slightly better than corneal vertex-centered FD-OCT. Mean corneal values had higher reproducibly than any of the individual values. CKE, K-a, K-p, and K-n obtained by FD-OCT were 0.62 to 0.68 diopters (D), 0.70 to 0.76 D, 0.11 to 0.13 D, and 0.93 to 0.94 D higher than those obtained by the Pentacam HR, respectively. CKE and K-a obtained with the RTVue were also 0.60 to 0.74 D higher than those obtained with the IOLMaster, respectively.Conclusions:The corneal power measurements obtained by the RTVue FD-OCT system showed high repeatability and reproducibility. Measurements obtained by FD-OCT with pupil centration were more reproducible than those obtained by FD-OCT with corneal vertex centration. We recommend that pupil-centered FD-OCT be used in clinical applications. Neither RTVue versus Pentacam HR nor RTVue versus IOLMaster can be used interchangeably.
引用
收藏
页码:1266 / 1271
页数:6
相关论文
共 50 条
  • [21] Assessment of corneal thickness measurement using swept-source Fourier-domain anterior segment optical coherence tomography and Scheimpflug camera
    Huang, Jinhai
    Feng, Yifan
    Wang, Qinmei
    Pesudovs, Konrad
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (07): : 1305 - 1306
  • [22] Corneal epithelial thickness measured by Fourier-domain optical coherence tomography
    Wang, Ting
    Li, Fengjie
    Wang, Shuting
    Shi, Weiyun
    SPEKTRUM DER AUGENHEILKUNDE, 2016, 30 (01) : 3 - 9
  • [23] Diagnosis of keratoconus progression by Fourier-domain optical coherence tomography should be correlated with Scheimpflug tomography
    Agrawal, Swati
    Khurana, Ashi
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2019, 45 (10): : 1523 - 1523
  • [24] Comparison of central corneal thickness measurements using optical low-coherence reflectometry, Fourier domain optical coherence tomography, and Scheimpflug camera
    Gonul, Saban
    Koktekir, Bengu Ekinci
    Bakbak, Berker
    Gedik, Sansal
    ARQUIVOS BRASILEIROS DE OFTALMOLOGIA, 2014, 77 (06) : 345 - 350
  • [25] Repeatability of central corneal thickness measurements of donor corneas in preservation chamber using Fourier-domain anterior segment optical coherence tomography
    Abhinav Golla
    Li Wang
    Caleb Morris
    Mitchell P. Weikert
    Sumitra S. Khandelwal
    Cell and Tissue Banking, 2018, 19 : 717 - 720
  • [26] Repeatability of central corneal thickness measurements of donor corneas in preservation chamber using Fourier-domain anterior segment optical coherence tomography
    Golla, Abhinav
    Wang, Li
    Morris, Caleb
    Weikert, Mitchell P.
    Khandelwal, Sumitra S.
    CELL AND TISSUE BANKING, 2018, 19 (04) : 717 - 720
  • [27] Fourier phase in Fourier-domain optical coherence tomography
    Uttam, Shikhar
    Liu, Yang
    JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 2015, 32 (12) : 2286 - 2306
  • [28] Consistency of corneal sublayer thickness measurements using Fourier-domain optical coherence tomography after phacoemulsification
    Lopez-Miguel, Alberto
    Calabuig-Goena, Maria
    Marques-Fernandez, Victoria
    Fernandez, Itziar
    Alio, Jorge L.
    Maldonado, Miguel J.
    EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2016, 26 (06) : 540 - 545
  • [29] Fourier-Domain Optical Coherence Tomography Imaging in Keratoconus A Corneal Structural Classification
    Sandali, Otman
    El Sanharawi, Mohamed
    Temstet, Cyril
    Hamiche, Taous
    Galan, Alice
    Ghouali, Wajdene
    Goemaere, Isabelle
    Basli, Elena
    Borderie, Vincent
    Laroche, Laurent
    OPHTHALMOLOGY, 2013, 120 (12) : 2403 - 2412
  • [30] In Vivo Comparison of Fourier-Domain Optical Coherence Tomography and Intravascular Ultrasonography
    Ramesh, Sathyadeepak
    Papayannis, Aristotelis
    Abdel-karim, Abdul-rahman
    Banerjee, Subhash
    Brilakis, Emmanouil
    JOURNAL OF INVASIVE CARDIOLOGY, 2012, 24 (03): : 111 - 115