Two-Dimensional Color Doppler Echocardiography for Left Ventricular Stroke Volume Assessment: A Comparison Study with Three-Dimensional Echocardiography

被引:9
|
作者
Da Silva, Cristina [1 ]
Pedro, Fatima [1 ]
Deister, Lizandra [1 ]
Sahlen, Anders [2 ]
Manouras, Aristomenis [2 ]
Shahgaldi, Kambiz [3 ]
机构
[1] Karolinska Univ Hosp, Dept Clin Physiol, Stockholm, Sweden
[2] Karolinska Univ Hosp, Dept Cardiol & Clin Physiol, Stockholm, Sweden
[3] Karolinska Univ Hosp, Dept Cardiol, Stockholm, Sweden
关键词
left ventricular outflow tract; left ventricular stroke volume; color Doppler; three-dimensional echocardiography; AORTIC-VALVE AREA; OUTFLOW TRACT; TRANSESOPHAGEAL ECHOCARDIOGRAPHY; EJECTION FRACTION; BIPLANE MODE; STENOSIS; ACCURACY; QUANTIFICATION; RECOMMENDATIONS;
D O I
10.1111/j.1540-8175.2012.01695.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aim: Whether measurement of left ventricular outflow tract diameter (LVOTd) using color Doppler (CD) in order to more accurately define LVOTd is more accurate for determination of stroke volume (SV) than gray scale and compare it with direct measurement of LVOT area (a) using three-dimensional echocardiography (3DE) for SV determination. Methods and Results: Twenty-one volunteers were examined. LVOTa was calculated by two-dimensional echocardiography (2DE) using the following formula: px (d/2)2, d = LVOT diameter by gray scale and CD, respectively. Planimetry of LVOTa was performed in parasternal long axis using 3DE. Eccentricity Index was calculated using the lateral and anterior-posterior LVOTd. SV was obtained by four different methods: (1) 2D gray scale, (2) 2D color, (3) LVOTa x LVOT velocity time integral, and (4) SV by Simpson's biplane method. Gray scale LVOTd was significantly smaller compared to LVOTd obtained with CD (P < 0.05). Significant differences occurred between LVOTa gray scale and CD (3.29 +/- 0.74 cm2 vs 3.67 +/- 0.70 cm2, P < 0.05) and between LVOTa calculated by gray scale in comparison to 3DE planimetry; (3.29 +/- 0.74 cm2 vs 3.61 +/- 0.89 cm2, P = 0.011). Half of the subjects had at least 17% difference between the lateral and anterior-posterior LVOTd. There were significant differences between SV by 2D gray scale and 2D CD (82.8 +/- 17.1 mL vs 92.4 +/- 16.8 mL, P < 0.05) and between 2D gray scale and 3DE planimetry (82.8 +/- 17.1 mL vs 90.7 +/- 19.8 mL, P = 0.025). Conclusion: Our study demonstrates LVOT being frequently elliptical. SV and LVOTa were found to be similar when comparing 2DE CD and 3DE planimetry and showed higher values in comparison to 2DE gray scale, which suggests 2DE CD to be an alternative approach for SV assessment.
引用
收藏
页码:766 / 772
页数:7
相关论文
共 50 条
  • [21] Comparison of three-dimensional echocardiography to two-dimensional echocardiography and fluoroscopy for monitoring of endomyocardial biopsy
    Amitai, Miriam E.
    Schnittger, Ingela
    Popp, Richard L.
    Chow, Judy
    Brown, Patricia
    Liang, David H.
    AMERICAN JOURNAL OF CARDIOLOGY, 2007, 99 (06): : 864 - 866
  • [22] Measurements of left ventricular volumes and ejection fraction with three-dimensional echocardiography: feasibility and agreement compared to two-dimensional echocardiography
    Vidar Ruddox
    Thor Edvardsen
    Morten Bækkevar
    Jan Erik Otterstad
    The International Journal of Cardiovascular Imaging, 2014, 30 : 1325 - 1330
  • [23] Measurements of left ventricular volumes and ejection fraction with three-dimensional echocardiography: feasibility and agreement compared to two-dimensional echocardiography
    Ruddox, Vidar
    Edvardsen, Thor
    Baekkevar, Morten
    Otterstad, Jan Erik
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2014, 30 (07): : 1325 - 1330
  • [24] Does Two-Dimensional Image Reconstruction from Three-Dimensional Full Volume Echocardiography Improve the Assessment of Left Ventricular Morphology and Function?
    Amzulescu, Mihaela Silvia
    Slavich, Massimo
    Florian, Anca
    Goetschalckx, Kaatje
    Voigt, Jens-Uwe
    ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 2013, 30 (01): : 55 - 63
  • [25] Left Ventricular Systolic Dyssynchrony Index by Three-Dimensional Echocardiography in Patients with Decreased Left Ventricular Function: Comparison with Tissue Doppler Echocardiography
    Tani, Tomoko
    Sumida, Toshiaki
    Tanabe, Kazuaki
    Ehara, Natsuhiko
    Yamaguchi, Kazuto
    Kawai, Junichi
    Yagi, Toshikazu
    Morioka, Shigefumi
    Fujiwara, Hiroshi
    Okada, Yukikatsu
    Kita, Toru
    Furukawa, Yutaka
    ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 2012, 29 (03): : 346 - 352
  • [26] CARDIAC TWIST BY TWO-DIMENSIONAL AND THREE-DIMENSIONAL ECHOCARDIOGRAPHY: AN IN VITRO COMPARISON STUDY
    Sahn, David J.
    Li, Karen
    Ashraf, Shiza
    Yang, Sarah
    Carlson, Max
    Newey, Colleen
    Xiong, Li
    Zhou, Zhiwen
    Ashraf, Muhammad
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2010, 55 (10)
  • [27] Validation of live three-dimensional echocardiography for quantification of absolute left ventricular volumes; comparison with volumetric three-dimensional echocardiography
    Shin, JH
    Shiota, T
    Qin, JX
    Agler, DA
    Saracino, G
    Greenberg, N
    Oryszak, S
    Thomas, JD
    EUROPEAN HEART JOURNAL, 2004, 25 : 260 - 260
  • [28] Quantitative assessment of left ventricular size and function in cardiac transplant recipients: Side-by-side comparison of real time two-dimensional echocardiography, contrast-enhanced two-dimensional echocardiography, three-dimensional echocardiography, and contrast-enhanced three-dimensional echocardiography as compared to magnetic resonance imaging
    Rodriguez-Manero, Moises
    Azcarate-Aguero, Pedro
    Kreidieh, Bahij
    Teresa Alvez, Maria
    Martinez-Monzonis, Amparo
    Diaz-Dorronsoro, Agnes
    Cid-Menendez, Adrian
    Ramon Gonzalez-Juanatey, Jose
    Barba-Cosials, Joaquin
    Rabago, Gregorio
    Bastarrika, Gorka
    ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 2019, 36 (02): : 306 - 311
  • [29] Comparison of two-dimensional and three-dimensional speckle tracking echocardiography for the assessment of myocardial viability
    Cwiek, E.
    Szymczyk, E.
    Kasprzak, J. D.
    Michalski, B.
    Stefanczyk, L.
    Wozniakowski, B.
    Rotkiewicz, A.
    Lipiec, P.
    EUROPEAN HEART JOURNAL, 2014, 35 : 551 - 552
  • [30] COLOR DOPPLER ASSESSMENT OF UNIVENTRICULAR ATRIOVENTRICULAR CONNECTION - COMPARISON WITH TWO-DIMENSIONAL ECHOCARDIOGRAPHY AND CINEANGIOGRAPHY
    VITARELLI, A
    BARILLA, F
    DADDIO, AP
    SCAPATO, A
    SAPONARO, A
    REALE, A
    CLINICAL RESEARCH, 1989, 37 (02): : A304 - A304