Slat collimator design issues for dual-head coincidence Imaging systems

被引:4
|
作者
Joung, J [1 ]
Miyaoka, RS [1 ]
Kohlmyer, SG [1 ]
Harrison, RL [1 ]
Lewellen, TK [1 ]
机构
[1] Univ Washington, Image Res Labs, Dept Radiol, Seattle, WA 98195 USA
关键词
dual-head coincidence imaging (DHCI); noise; equivalent count (NEC); salt collimator;
D O I
10.1109/TNS.2002.998741
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper investigates optimum slat collimator design parameters for dual-head coincidence imaging (DHCI) systems. Noise equivalent count (NEC) rate was examined with respect to the activity concentration under various system conditions. All results are derived from Monte Carlo simulations with a digital anthropomorphic (Zubal) phantom. The DHCI system was modeled after the Millennium VG gamma camera (GEMS, Waukesha, WI). The dead-time characteristics of the camera were experimentally determined. Our results suggests that substantial NEC gains can be achieved by varying the slat-to-slat separation, such that the peak of the NEC curve is located at clinically relevant levels (i.e., 0.07 similar to 0.10 muCi/cc). The NEC was also found to increase with the use of longer slats with appropriately selected slat-to-slat separation. Furthermore, the NEC performance also depends on the count-rate performance (i.e., dead-time losses) of the system. Therefore, as improvements are made to the count-rate capabilities of DHCI systems, the slat geometry should be modified. Further study is required to determine the effect that slat collimator design has on image quality and lesion detection for clinically realistic imaging situations.
引用
收藏
页码:141 / 146
页数:6
相关论文
共 50 条
  • [31] Initial experience with a prototype dual-crystal (LSO/NaI) dual-head coincidence camera in oncology
    Joshi, U
    Hoekstra, OS
    Boellaard, R
    Comans, EFI
    Raijmakers, PGHM
    Pijpers, RJ
    Miller, SD
    Teule, GJJ
    van Lingen, A
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2004, 31 (04) : 596 - 598
  • [32] Simultaneous multi-nuclide imaging via reconstruction-free double-photon emission coincidence imaging method with parallel hole collimator and slat collimator
    Yan, Linlin
    Uenomachi, Mizuki
    Ueki, Taisei
    Kim, Donghwan
    Shimazoe, Kenji
    Takahashi, Hiroyuki
    Kamada, Kei
    Shigekawa, Yudai
    Nambu, Akihiro
    Yin, Xiaojie
    Wang, Yang
    Haba, Hiromitsu
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2023, 1053
  • [33] Cardiac fluorine-18 fluorodeoxyglucose imaging using a dual-head gamma camera with coincidence detection: a clinical pilot study
    Johan De Sutter
    Frederic De Winter
    Christophe Van de Wiele
    Pieter De Bondt
    Yves D’Asseler
    Rudi Dierckx
    European Journal of Nuclear Medicine, 2000, 27 : 676 - 685
  • [34] Cardiac fluorine-18 fluorodeoxyglucose imaging using a dual-head gamma camera with coincidence detection: a clinical pilot study
    De Sutter, J
    De Winter, F
    Van de Wiele, C
    De Bondt, P
    D'Asseler, Y
    Dierckx, R
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 2000, 27 (06) : 676 - 685
  • [35] Assessment of malignancy in pulmonary lesions: FDG dual-head coincidence gamma camera imaging in association with serum tumor marker measurement
    Bousson, V
    Moretti, JL
    Weinmann, P
    Safi, N
    Tamgac, F
    Groiselle, C
    de Beco, V
    Hillali, Y
    Valeyre, D
    Breau, JL
    JOURNAL OF NUCLEAR MEDICINE, 2000, 41 (11) : 1801 - 1807
  • [36] Design of a novel slit-slat collimator system for SPECT imaging of the human brain
    Mahmood, S. T.
    Erlandsson, K.
    Cullum, I.
    Hutton, B. F.
    PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (11): : 3433 - 3449
  • [37] Rotating slat collimator design for high-energy near-field imaging
    Sharma, Amy
    Floyd, Carey
    Harrawood, Brian
    Tourassi, Georgia
    Kapadia, Anuj
    Bender, Janelle
    Lo, Joseph
    Howell, Calvin
    MEDICAL IMAGING 2006: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2006, 6142
  • [38] Role of 18F-FDG dual-head gamma-camera coincidence imaging in recurrent or metastatic colorectal carcinoma
    Even-Sapir, E
    Lerman, H
    Figer, A
    Rabau, M
    Livshitz, G
    Inbar, M
    Gutman, M
    JOURNAL OF NUCLEAR MEDICINE, 2002, 43 (05) : 603 - 609
  • [39] Fluorine-18 fluorodeoxyglucose imaging using dual-head coincidence positron emission tomography without attenuation correction in patients with head and neck cancer
    Pai, M
    Park, CH
    Suh, JH
    Koh, JH
    CLINICAL NUCLEAR MEDICINE, 1999, 24 (07) : 495 - 500
  • [40] The clinical value of 18F-FDG detection with a dual-head coincidence camera:: a review