Reproducibility of bone mineral density measurements using dual X-ray absorptiometry in daily clinical practice

被引:73
|
作者
El Maghraoui, A [1 ]
Zounon, AAD
Jroundi, I
Nouijai, A
Ghazi, M
Achemlal, L
Bezza, A
Tazi, MA
Abouqual, R
机构
[1] Military Hosp Mohammed 5, Rheumatol & Phys Rehabil Dept, Rabat, Morocco
[2] Fac Med & Pharm, Dept Epidemiol & Biostat, Rabat, Morocco
[3] Minist Hlth, Dept Epidemiol, Rabat, Morocco
关键词
chronic rheumatic diseases; dual energy X-ray absorptiometry (DXA); healthy volunteers; osteoporosis; post-menopausal women; reproducibility;
D O I
10.1007/s00198-005-1916-2
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Bone mineral density (BMD) measurements are frequently performed repeatedly for each patient. Subsequent BMD measurements allow reproducibility to be assessed. Previous studies have suggested that reproducibility may be influenced by age and clinical status. The purpose of the study was to examine the reproducibility of BMD by dual energy X-ray absorptiometry (DXA) and to investigate the practical value of different measures of reproducibility in three distinct groups of subjects: healthy young volunteers, postmenopausal women and patients with chronic rheumatic diseases. Two hundred twenty-two subjects underwent two subsequent BMD measurements of the spine and hip. There were 60 young healthy subjects, 102 postmenopausal women and 60 patients with chronic rheumatic diseases (33 rheumatoid arthritis, 10 ankylosing spondylitis and 10 other systemic diseases). Forty-five patients (75%) among the third group were receiving corticosteroids. Reproducibility was expressed as the smallest detectable difference (SDD), coefficient of variation (CV), least significant change (LSC) and intraclass correlation coefficient (ICC). Sources of variation were investigated by linear regression analysis. The median interval between measurements was 0 days (range 0-7). The mean difference (SD) between the measurements (g/cm(2)) was -0.0001 (+/- 0.003) and -0.0004 (+/- 0.002) at L1-L4 and the total hip, respectively. At L1-L4 and the total hip, SDD (g/cm(2)) was +/- 0.04 and +/- 0.02, CV (%) was 2.02 and 1.29, and LSC (%) 5.60 and 3.56, respectively. The ICC at the spine and hip was 0.99 and 0.99, respectively. Only a minimal difference existed between the groups. Reproducibility in the three groups studied was good. In a repeated DXA scan, a BMD change, the least significant change (LSC) or the SDD should be regarded as significant. Use of the SDD is preferable to use of the CV and LSC because of its independence from BMD and its expression in absolute units. Expressed as SDD, a BMD change of at least +/- 0.04 g/cm(2) at L1-L4 and +/- 0.02 g/cm(2) at the total hip should be considered significant. This reproducibility seems independent from age and clinical status and improved in the hips by measuring the dual femur.
引用
收藏
页码:1742 / 1748
页数:7
相关论文
共 50 条
  • [1] Reproducibility of bone mineral density measurements using dual X-ray absorptiometry in daily clinical practice
    A. El Maghraoui
    A. A. Do Santos Zounon
    I. Jroundi
    A. Nouijai
    M. Ghazi
    L. Achemlal
    A. Bezza
    M. A. Tazi
    R. Abouqual
    Osteoporosis International, 2005, 16 : 1742 - 1748
  • [2] Reproducibility of bone mineral density measurement using dual x-ray absorptiometry in daily clinical practice
    El Maghraoui, A
    Do Santos, AAW
    Jroundi, I
    Nouijal, A
    Ghazi, M
    Achemlal, L
    Bezza, A
    Tazi, MA
    Abouqual, R
    ANNALS OF THE RHEUMATIC DISEASES, 2005, 64 : 381 - 381
  • [3] Reproducibility of bone mineral density measurement using dual X-ray absorptiometry in daily clinical practice.
    El Maghraoui, A
    Do Santos, AA
    Jroundi, I
    Nouijai, A
    Ghazi, M
    Achemlal, L
    Bezza, A
    Tazi, MA
    Abouqual, R
    ARTHRITIS AND RHEUMATISM, 2005, 52 (09): : S290 - S290
  • [4] Bone mineral density measurements in otosclerosis using dual-energy X-ray absorptiometry
    Tonuk, S. B.
    Koybasi, S.
    Bicer, Y. O.
    Alkan Temel, S.
    CLINICAL OTOLARYNGOLOGY, 2016, 41 (02) : 149 - 153
  • [5] THE REPRODUCIBILITY OF BONE-DENSITY ANALYSES USING DUAL X-RAY ABSORPTIOMETRY
    EL-HAJJ FULEIHAN, G
    ANGELL, J
    PORRINO, N
    CHUNG, S
    LEBOFF, MS
    JOURNAL OF BONE AND MINERAL RESEARCH, 1992, 7 : S190 - S190
  • [6] Influence of strontium on bone mineral density and bone mineral content measurements by dual X-ray absorptiometry
    Nielsen, SP
    Slosman, D
    Sorensen, OH
    Basse-Cathalinat, B
    De Cassin, P
    Roux, C
    Meunier, PJ
    JOURNAL OF CLINICAL DENSITOMETRY, 1999, 2 (04) : 371 - 379
  • [7] BONE-MINERAL DENSITY IN HEALTHY NORMAL WOMEN AND REPRODUCIBILITY OF MEASUREMENTS IN SPINE AND HIP USING DUAL-ENERGY X-RAY ABSORPTIOMETRY
    HADDAWAY, MJ
    DAVIE, MWJ
    MCCALL, IW
    BRITISH JOURNAL OF RADIOLOGY, 1992, 65 (771): : 213 - 217
  • [8] The reliability of dual-energy X-ray absorptiometry measurements of bone mineral density in the metatarsals
    Joel T Fuller
    Jane Archer
    Jonathan D Buckley
    Margarita D Tsiros
    Dominic Thewlis
    Skeletal Radiology, 2016, 45 : 135 - 140
  • [9] Measurements and Evaluation of Proximal Femoral Bone Mineral Density with Dual Energy X-ray Absorptiometry
    Hayashida, Keiichi
    Takeda, Yoshihiro
    Katsuda, Toshizo
    Yamamoto, Kenyu
    Suesada, Yasuhide
    Shibata, Moeko
    Azuma, Masami
    ACTA MEDICA OKAYAMA, 2012, 66 (01) : 17 - 21
  • [10] Measurements and Evaluation of Proximal Femoral Bone Mineral Density with Dual Energy X-ray Absorptiometry
    Hayashida, K.
    Katsuda, T.
    Yamamoto, K.
    Takeda, Y.
    5TH EUROPEAN CONFERENCE OF THE INTERNATIONAL FEDERATION FOR MEDICAL AND BIOLOGICAL ENGINEERING, PTS 1 AND 2, 2012, 37 : 703 - +