Reproducibility of bone mineral density measurements using dual X-ray absorptiometry in daily clinical practice

被引:73
|
作者
El Maghraoui, A [1 ]
Zounon, AAD
Jroundi, I
Nouijai, A
Ghazi, M
Achemlal, L
Bezza, A
Tazi, MA
Abouqual, R
机构
[1] Military Hosp Mohammed 5, Rheumatol & Phys Rehabil Dept, Rabat, Morocco
[2] Fac Med & Pharm, Dept Epidemiol & Biostat, Rabat, Morocco
[3] Minist Hlth, Dept Epidemiol, Rabat, Morocco
关键词
chronic rheumatic diseases; dual energy X-ray absorptiometry (DXA); healthy volunteers; osteoporosis; post-menopausal women; reproducibility;
D O I
10.1007/s00198-005-1916-2
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Bone mineral density (BMD) measurements are frequently performed repeatedly for each patient. Subsequent BMD measurements allow reproducibility to be assessed. Previous studies have suggested that reproducibility may be influenced by age and clinical status. The purpose of the study was to examine the reproducibility of BMD by dual energy X-ray absorptiometry (DXA) and to investigate the practical value of different measures of reproducibility in three distinct groups of subjects: healthy young volunteers, postmenopausal women and patients with chronic rheumatic diseases. Two hundred twenty-two subjects underwent two subsequent BMD measurements of the spine and hip. There were 60 young healthy subjects, 102 postmenopausal women and 60 patients with chronic rheumatic diseases (33 rheumatoid arthritis, 10 ankylosing spondylitis and 10 other systemic diseases). Forty-five patients (75%) among the third group were receiving corticosteroids. Reproducibility was expressed as the smallest detectable difference (SDD), coefficient of variation (CV), least significant change (LSC) and intraclass correlation coefficient (ICC). Sources of variation were investigated by linear regression analysis. The median interval between measurements was 0 days (range 0-7). The mean difference (SD) between the measurements (g/cm(2)) was -0.0001 (+/- 0.003) and -0.0004 (+/- 0.002) at L1-L4 and the total hip, respectively. At L1-L4 and the total hip, SDD (g/cm(2)) was +/- 0.04 and +/- 0.02, CV (%) was 2.02 and 1.29, and LSC (%) 5.60 and 3.56, respectively. The ICC at the spine and hip was 0.99 and 0.99, respectively. Only a minimal difference existed between the groups. Reproducibility in the three groups studied was good. In a repeated DXA scan, a BMD change, the least significant change (LSC) or the SDD should be regarded as significant. Use of the SDD is preferable to use of the CV and LSC because of its independence from BMD and its expression in absolute units. Expressed as SDD, a BMD change of at least +/- 0.04 g/cm(2) at L1-L4 and +/- 0.02 g/cm(2) at the total hip should be considered significant. This reproducibility seems independent from age and clinical status and improved in the hips by measuring the dual femur.
引用
收藏
页码:1742 / 1748
页数:7
相关论文
共 50 条
  • [41] Dual energy X-ray absorptiometry in small rats with low bone mineral density
    Petersen, MM
    Nielsen, PK
    Lewin, E
    Olgaard, K
    CALCIFIED TISSUE INTERNATIONAL, 2000, 67 (06) : 455 - 459
  • [42] Bone mineral density and width of the femoral shaft by dual energy X-ray absorptiometry
    Kolta, S
    Fechtenbaum, J
    Listrat, V
    Dougados, M
    Roux, C
    JOURNAL OF BONE AND MINERAL RESEARCH, 1996, 11 : S581 - S581
  • [43] Assessment of bone mineral density by dual energy x-ray absorptiometry in patients with mucopolysaccharidoses
    Lin, Hsiang-Yu
    Shih, Shou-Chuan
    Chuang, Chih-Kuang
    Chen, Ming-Ren
    Niu, Dau-Ming
    Lin, Shuan-Pei
    ORPHANET JOURNAL OF RARE DISEASES, 2013, 8
  • [44] Evaluation of bone mineral density in CAPD patients with dual energy x-ray absorptiometry
    Pasadakis, P
    Thodis, E
    Mourvati, E
    Euthimiadou, A
    Margaritis, D
    Manavis, J
    Vargemezis, V
    ADVANCES IN PERITONEAL DIALYSIS/1996, VOL 12, 1996, 12 : 245 - 249
  • [45] Measurement of bone mineral density in osteoporotic fractures of the spine using dual energy x-ray absorptiometry
    Jahng, JS
    Lee, WI
    ORTHOPEDICS, 1996, 19 (11) : 951 - 954
  • [46] Hazards with evaluation of bone mineral density by dual-energy x-ray absorptiometry
    Bolosiu, CR
    OSTEOPOROSIS INTERNATIONAL, 2002, 13 : S127 - S127
  • [47] Development of Bone Mineral Density and Bone Mineral Content Measurements System Using a Dual Energy X-ray
    Promworn, Y.
    Pintavirooj, C.
    5TH BIOMEDICAL ENGINEERING INTERNATIONAL CONFERENCE (BMEICON 2012), 2012, : 102 - 102
  • [48] Development of Bone Mineral Density and Bone Mineral Content Measurements System Using a Dual Energy X-ray
    Promworn, Y.
    Pintavirooj, C.
    5TH BIOMEDICAL ENGINEERING INTERNATIONAL CONFERENCE (BMEICON 2012), 2012,
  • [49] DUAL ENERGY X-RAY ABSORPTIOMETRY AND QUANTITATIVE ULTRASOUND MEASUREMENTS OF BONE MINERAL DENSITY OF TRABECULAR BONE PHANTOMS IN THE PRESENCE OF STRONTIUM
    Pejovic-Milic, A. M.
    Jang, D. Hyun
    Da Silva, E.
    Slatkovska, L.
    Cheung, A. M.
    Tavakkoli, J.
    OSTEOPOROSIS INTERNATIONAL, 2016, 27 : S242 - S242
  • [50] EFFECT OF FEMORAL ROTATION ON BONE-MINERAL DENSITY-MEASUREMENTS WITH DUAL-ENERGY X-RAY ABSORPTIOMETRY
    GOH, JCH
    LOW, SL
    BOSE, K
    CALCIFIED TISSUE INTERNATIONAL, 1995, 57 (05) : 340 - 343