Comparison of air kerma standards for medium-energy x radiation between the MKEH and the IAEA

被引:1
|
作者
Csete, Istvan [1 ]
Czap, Ladislav [2 ]
Gomola, Igor [2 ]
机构
[1] Hungarian Trade Licensing Off MKEH, Budapest, Hungary
[2] IAEA, A-1400 Vienna, Austria
关键词
D O I
10.1088/0026-1394/49/1A/06011
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A comparison of the air kerma standards for medium-energy x radiation was performed between the MKEH and the IAEA. Two reference class ionization chambers of the IAEA, traceable to the PTB with volumes 1000 cm(3) and 100 cm(3), and the ISO 4037 N-40, N-60, N-120 and N-300 standard beam qualities were selected for the comparison. The calibration coefficients, N-K, were determined for both chambers at the MKEH in August 2011 and before and after this at the IAEA Dosimetry Laboratory. The comparison ratio of the calibration coefficients normalized to the previous comparison results of the PTB and MKEH primary air kerma standards were in the range from 0.989 to 0.994 for the beam qualities used. The relative expanded ( k = 2) uncertainty for each of these ratios is around 2 %.
引用
收藏
页数:7
相关论文
共 50 条
  • [31] Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the LNE-LNHB, France and the BIPM in medium-energy x-rays
    Burns, D. T.
    Kessler, C.
    Denoziere, M.
    Ksouri, W.
    METROLOGIA, 2008, 45
  • [32] Key comparison BIPM.RI(I)-K2 of the air-kerma standards of the MKEH, Hungary and the BIPM in low-energy x-rays
    Burns, D. T.
    Kessler, C.
    Roger, P.
    Csete, I.
    METROLOGIA, 2012, 49
  • [33] APMP/TCRI Key Comparison Report of Measurement of Air Kerma for Medium-Energy X-Rays (APMP.RI(I)-K3.2013)
    Chu C.H.
    Butler D.J.
    Tanaka T.
    Sathian V.
    Budiantari C.T.
    Yi C.-Y.
    Kim Y.H.
    Wu J.
    El-Sersy A.R.
    Dolah M.T.
    Jozela S.
    Ngcezu S.A.
    Patrao K.
    Metrologia, 2023, 60 (1 A)
  • [34] Comparison of the NIST and PTB Air-Kerma Standards for Low-Energy X-Rays
    O'Brien, Michelle
    Bueermann, Ludwig
    JOURNAL OF RESEARCH OF THE NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY, 2009, 114 (06) : 321 - 331
  • [35] APMP/TCRI key comparison report of measurement of air kerma for medium-energy x-rays (APMP.RI(I)-K3)
    Lee, J. H.
    Hwang, W. S.
    Kotler, L. H.
    Webb, D. V.
    Bueermann, L.
    Burns, D. T.
    Takeyeddin, M.
    Shaha, V. V.
    Srimanoroth, S.
    Meghzifene, A.
    Hah, S. H.
    Chun, K. J.
    Kadni, T. B.
    Takata, N.
    Msimang, Z.
    METROLOGIA, 2008, 45
  • [36] Comparison of the NIST and ENEA air kerma standards
    Laitano, RF
    Lamperti, PJ
    Toni, MP
    JOURNAL OF RESEARCH OF THE NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY, 1998, 103 (04) : 365 - 378
  • [37] Comparison of the air kerma standards for 137Cs and 60Co gamma-ray beams between the IAEA and the NIST
    Minniti, Ronaldo
    Czap, Ladislav
    METROLOGIA, 2011, 48
  • [38] Key comparison BIPM.RI(I)-K1 of the air-kerma standards of the MKEH, Hungary and the BIPM in 60Co gamma radiation
    Kessler, C.
    Burns, D.
    Machula, G.
    METROLOGIA, 2018, 55
  • [39] Comparison of the standards for air kerma of the ARPANSA and the BIPM for 60Co gamma radiation
    Kessler, C.
    Allisy-Roberts, P. J.
    Lye, J.
    Oliver, C.
    METROLOGIA, 2011, 48
  • [40] Comparison of the standards for air kerma of the BEV and the BIPM for 137Cs gamma radiation
    Kessler, C.
    Allisy-Roberts, P. J.
    Steurer, A.
    Tiefenboeck, W.
    Gabris, F.
    METROLOGIA, 2010, 47