Surface Deformation Analysis of End-to-End Stapled Intestinal Anastomosis

被引:4
|
作者
Yang, Jian [3 ,4 ,5 ]
Soltz, Michael [2 ]
Russell, Heinrich [2 ]
Beres, Jeff [2 ]
Zhao, Jingbo [3 ,4 ,5 ]
Liao, Donghua [3 ,4 ,5 ]
Gregersen, Hans [1 ,3 ,4 ,5 ]
机构
[1] Sino Danish Ctr Educ & Res, DK-8000 Aarhus, Denmark
[2] Covidien, New Haven, CT USA
[3] Aarhus Univ, DK-8000 Aarhus C, Denmark
[4] Aarhus Univ Hosp, DK-8000 Aarhus, Denmark
[5] Aarhus Univ Hosp, Aalborg, Denmark
关键词
anastomosis leakage; surface deformation; biomechanics; COLLAGEN CONTENT; GENE-EXPRESSION; WALL; COARCTATION;
D O I
10.1177/1553350611426566
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background. Stapling devices for creating anastomosis in internal organs are commonly used during surgery. Despite the obvious advantages of shortened procedure duration and fewer complications to manual suturing, staple-line leakage during intestinal anastomosis likely relates to the interaction between the staples and the tissue and to the tissue mechanical properties. The authors studied the deformation pattern close to the anastomosis to learn more about the mechanism involved in leakage. Methods. End-to-end anastomosis in pig small intestine was done using 21-mm circular staplers. Distension with pressure up to 100 cm H2O was done on the anastomosed segment. Surface markers were tracked using a microscope and a CCD camera. Circumferential and longitudinal strains were computed. Results. The staples restricted the deformation both in circumferential and longitudinal directions and induced a heterogeneous strain distribution. Circumferential strains were bigger between the staples (range 0.5-1) than inside the staples (range 0-0.3). The longitudinal strain ranged from 0 to slightly negative between the staples, indicating longitudinal compression. The negative strains turned into positive strains with increasing distance from the anastomosis. Further away from the anastomosis the longitudinal strain was in the range 0.3 to 0.5. Conclusion. The surface strain field was heterogeneously close to the stapled anastomosis. The longitudinal compression between staples in the longitudinal direction during inflation may have a beneficial effect preventing leakage, a phenomenon that needs further studies. The method may be useful in the design and validation of new staplers.
引用
收藏
页码:281 / 287
页数:7
相关论文
共 50 条
  • [41] Intrathoracic Linear Stapled Esophagogastric Anastomosis: An Alternative to the End to End Anastomosis
    Gorenstein, Lyall A.
    Bessler, Marc
    Sonett, Joshua R.
    ANNALS OF THORACIC SURGERY, 2011, 91 (01): : 314 - 316
  • [42] A COMPARATIVE STUDY OF INVERTING END-TO-END ANASTOMOSIS VS TELESCOPING END-TO-END ANASTOMOSIS IN DOG INTESTINE
    WHITE, D
    SINGLETO.AO
    TEXAS REPORTS ON BIOLOGY AND MEDICINE, 1967, 25 (03) : 502 - &
  • [43] Comparative analysis of Side-to-End and End-to-End intestinal anastomosis techniques: insights from a rat model study
    Azita shishegar
    Mahyar Tahmasian
    Ali Ashjaei
    Iraj Mirzaii-Dizgah
    Esmaeil Samizadeh
    BMC Surgery, 24 (1)
  • [44] INTESTINAL OBSTRUCTION IN A PREMATURE INFANT TREATED BY RESECTION AND PRIMARY END-TO-END ANASTOMOSIS
    BIERMAN, CW
    DAVIS, JK
    BIEHUSEN, FC
    JOURNAL OF PEDIATRICS, 1952, 40 (04): : 474 - 478
  • [45] END-TO-END ANASTOMOSIS IN THE SMALL INTESTINE
    TURNBULL, WR
    BRITISH MEDICAL JOURNAL, 1945, 2 (4420): : 405 - 405
  • [46] PARTIAL ESOPHAGECTOMY WITH END-TO-END ANASTOMOSIS
    KAY, EB
    BROWN, KL
    SURGERY, 1951, 29 (05) : 658 - 662
  • [47] End-to-end anastomosis of the pancreatic duct
    Losanoff, JE
    Richman, BW
    Jones, JW
    SURGERY, 2003, 133 (03) : 351 - 351
  • [48] EVALUATION OF INTESTINAL STAPLES FOR END-TO-END ANASTOMOSIS OF THE SMALL-INTESTINE IN THE HORSE
    SULLINS, KE
    STASHAK, TS
    MERO, KN
    VETERINARY SURGERY, 1985, 14 (02) : 87 - 92
  • [49] Aseptic end-to-end anastomosis of the intestine
    Burns, JG
    ANNALS OF SURGERY, 1925, 81 : 670 - 673
  • [50] Finite element modelling of stapled colorectal end-to-end anastomosis: Advantages of variable height stapler design
    Novacek, V.
    Tran, T. N.
    Klinge, U.
    Tolba, R. H.
    Staat, M.
    Bronson, D. G.
    Miesse, A. M.
    Whiffen, J.
    Turquier, F.
    JOURNAL OF BIOMECHANICS, 2012, 45 (15) : 2693 - 2697