Surface Deformation Analysis of End-to-End Stapled Intestinal Anastomosis

被引:4
|
作者
Yang, Jian [3 ,4 ,5 ]
Soltz, Michael [2 ]
Russell, Heinrich [2 ]
Beres, Jeff [2 ]
Zhao, Jingbo [3 ,4 ,5 ]
Liao, Donghua [3 ,4 ,5 ]
Gregersen, Hans [1 ,3 ,4 ,5 ]
机构
[1] Sino Danish Ctr Educ & Res, DK-8000 Aarhus, Denmark
[2] Covidien, New Haven, CT USA
[3] Aarhus Univ, DK-8000 Aarhus C, Denmark
[4] Aarhus Univ Hosp, DK-8000 Aarhus, Denmark
[5] Aarhus Univ Hosp, Aalborg, Denmark
关键词
anastomosis leakage; surface deformation; biomechanics; COLLAGEN CONTENT; GENE-EXPRESSION; WALL; COARCTATION;
D O I
10.1177/1553350611426566
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background. Stapling devices for creating anastomosis in internal organs are commonly used during surgery. Despite the obvious advantages of shortened procedure duration and fewer complications to manual suturing, staple-line leakage during intestinal anastomosis likely relates to the interaction between the staples and the tissue and to the tissue mechanical properties. The authors studied the deformation pattern close to the anastomosis to learn more about the mechanism involved in leakage. Methods. End-to-end anastomosis in pig small intestine was done using 21-mm circular staplers. Distension with pressure up to 100 cm H2O was done on the anastomosed segment. Surface markers were tracked using a microscope and a CCD camera. Circumferential and longitudinal strains were computed. Results. The staples restricted the deformation both in circumferential and longitudinal directions and induced a heterogeneous strain distribution. Circumferential strains were bigger between the staples (range 0.5-1) than inside the staples (range 0-0.3). The longitudinal strain ranged from 0 to slightly negative between the staples, indicating longitudinal compression. The negative strains turned into positive strains with increasing distance from the anastomosis. Further away from the anastomosis the longitudinal strain was in the range 0.3 to 0.5. Conclusion. The surface strain field was heterogeneously close to the stapled anastomosis. The longitudinal compression between staples in the longitudinal direction during inflation may have a beneficial effect preventing leakage, a phenomenon that needs further studies. The method may be useful in the design and validation of new staplers.
引用
收藏
页码:281 / 287
页数:7
相关论文
共 50 条
  • [21] MODIFICATION OF THE END-TO-END STAPLED ANASTOMOSIS UTILIZING INTERSECTING STAPLE LINES
    LEBLANC, KA
    RUSSO, VR
    ARCHIVES OF SURGERY, 1988, 123 (01) : 110 - 112
  • [22] END-TO-END ASEPTIC INTESTINAL ANASTOMOSIS IN INFANTS AND CHILDREN
    SWENSON, O
    SURGERY, 1954, 36 (02) : 192 - 197
  • [23] CONJOINED END-TO-END URETERO-INTESTINAL ANASTOMOSIS
    ALBERT, DJ
    PERSKY, L
    JOURNAL OF UROLOGY, 1971, 105 (02): : 201 - &
  • [24] Transcolonic approach for a stapled end-to-end colorectal anastomosis in low anterior resection
    Korenaga, D
    Orita, H
    Toh, Y
    Maekawa, S
    Ikeda, T
    Sugimachi, K
    HEPATO-GASTROENTEROLOGY, 1996, 43 (10) : 832 - 834
  • [25] END-TO-END ANASTOMOSIS
    MARIK, JJ
    ZANUTTA, EU
    INTERNATIONAL JOURNAL OF FERTILITY, 1983, 28 (01) : 29 - 30
  • [26] Long-term survival and complications following small intestinal resection and partially stapled, functional end-to-end anastomosis
    Kopec, Elizabeth Katherine
    Stevens, Martha
    Crowe, Oliver
    Wright, Chris
    Suthers, Joanna
    VETERINARY SURGERY, 2025, 54 (01) : 172 - 181
  • [27] Is "functional end-to-end anastomosis" really functional? A review of the literature on stapled anastomosis using linear staplers
    Kano, Masayuki
    Hanari, Naoyuki
    Gunji, Hisashi
    Hayano, Koichi
    Hayashi, Hideki
    Matsubara, Hisahiro
    SURGERY TODAY, 2017, 47 (01) : 1 - 7
  • [28] Is “functional end-to-end anastomosis” really functional? A review of the literature on stapled anastomosis using linear staplers
    Masayuki Kano
    Naoyuki Hanari
    Hisashi Gunji
    Koichi Hayano
    Hideki Hayashi
    Hisahiro Matsubara
    Surgery Today, 2017, 47 : 1 - 7
  • [29] A new instrument and operation for aseptic end-to-end intestinal anastomosis
    Webster, JP
    PROCEEDINGS OF THE SOCIETY FOR EXPERIMENTAL BIOLOGY AND MEDICINE, 1924, 21 : 579 - 581
  • [30] Early Experience of Robot-Assisted Esophagectomy With Circular End-to-End Stapled Anastomosis
    Wee, Jon O.
    Bravo-Iniguez, Carlos E.
    Jaklitsch, Michael T.
    ANNALS OF THORACIC SURGERY, 2016, 102 (01): : 253 - 259