Results from an amino acid racemization inter-laboratory proficiency study; design and performance evaluation

被引:17
|
作者
Powell, Joanne [1 ]
Collins, Matthew J. [1 ]
Cussens, James [3 ]
MacLeod, Norman [4 ]
Penkman, Kirsty E. H. [2 ]
机构
[1] Univ York, Dept Archaeol, BioArCh, York YO10 5DD, N Yorkshire, England
[2] Univ York, Dept Chem, BioArCh, York YO10 5DD, N Yorkshire, England
[3] Univ York, Dept Comp Sci, YCCSA, York YO10 5DD, N Yorkshire, England
[4] Nat Hist Museum, Dept Palaeontol, London SW7 5BD, England
基金
英国艺术与人文研究理事会;
关键词
Amino acid racemization; Geochronology; Inter-laboratory comparison; Proficiency test; Accuracy; Precision; Bias; Measurement uncertainty;
D O I
10.1016/j.quageo.2012.11.001
中图分类号
P9 [自然地理学];
学科分类号
0705 ; 070501 ;
摘要
It is nearly thirty years since the last inter-laboratory study was carried out for amino acid racemization (AAR) analysis using powdered fossil material (Wehmiller 1984). Since then there have been major changes in sample preparation and instrumentation, and it was considered timely to coordinate a new inter-laboratory study in support of current methodologies. In 2010, two such studies were undertaken. The first of these, coordinated by Wehmiller (this edition), used homogeneous hydrolysates of Pleistocene mollusc and eggshell materials and focused on the agreement of analytical measurements between laboratories, without interference from differing sample preparation procedures. The second (this study) was designed specifically as a proficiency test. Unlike previous inter-laboratory comparisons that have focussed on precision estimates, the purpose of this study was to carry out an evaluation of measurement bias by comparing the measurement results of laboratories carrying out their routine methods, including extraction, against the consensus values. Participants were sent one dried sample of a mixed amino acid standards solution and five homogeneous powders: two Pleistocene mollusc test materials prepared from material (ILC-A) supplied and used by Wehmiller in previous inter-laboratory studies (1984; and this edition), one Pleistocene opercula test material from the terrestrial gastropod, Bithynia tentaculata, and two heat-treated modern ostrich eggshell test materials. Results from this study demonstrate that whilst individual laboratory precision may be excellent, suggesting good control of random error influences (less than 1% for replicate measurements by some individual laboratories), agreement between methods, or even between laboratories carrying out the same method, may be very different. Trueness evaluation (determined as the relative percentage bias) reveals the extent of the disagreement reflected by the interlaboratory variability. Individual laboratory D/L value biases of 10-30% or more when compared to the consensus values are not uncommon. We demonstrate why bias contributions should also be included in AAR uncertainty estimation and recommend that the preparation of defined reference materials are seen as a priority in order to control and correct for systematic error influences in the analytical system. (C) 2012 Elsevier B.V. All rights reserved.
引用
收藏
页码:183 / 197
页数:15
相关论文
共 50 条
  • [21] RECENT PARTICIPATIONS OF THE RADIONUCLIDE METROLOGY LABORATORY FROM IFIN-HH TO PROFICIENCY TESTS AND INTER-LABORATORY COMPARISONS
    Luca, Aurelian
    Sahagia, Maria
    Antohe, Andrei
    Sulea, Cristina
    Berenstain, Robert
    PROCEEDINGS OF THE 3RD INTERNATIONAL PROFICIENCY TESTING CONFERENCE, PTCONF 2011, 2011, : 120 - 127
  • [22] A METHOD FREE FROM THE DISTRIBUTION FOR THE ANALYSIS OF RESULTS OF INTER-LABORATORY EXPERIMENTS
    LEONOV, VV
    FIRSANOV, VA
    INDUSTRIAL LABORATORY, 1981, 47 (03): : 287 - 290
  • [23] Viral Encephalo-Retinopathy Inter-laboratory Proficiency Test (VER-IPT): results from two years' experience
    Toffan, A.
    Buratin, A.
    Toson, M.
    Abbadi, M.
    Leardini, S.
    Pascoli, F.
    BULLETIN OF THE EUROPEAN ASSOCIATION OF FISH PATHOLOGISTS, 2021, 41 (02): : 59 - 69
  • [24] GRAPHICAL EVALUATION OF INTER-LABORATORY TEST RESULTS .2. APPLICATION IN PRACTICE
    BECKER, K
    SCHULTZE, GR
    ERDOL UND KOHLE ERDGAS PETROCHEMIE VEREINIGT MIT BRENNSTOFF-CHEMIE, 1970, 23 (07): : 423 - &
  • [25] EVALUATION OF INTER-LABORATORY PERFORMANCE FOR CREATINE-KINASE (CK) AND CK ISOENZYMES
    BOONE, J
    SAMPSON, E
    WHITNER, V
    LEWIS, S
    MCKNEALLY, S
    CLINICAL CHEMISTRY, 1979, 25 (06) : 1109 - 1109
  • [26] Evaluation of laboratory performance using proficiency test exercise results
    Siddique, N.
    Waheed, S.
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2012, 291 (03) : 817 - 823
  • [27] Evaluation of laboratory performance using proficiency test exercise results
    N. Siddique
    S. Waheed
    Journal of Radioanalytical and Nuclear Chemistry, 2012, 291 : 817 - 823
  • [28] INTER-LABORATORY VARIATION OF LEUKOCYTE DIFFERENTIAL COUNTS - RESULTS FROM THE FINNISH PROFICIENCY TESTING PROGRAM IN HEMATOLOGICAL MORPHOLOGY, 1974-1977
    RAJAMAKI, A
    SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 1979, 39 (07): : 613 - 617
  • [29] Proof-of-concept study for successful inter-laboratory comparison of MLVA results
    Larsson, J. T.
    Torpdahl, M.
    Nielsen, E. Moller
    EUROSURVEILLANCE, 2013, 18 (35) : 20 - 27
  • [30] An inter-laboratory comparison of probabilistic genotyping parameters and evaluation of performance on DNA mixtures from different laboratories
    Boodoosingh, Safia
    Kelly, Hannah
    Curran, James M.
    Kalafut, Tim
    FORENSIC SCIENCE INTERNATIONAL-GENETICS, 2024, 71