Comparison of Diagnostic Accuracy between Octopus 900 and Goldmann Kinetic Visual Fields

被引:21
|
作者
Rowe, Fiona J. [1 ]
Rowlands, Alison [2 ]
机构
[1] Univ Liverpool, Dept Hlth Serv Res, Liverpool L69 3GB, Merseyside, England
[2] Warrington & Halton Hosp NHS Fdn Trust, Dept Ophthalmol, Warrington WA5 1QG, Cheshire, England
关键词
IDIOPATHIC INTRACRANIAL HYPERTENSION; PERIMETRY;
D O I
10.1155/2014/214829
中图分类号
Q81 [生物工程学(生物技术)]; Q93 [微生物学];
学科分类号
071005 ; 0836 ; 090102 ; 100705 ;
摘要
Purpose. To determine diagnostic accuracy of kinetic visual field assessment by Octopus 900 perimetry compared with Goldmann perimetry. Methods. Prospective cross section evaluation of 40 control subjects with full visual fields and 50 patients with known visual field loss. Comparison of test duration and area measurement of isopters for Octopus 3, 5, and 10 degrees/sec stimulus speeds. Comparison of test duration and type of visual field classification for Octopus versus Goldmann perimetry. Results were independently graded for presence/absence of field defect and for type and location of defect. Statistical evaluation comprised of ANOVA and paired t test for evaluation of parametric data with Bonferroni adjustment. Bland Altman and Kappa tests were used for measurement of agreement between data. Results. Octopus 5 degrees/sec perimetry had comparable test duration to Goldmann perimetry. Octopus perimetry reliably detected type and location of visual field loss with visual fields matched to Goldmann results in 88.8% of results (K = 0.775). Conclusions. Kinetic perimetry requires individual tailoring to ensure accuracy. Octopus perimetry was reproducible for presence/absence of visual field defect. Our screening protocol when using Octopus perimetry is 5 degrees/sec for determining boundaries of peripheral isopters and 3 degrees/sec for blind spot mapping with further evaluation of area of field loss for defect depth and size.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] Comparison of Kinetic Visual Fields (Goldmann and Octopus) and Threshold Visual Fields (Octopus macular static and Full-Field Stimulus Threshold) in Patients with Inherited Retinal Diseases
    Schuchard, Ronald A.
    Birch, David G.
    Dagnelie, Gislin
    Koenekoop, Robert K.
    Bittner, Ava K.
    Wood, Leah M.
    Barnes, Claire S.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2014, 55 (13)
  • [2] Semi-automated kinetic perimetry: Comparison of the Octopus 900 and Humphrey visual field analyzer 3 versus Goldmann perimetry
    Bevers, Claudia
    Blanckaert, Gauthier
    Van Keer, Karel
    Fils, Jean-Francois
    Vandewalle, Evelien
    Stalmans, Ingeborg
    ACTA OPHTHALMOLOGICA, 2019, 97 (04) : E499 - E505
  • [3] MATURATION OF GOLDMANN KINETIC VISUAL-FIELDS
    GOLDBERG, MC
    PALAFOX, G
    MAYER, DL
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1992, 33 (04) : 713 - 713
  • [4] AUTOMATIC PERIMETER OCTOPUS - CLINICAL COMPARISON WITH THE VISUAL-FIELD DEVICE OF GOLDMANN
    KAMPIK, A
    LUND, OE
    GREITE, JH
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 1979, 175 (01) : 72 - 81
  • [5] Comparison between semiautomated kinetic perimetry and conventional Goldmann manual kinetic perimetry in advanced visual field loss
    Nowomiejska, K
    Vonthein, R
    Paetzold, J
    Zagorski, Z
    Kardon, R
    Schiefer, U
    OPHTHALMOLOGY, 2005, 112 (08) : 1343 - 1354
  • [6] COMPARISON OF MEASUREMENTS OF THE NEURORETINAL MARGIN ZONE WITH OCTOPUS VISUAL-FIELDS
    CLOUXFEY, U
    GLOOR, B
    KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 1986, 188 (05) : 485 - 485
  • [7] Study of Optimal Perimetric Testing In Children (OPTIC) - Comparison of outputs and normative data for Goldmann and Octopus kinetic perimetry
    Patel, Dipesh E.
    Walters, Bronwen C.
    Cumberland, Phillippa M.
    Russell-Eggitt, Isabelle
    Cortina-Borja, Mario
    Rahi, Jugnoo S.
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2015, 56 (07)
  • [8] Normal Values for the Full Visual Field, Corrected for Age- and Reaction Time, Using Semiautomated Kinetic Testing on the Octopus 900 Perimeter
    Grobbel, Julia
    Dietzsch, Janko
    Johnson, Chris A.
    Vonthein, Reinhard
    Stingl, Katarina
    Weleber, Richard G.
    Schiefer, Ulrich
    TRANSLATIONAL VISION SCIENCE & TECHNOLOGY, 2016, 5 (02):
  • [9] COMPARISON OF AUTOMATED STATIC PERIMETRY AND KINETIC GOLDMANN PERIMETRY IN THE ASSESSMENT OF VISUAL-FIELD IMPAIRMENT INDUCED BY BLEPHAROPTOSIS
    VERGARA, ME
    AGHAI, M
    MEYER, DR
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1994, 35 (04) : 1509 - 1509
  • [10] Comparison of Test Duration and Test Reliability Between Octopus 900 Perimetry and Humphrey SITA-Standard Perimetry
    Beyer, Michael
    Lodewyk, Kevin
    Nassiri, Nariman
    Kim, Chaesik
    Hughes, Bret A.
    Juzych, Mark S.
    Cyrlin, Marshall
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2017, 58 (08)