Four-channel surface coil array for sequential CW-EPR image acquisition

被引:6
|
作者
Enomoto, Ayano [1 ]
Emoto, Miho [2 ]
Fujii, Hirotada [2 ]
Hirata, Hiroshi [1 ]
机构
[1] Hokkaido Univ, Grad Sch Informat Sci & Technol, Div Bioengn & Bioinformat, Kita Ku, Sapporo, Hokkaido 0600814, Japan
[2] Sapporo Med Univ, Ctr Med Educ, Chuo Ku, Sapporo, Hokkaido 0608556, Japan
基金
日本学术振兴会;
关键词
Surface coil; Array; Decoupling; CW-EPR imaging; NITROXYL RADICALS; RESONATOR; SENSITIVITY; GENERATION; DESIGN; BRAIN; MOUSE; LIGHT; MICE; BODY;
D O I
10.1016/j.jmr.2013.06.007
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
This article describes a four-channel surface coil array to increase the area of visualization for continuous-wave electron paramagnetic resonance (CW-EPR) imaging. A 776-MHz surface coil array was constructed with four independent surface coil resonators and three kinds of switches. Control circuits for switching the resonators were also built to sequentially perform EPR image acquisition for each resonator. The resonance frequencies of the resonators were shifted using PIN diode switches to decouple the inductively coupled coils. To investigate the area of visualization with the surface coil array, three-dimensional EPR imaging was performed using a glass cell phantom filled with a solution of nitroxyl radicals. The area of visualization obtained with the surface coil array was increased approximately 3.5-fold in comparison to that with a single surface coil resonator. Furthermore, to demonstrate the applicability of this surface coil array to animal imaging, three-dimensional EPR imaging was performed in a living mouse with an exogenously injected nitroxyl radical imaging agent. (C) 2013 Elsevier Inc. All rights reserved.
引用
收藏
页码:21 / 29
页数:9
相关论文
共 50 条
  • [1] Sequential CW-EPR image acquisition with 760-MHz surface coil array
    Enomoto, Ayano
    Hirata, Hiroshi
    JOURNAL OF MAGNETIC RESONANCE, 2011, 209 (02) : 244 - 249
  • [2] Four-Channel Surface Coil Array for 300-MHz Pulsed EPR Imaging: Proof-of-Concept Experiments
    Enomoto, Ayano
    Hirata, Hiroshi
    Matsumoto, Shingo
    Saito, Keita
    Subramanian, Sankaran
    Krishna, Murali C.
    Devasahayam, Nallathamby
    MAGNETIC RESONANCE IN MEDICINE, 2014, 71 (02) : 853 - 858
  • [3] A Four-Channel Broadband MRI Receive Array Coil
    Hou, Jue
    Bauer, Courtney C.
    Sun, Chenhao
    Malone, Ben
    Griffin, Jay
    Wright, Steven M.
    2023 45TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE & BIOLOGY SOCIETY, EMBC, 2023,
  • [4] Sodium imaging of the heart at 7T: design, evaluation and application of a four-channel transmit/receive surface coil array
    A Graessl
    A Ruehle
    W Renz
    L Winter
    H Pfeiffer
    J Ruff
    J Rieger
    T Niendorf
    Journal of Cardiovascular Magnetic Resonance, 15 (Suppl 1)
  • [5] Nineteen-Channel Receive Array and Four-Channel Transmit Array Coil for Cervical Spinal Cord Imaging at 7T
    Zhao, Wei
    Cohen-Adad, Julien
    Polimeni, Jonathan R.
    Keil, Boris
    Guerin, Bastien
    Setsompop, Kawin
    Serano, Peter
    Mareyam, Azma
    Hoecht, Philipp
    Wald, Lawrence L.
    MAGNETIC RESONANCE IN MEDICINE, 2014, 72 (01) : 291 - 300
  • [7] FLEXIBLE READOUT IMAGE CAPTURE WITH A FOUR-CHANNEL CFA
    Pillman, Bruce
    Deever, Aaron
    Kumar, Mrityunjay
    2010 IEEE INTERNATIONAL CONFERENCE ON IMAGE PROCESSING, 2010, : 561 - 564
  • [8] Four-channel optically pumped magnetometer for a magnetoencephalography sensor array
    Iivanainen, Joonas
    Carter, Tony R.
    Dhombridge, Jonathan E.
    Read, Timothy S.
    Campbell, Kaleb
    Abate, Quinn
    Ridley, David M.
    Borna, Amir
    Schwindt, Peter D. D.
    OPTICS EXPRESS, 2024, 32 (10): : 18334 - 18351
  • [9] Four-channel EA array operates at 10 Gbit/s
    Powell, PN
    LASER FOCUS WORLD, 2000, : 10 - 10
  • [10] Four-channel EA array operates at 10 Gbit/s
    2000, PennWell Publ Co, Tulsa, OK, USA (36):