Characteristics and factors associated with intraocular lens tilt and decentration after cataract surgery

被引:52
|
作者
Chen, Xiaoyun [1 ]
Gu, Xiaoxun [1 ]
Wang, Wei [1 ]
Xiao, Wei [1 ]
Jin, Guangming [1 ]
Wang, Lanhua [1 ]
Dai, Ye [1 ]
Zhang, Enen [1 ]
Ruan, Xiaoting [1 ]
Liu, Zhenzhen [1 ]
Luo, Lixia [1 ]
Liu, Yizhi [1 ]
机构
[1] Sun Yat Sen Univ, Zhongshan Ophthalm Ctr, State Key Lab Ophthalmol, Guangzhou, Guangdong, Peoples R China
来源
基金
中国国家自然科学基金;
关键词
CAPSULAR BAG PERFORMANCE; ROTATIONAL STABILITY; ABERRATION; POSITION; 1-PIECE; SIZE;
D O I
10.1097/j.jcrs.0000000000000219
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To analyze the characteristics and factors associated with intraocular lens (IOL) bit and decentration after uneventful phacoemulsification with IOL implantation. Setting: Zhongshan Ophthalmic Center, Sun Yat-sen University, Guangzhou, China. Design: Cross-sectional study. Methods: All patients underwent a general ophthalmologic examination and anterior segment photography. IOL tilt and decentration were measured with a second-generation anterior segment optical coherence tomography (Casia2). Univariate and multivariate regression analyses were performed to assess the association between IOL bit and decentration with ocular biometric and systemic parameters, and the visual acuity. Results: A total of 196 eyes of 196 patients were enrolled in this study. IOLs showed a mean tilt of 4.8 degrees toward the inferotemporal direction, and the mean decentration was 0.21 mm. Both eyes presented a mirror symmetry relationship. Twenty-two eyes (11.22%) had a tilt greater than 7 degrees, and 21 eyes (10.72%) had a decentration more than 0.4 mm. Multivariate regression analysis showed previous pars plane vitrectomy (PPV) and short axial length (AL) were associated with greater IOL tilt (P = .014 and P < .001). In addition, long AL, thicker lens, and less capsulorhexis-OL overlap were positively correlated with decentration (P < .001, P = .029, and P = .026). Corrected cstanoe visual acuity did not directly correlate with IOL tilt and decentration (P = .417 and P = .550). Conclusions: PPV history and short AL were associated with greater IOL tilt, whereas longer AL, thicker lens, and overlarge capsulorhexis contribute to greater decentration. Implantation of toric and multifocal IOLs in these patients should be cautious. Copyright (C) 2020 Published by Wolters Kluwer on behalf of ASCRS and ESCRS
引用
收藏
页码:1126 / 1131
页数:6
相关论文
共 50 条
  • [41] Determination of Intraocular Lens Tilt and Decentration Using Simple and Rapid Method
    Mingxing Wu Huaming Li Weirong ChengZhongshan Ophthalmic Center Sun Yatsen University of Medical Sciences Guangzhou China
    眼科学报, 1998, (01) : 13 - 16+26
  • [42] Incidence of Intraocular Lens Exchange after Cataract Surgery
    Elsayed, Maram E. A. Abdalla
    Ahmad, Khabir
    Al-Abdullah, Abdulelah A.
    Malik, Rizwan
    Khandekar, Rajiv
    Martinez-Osorio, Hernan
    Mura, Marco
    Schatz, Patrik
    SCIENTIFIC REPORTS, 2019, 9 (1)
  • [43] Incidence of Intraocular Lens Exchange after Cataract Surgery
    Maram E. A. Abdalla Elsayed
    Khabir Ahmad
    Abdulelah A. Al-Abdullah
    Rizwan Malik
    Rajiv Khandekar
    Hernan Martinez-Osorio
    Marco Mura
    Patrik Schatz
    Scientific Reports, 9
  • [44] Metaanalysis of cataract after phakic intraocular lens surgery
    Budo, Camille J.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2008, 34 (11): : 1818 - 1819
  • [45] UNEXPLAINED INTRAOCULAR TOXICITY AFTER CATARACT-INTRAOCULAR LENS SURGERY
    OLSON, RJ
    APPLE, DJ
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 1987, 13 (06): : 688 - 689
  • [46] A theoretical study of intraocular lens tilt and decentration on perceptual image quality
    Turuwhenua, J
    OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2005, 25 (06) : 556 - 567
  • [47] Rotational Stability, Decentration, and Tilt of a New Hydrophobic Acrylic Intraocular Lens
    Roggla, Veronika
    Schartmueller, Daniel
    Schwarzenbacher, Luca
    Leydolt, Christina
    Menapace, Rupert
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2023, 250 : 149 - 156
  • [48] INVIVO MEASUREMENT OF POSTERIOR CHAMBER INTRAOCULAR-LENS DECENTRATION AND TILT
    AURAN, JD
    KOESTER, CJ
    DONN, A
    ARCHIVES OF OPHTHALMOLOGY, 1990, 108 (01) : 75 - 79
  • [49] Intraocular lens tilt and decentration, anterior chamber depth, and refractive error after trans-scleral suture fixation surgery
    Hayashi, K
    Hayashi, H
    Nakao, F
    Hayashi, F
    OPHTHALMOLOGY, 1999, 106 (05) : 878 - 882
  • [50] Possible Predisposing Factors for Late Intraocular Lens Dislocation After Routine Cataract Surgery
    Krepste, Lina
    Kuzmiene, Loreta
    Miliauskas, Arunas
    Januleviciene, Ingrida
    MEDICINA-LITHUANIA, 2013, 49 (05): : 229 - 234