Comparison of optic nerve head parameters using Heidelberg Retina Tomograph 3 and spectral-domain optical coherence tomography

被引:16
|
作者
Sato, Shino [1 ]
Hirooka, Kazuyuki [1 ]
Baba, Tetsuya [1 ]
Shiraga, Fumio [1 ]
机构
[1] Kagawa Univ, Fac Med, Dept Ophthalmol, Miki, Kagawa 7610793, Japan
来源
关键词
Heidelberg Retina Tomograph; optic nerve head; optical coherence tomography; FIBER LAYER THICKNESS; SCANNING LASER OPHTHALMOSCOPY; STRATUS OCT; GLAUCOMATOUS EYES; AGREEMENT; REPRODUCIBILITY; TOPOGRAPHY;
D O I
10.1111/j.1442-9071.2012.02782.x
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background: To assess the agreement between the Heidelberg Retina Tomograph and Cirrus spectral-domain optical coherence tomography (Cirrus HD-OCT [Carl Zeiss Meditec, Dublin, CA, USA]) when measuring optic disc parameters. Design: Prospective, cross-sectional study. Participants: A total of 96 glaucoma patients and 21 normal subjects were analysed. Methods: Optic nerve head measurements, including disc area, rim area, cup-to-disc ratio and cup volume were obtained using both the Cirrus HD-OCT and Heidelberg Retina Tomograph 3 (Heidelberg Engineering GmbH, Heidelberg, Germany). Main Outcome Measurements: BlandAltman plots were used to evaluate the agreement between each of the optic disc parameters. Results: Although Cirrus HD-OCT values were smaller than Heidelberg Retina Tomograph 3 values for the disc and rim areas, overall the Cirrus HD-OCT and Heidelberg Retina Tomograph 3 measurements were highly correlated (r = 0.657 to 0.821). As compared with Heidelberg Retina Tomograph 3, disc and rim areas tended to be minimized by Cirrus HD-OCT, especially when there were large disc and rim areas, although there was no relationship between the discrepancy and the level of measurement of the cup-to-disc ratio. The optical coherence tomography overestimated the Heidelberg Retina Tomograph-determined cup volume, especially when there were large cup volumes. Conclusions: BlandAltman analyses revealed that with the exception of the cup-to-disc ratio, there was poor agreement between the Cirrus HD-OCT and Heidelberg Retina Tomograph 3 morphometric measurements. In addition, with the exception of the cup-to-disc ratio, the optic nerve head measurements could not be directly compared between the Cirrus HD-OCT and Heidelberg Retina Tomograph 3.
引用
收藏
页码:721 / 726
页数:6
相关论文
共 50 条
  • [1] Comparison of Spectral-Domain Optical Coherence Tomography and Heidelberg Retina Tomograph III Optic Nerve Head Parameters in Glaucoma
    Seymenoglu, Goktug
    Baser, Esin
    Ozturk, Bilge
    OPHTHALMOLOGICA, 2013, 229 (02) : 101 - 105
  • [2] Effect of head tilt on repeatability of optic nerve head parameters using Cirrus spectral-domain optical coherence tomography
    Koh, Lilian Hui Li
    Ismail, Muhammad Amir
    Yap, Sae Cheong
    Wong, Elizabeth Poh Ying
    Yip, Leonard Wei Leon
    INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2016, 9 (08) : 1170 - 1175
  • [3] Effect of head tilt on repeatability of optic nerve head parameters using Cirrus spectral-domain optical coherence tomography
    Lilian Hui Li Koh
    Muhammad Amir Ismail
    Sae Cheong Yap
    Elizabeth Poh Ying Wong
    Leonard Wei Leon Yip
    International Journal of Ophthalmology, 2016, (08) : 1170 - 1175
  • [4] Optic nerve head assessment: comparison of Cirrus optic coherence tomography and Heidelberg Retinal Tomograph 3
    Kratz, Assaf
    Lim, Ridia
    Goldberg, Ivan
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2014, 42 (08): : 734 - 744
  • [5] Objective optic nerve head assessment using optical coherence tomography and Heidelberg retinal tomograph
    Morgan, William H.
    Kang, Min-Hye
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2014, 42 (08): : 711 - 712
  • [6] Optic Disc Imaging with Spectral-Domain Optical Coherence Tomography Variability and Agreement Study with Heidelberg Retinal Tomograph
    Yang, Bingzhi
    Ye, Cong
    Yu, Marco
    Liu, Shu
    Lam, Dennis Shun Chiu
    Leung, Christopher Kai-shun
    OPHTHALMOLOGY, 2012, 119 (09) : 1852 - 1857
  • [7] Morphologic Characteristics of Optic Nerve Head Drusen on Spectral-Domain Optical Coherence Tomography
    Lee, Kyoung Min
    Woo, Se Joon
    Hwang, Jeong-Min
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2013, 155 (06) : 1139 - 1147
  • [8] Spectral-domain optical coherence tomography evaluation of the cornea, retina, and optic nerve in normal horses
    Pinto, Nelson I.
    Gilger, Brian C.
    VETERINARY OPHTHALMOLOGY, 2014, 17 : 140 - 148
  • [9] Comparison of Optic Nerve Head Parameter Measurements Obtained by Time-domain and Spectral-domain Optical Coherence Tomography
    Savini, Giacomo
    Barboni, Piero
    Carbonelli, Michele
    Sbreglia, Alessandra
    Deluigi, Giulia
    Parisi, Vincenzo
    JOURNAL OF GLAUCOMA, 2013, 22 (05) : 384 - 389
  • [10] Differentiation of Optic Nerve Head Drusen and Optic Disc Edema with Spectral-Domain Optical Coherence Tomography
    Lee, Kyoung Min
    Woo, Se Joon
    Hwang, Jeong-Min
    OPHTHALMOLOGY, 2011, 118 (05) : 971 - 977