Effect of head tilt on repeatability of optic nerve head parameters using Cirrus spectral-domain optical coherence tomography

被引:0
|
作者
Lilian Hui Li Koh [1 ]
Muhammad Amir Ismail [1 ]
Sae Cheong Yap [1 ]
Elizabeth Poh Ying Wong [1 ]
Leonard Wei Leon Yip [1 ]
机构
[1] Department of Ophthalmology, Tan Tock Seng Hospital,National Healthcare Group Eye Institute
关键词
glaucoma; imaging; artifacts; optical coherence tomography; head movements;
D O I
暂无
中图分类号
R770.4 [眼科诊断学];
学科分类号
100212 ;
摘要
AIM: To assess the repeatability of measuring optic nerve head(ONH) parameters using the Cirrus optical coherence tomography(OCT), as well as to assess the effect of head tilt on these measurements.METHODS: Thirty healthy participants with no evidence of glaucoma were recruited for the study. Visual acuity, intraocular pressure, standard automated perimetry and ocular examination were performed for each participant. One eye was then randomly selected and scanned undilated with the Cirrus OCT in 3 positions(neutral, 30° right tilt and 30° left tilt).RESULTS: Data collected from 29 eyes were used for analysis. One patient was omitted due to poor scan quality. The repeatability of the ONH parameters was analyzed using analysis of variance, coefficient of variation(COV) and intraclass correlation coefficient(ICC). Analysis of variance showed no statistically significant difference between 3 scans in a single position. There was good agreement between measurements(ICC 0.919-0.996, COV 1.94%-5.48%).Even with the presence of head tilt, repeated scans in the 3 positions showed good agreement as well(ICC 0.888-0.996, COV 2.04%-5.39%).CONCLUSION: Serial measurements of ONH parameters using the Cirrus OCT are found to have good repeatability. The ONH parameters with Cirrus OCT also maintain good repeatability despite head tilt.
引用
收藏
页码:1170 / 1175
页数:6
相关论文
共 50 条
  • [1] Effect of head tilt on repeatability of optic nerve head parameters using Cirrus spectral-domain optical coherence tomography
    Koh, Lilian Hui Li
    Ismail, Muhammad Amir
    Yap, Sae Cheong
    Wong, Elizabeth Poh Ying
    Yip, Leonard Wei Leon
    INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2016, 9 (08) : 1170 - 1175
  • [2] Effect of Head Tilt on Retinal Nerve Fiber Layer Thickness and Cup Disc Parameters by Cirrus and Spectralis Spectral-Domain Optical Coherence Tomography
    Koh, Lilian
    Bin Ismail, Muhammad Amir
    Yap, Sae Chong
    Yip, Leonard
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2014, 55 (13)
  • [3] Morphologic Characteristics of Optic Nerve Head Drusen on Spectral-Domain Optical Coherence Tomography
    Lee, Kyoung Min
    Woo, Se Joon
    Hwang, Jeong-Min
    AMERICAN JOURNAL OF OPHTHALMOLOGY, 2013, 155 (06) : 1139 - 1147
  • [4] Comparison of optic nerve head parameters using Heidelberg Retina Tomograph 3 and spectral-domain optical coherence tomography
    Sato, Shino
    Hirooka, Kazuyuki
    Baba, Tetsuya
    Shiraga, Fumio
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2012, 40 (07): : 721 - 726
  • [5] Optic Nerve Head Parameters Measured with Spectral-Domain Optical Coherence Tomography in Healthy Turkish Children: Normal Values, Repeatability, and interocular Symmetry
    Kiziloglu, Ozge Yabas
    Toygar, Okan
    Toygar, Baha
    Hacimustafaoglu, Ali Murat
    NEURO-OPHTHALMOLOGY, 2018, 42 (02) : 83 - 89
  • [6] Differentiation of Optic Nerve Head Drusen and Optic Disc Edema with Spectral-Domain Optical Coherence Tomography
    Lee, Kyoung Min
    Woo, Se Joon
    Hwang, Jeong-Min
    OPHTHALMOLOGY, 2011, 118 (05) : 971 - 977
  • [7] The effect of myopic optic disc tilt on measurement of spectral-domain optical coherence tomography parameters
    Shin, Hye-Young
    Park, Hae-Young Lopilly
    Park, Chan Kee
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2015, 99 (01) : 69 - 74
  • [8] Optic Nerve Head Development in Healthy Infants and Children Using Handheld Spectral-Domain Optical Coherence Tomography
    Patel, Aarti
    Purohit, Ravi
    Lee, Helena
    Sheth, Viral
    Maconachie, Gail
    Papageorgiou, Eleni
    McLean, Rebecca J.
    Gottlob, Irene
    Proudlock, Frank A.
    OPHTHALMOLOGY, 2016, 123 (10) : 2147 - 2157
  • [9] Optic nerve head analysis using confocal scanning laser ophthalmoscopy and spectral-domain optical coherence tomography
    Roberti, Gloria
    Oddone, Francesco
    Tanga, Lucia
    Michelessi, Manuele
    De Geronimo, Daniele
    Centofanti, Marco
    Manni, Gianluca
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2013, 54 (15)
  • [10] Comparison of Spectral-Domain Optical Coherence Tomography and Heidelberg Retina Tomograph III Optic Nerve Head Parameters in Glaucoma
    Seymenoglu, Goktug
    Baser, Esin
    Ozturk, Bilge
    OPHTHALMOLOGICA, 2013, 229 (02) : 101 - 105