Evaluation of the automated coagulation analyzer SYSMEX CA 6000

被引:17
|
作者
Quehenberger, P [1 ]
Kapiotis, S [1 ]
Handler, S [1 ]
Ruzicka, K [1 ]
Speiser, W [1 ]
机构
[1] Univ Vienna, Clin Inst Med & Chem Lab Diagnost, A-1090 Vienna, Austria
关键词
evaluation; coagulation analyzer; SYSMEX CA 6000;
D O I
10.1016/S0049-3848(99)00069-9
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
In the present study the coagulation analyzer SYSMEX CA 6000 (TOA Medical Electronics Co., Kobe, Japan), an analyzer equipped with a photo-optical clot detection unit and a cap-piercing system, was evaluated with respect to its technical characteristics in the determination of standard coagulation tests (prothrombin time, activated partial thromboplastin time, thrombin time, fibrinogen, and antithrombin) and in the determination of coagulation single factor activities. In the normal and in the pathological range the intraassay coefficients of variation (CV) and interassay CV for most parameters were below 5% (exceptions: intraassay CV 5.4% for prolonged thrombin time; intraassay CV 9.26% and interassay CV 10.7% for decreased antithrombin; interassay CV 5.62% for fibrinogen in the normal range, intraassay CV 10.1% for fibrinogen greater than 7.0 g/L; intraassay CV 6.36% and interassay CV 11.7% for decreased fibrinogen; interassay CV 11.6% for prolonged activated partial thromboplastin time; interassay CV 6.12% for decreased factor VII). Interference studies with lipemic, icteric, and hemolytic samples showed just minor influences of these abnormal sample characteristics on prothrombin time, activated partial thromboplastin time, fibrinogen, and antithrombin measurements when compared to the results obtained by using mechanical clot detection (STA, Stage Diagnostica, Asnieres-Sur-Seine, France). No carryover was detected in alternating measurements of heparinized (3 U/mL unfractionated heparin) and normal plasma samples. Measurement of the activities of clotting factors V, VII, VIII, and IX showed a good correlation (r=0.993 to r=0.977) between SYSMEX CA 6000 and STA. Our results demonstrate that using SYSMEX CA 6000 analyzer basal routine coagulation testing as well as specialized tests for single factor activities can be performed with satisfactory precision; in particular, the cap-piercing system has no negative effect on the performance of the analyzer. (C) 1999 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:65 / 71
页数:7
相关论文
共 50 条
  • [31] Evaluation and performance characteristics of the automated coagulation analyzer ACL TOP
    Appert-Flory, Anny
    Fischer, Florence
    Jambou, Didier
    Toulon, Pierre
    THROMBOSIS RESEARCH, 2007, 120 (05) : 733 - 743
  • [32] Automated Plasmodium detection by the Sysmex XN hematology analyzer
    Dumas, Cecile
    Bienvenu, Anne-Lise
    Girard, Sandrine
    Picot, Stephane
    Debize, Gisele
    Durand, Brigitte
    JOURNAL OF CLINICAL PATHOLOGY, 2018, 71 (07) : 594 - 599
  • [33] PERFORMANCE EVALUATION OF ASSAY APPLICATIONS ON SYSMEX® CS-2000I COAGULATION SYSTEM ANALYZER
    Markowski, Mary Lou
    Kottke-Marchant, Kandice
    Gosselin, Robert
    Dwyre, Denis
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2012, 34 : 161 - 162
  • [34] Sysmex CA-6000: A new tool for old assays
    Proulle, V
    Duchemin, J
    Coville, E
    LeLeu, P
    Tchernia, G
    Dreyfus, M
    THROMBOSIS AND HAEMOSTASIS, 1997, : P2590 - P2590
  • [35] Evaluation of the performance of sysmex XN-3100 automated hematology analyzer regarding the sysmex XE-2100 and microscopic examination
    Incir, Said
    Palaoglu, Kerim Erhan
    TURKISH JOURNAL OF BIOCHEMISTRY-TURK BIYOKIMYA DERGISI, 2021, 46 (01): : 27 - 35
  • [36] Evaluation of the Succeeder SF-8200 Fully Automated Coagulation Analyzer
    Zengi, Oguzhan
    Kucuk, Suat H.
    CLINICAL LABORATORY, 2021, 67 (10) : 2316 - 2322
  • [37] PERFORMANCE EVALUATION OF THE NEW MEASUREMENT CHANNELS ON THE AUTOMATED SYSMEX XN-9000 HEMATOLOGY ANALYZER
    Schapkaitz, Elise
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2018, 40 : 127 - 127
  • [38] Evaluation of bone marrow aspirates using the automated hematology analyzer Sysmex XN-3000
    Tsuchiya, Koji
    Kimura, Konobu
    Ai, Tomohiko
    Yamamoto, Takamasa
    Nagasaka, Kaori
    Takemura, Hiroyuki
    Edahiro, Yoko
    Saito, Kaori
    Horiuchi, Yuki
    Misawa, Shigeki
    Takaku, Tomoiku
    Ando, Jun
    Ando, Miki
    Miida, Takashi
    Uchihashi, Kinya
    Tabe, Yoko
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2023, 45 (04) : 460 - 468
  • [39] Performance evaluation of the new measurement channels on the automated Sysmex XN-9000 hematology analyzer
    Schapkaitz, E.
    Raburabu, S.
    CLINICAL BIOCHEMISTRY, 2018, 53 : 132 - 138
  • [40] AUTOMATED IMMATURE GRANULOCYTE COUNTS ON THE NEW SYSMEX XN AUTOMATED HEMATOLOGY ANALYZER
    Rosenthal, Nancy
    Connell, Barbara
    Brown, Bonnie
    Kruger, Julie
    Capper, Mary
    Blaine, Kimberly
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2012, 34 : 130 - 130