Applicability of new configuration of open tubular solid phase microextraction for determination of free (unconjugated) testosterone esters by gas chromatography with barrier ionization discharge detector
Testosterone esters;
Barrier ionization discharge detector;
Open tube trapping solid-phase micro;
extraction;
Validation;
Gas chromatography;
MASS-SPECTROMETRY;
ANABOLIC-STEROIDS;
CAPILLARY MICROEXTRACTION;
THERMAL-DESORPTION;
WATER SAMPLES;
HUMAN HAIR;
VALIDATION;
AGENTS;
DRUGS;
D O I:
10.1016/j.microc.2022.108104
中图分类号:
O65 [分析化学];
学科分类号:
070302 ;
081704 ;
摘要:
The methodology has been developed for analyzing testosterone esters in urine by gas chromatography with a barrier ionization discharge (GC-BID) detector for anti-doping analysis. For sample preparation, poly-dimethylsiloxane (PDMS) was used as a trapping polymer in the new configuration of an open tube trapping solid-phase microextraction (OTT-SPME). The new device developed by this research group was called Intra Tube Flux Extraction (IT-FEX-SPME). The performance of the IT-FEX-SPME-GC-BID method was validated concerning extraction kinetics, sensitivity: the limit of detection (LOD 0.5-0.8 ng mL-1) and the limit of quantification (LOQ 1.5-2.4 ng mL-1), repeatability (<8%), reproducibility <7.5 %), linearity (R2>0.991), re-covery (mean 102.44 %), and stability (<30 %). The values obtained are acceptable, and the results showed the potentiality of the methodology to control testosterone ester doping. The qualitative performance of the pro-posed method is compared with those that use a mass spectrometer as a detector.