The reliability of computed tomography (CT) values and dimensional measurements of the oropharyngeal region using cone beam CT: comparison with multidetector CT

被引:89
|
作者
Yamashina, A. [1 ]
Tanimoto, K. [1 ]
Sutthiprapaporn, P. [1 ,2 ]
Hayakawa, Y. [3 ]
机构
[1] Hiroshima Univ, Grad Sch Biomed Sci, Dept Oral & Maxillofacial Surg, Hiroshima 730, Japan
[2] Khon Kaen Univ, Fac Dent, Dept Oral Diag, Khon Kaen, Thailand
[3] Kitami Inst Technol, Fac Engn, Dept Comp Sci, Kitami, Hokkaido, Japan
关键词
computed tomography; cone beam; computed tomography value; oropharynx;
D O I
10.1259/dmfr/45926904
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: To evaluate the reliability of cone beam CT (CBCT) values and dimensional measurements of oropharyngeal air spaces as compared with those of multidetector row CT (MDCT). Methods: A soft tissue equivalent phantom with different sized holes was used. The phantom was scanned using CBCT and MDCT. In addition, a volunteer was examined using both methods for clinical comparison. The CT data were retrieved to a personal computer and treated using image processing software for three-dimensional rendering and dimensional measurements. The CT values of air, water and soft tissues were measured experimentally and the CT values of air, fat and muscle were measured in a clinical case. The diameter of each hole drilled in the phantom was measured on CBCT and MDCT images using the software tool and digimatic callipers. Results: The CBCT values were quite different from the Hounsfield units obtained with a typical MDCT system. The standard deviations were almost ten times larger with CBCT. In vivo assessment showed that the CBCT values for fat had a wide range that partially overlapped the values for muscle. The measurement of distances greater than 4 mm was consistent for all methods. The difference for holes was less than 0.2 mm. Conclusions: The phantom study showed that CBCT provides limited quantitative CT values for each pixel on sliced images for differentiating air, water and soft tissues. However, the measurement of air spaces with CBCT was quite accurate.
引用
收藏
页码:245 / 251
页数:7
相关论文
共 50 条
  • [11] Region of interest cone beam tomography with prior CT data
    Ramamurthi, K
    Prince, JL
    CONFERENCE RECORD OF THE THIRTY-SEVENTH ASILOMAR CONFERENCE ON SIGNALS, SYSTEMS & COMPUTERS, VOLS 1 AND 2, 2003, : 1924 - 1927
  • [12] Cone beam CT paranasal sinuses versus standard multidetector and low dose multidetector CT studies
    Al Abduwani, Janan
    ZilinSkiene, Laura
    Colley, Steve
    Ahmed, Shahzada
    AMERICAN JOURNAL OF OTOLARYNGOLOGY, 2016, 37 (01) : 59 - 64
  • [13] Influence of anatomical location on CT numbers in cone beam computed tomography
    Oliveira, Matheus L.
    Tosoni, Guilherme M.
    Lindsey, David H.
    Mendoza, Kristopher
    Tetradis, Sotirios
    Mallya, Sanjay M.
    ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY, 2013, 115 (04): : 558 - 564
  • [14] Visualization of anatomical structures in the carpal region of the horse using cone beam computed tomography in comparison with conventional multidetector computed tomography
    Hagenbach, M.
    Bierau, J.
    Cruz, A. M.
    Koch, C.
    Manso-Diaz, G.
    Buettner, K.
    Staszyk, C.
    Roecken, M.
    FRONTIERS IN VETERINARY SCIENCE, 2024, 11
  • [15] Visualization of anatomical structures in the fetlock region of the horse using cone beam computed tomography in comparison with conventional multidetector computed tomography
    Bierau, Jonathan
    Cruz, Antonio M.
    Koch, Christoph
    Manso-Diaz, Gabriel
    Buettner, Kathrin
    Staszyk, Carsten
    Roecken, Michael
    FRONTIERS IN VETERINARY SCIENCE, 2024, 10
  • [16] Accuracy of three-dimensional measurements using cone-beam CT
    Pinsky, H. M.
    Dyda, S.
    Pinsky, R. W.
    Misch, K. A.
    Sarment, D. P.
    DENTOMAXILLOFACIAL RADIOLOGY, 2006, 35 (06) : 410 - 416
  • [17] Comparing Effective Doses for Hip Imaging Using Cone Beam CT (CBCT) and Multidetector CT (MDCT)
    Dave, J.
    Jiang, X.
    Machado, P.
    Roth, C.
    Minch, T.
    MEDICAL PHYSICS, 2022, 49 (06) : E629 - E629
  • [18] Picture quality and Hounsfield values of cone beam CT's in comparison to diag. CT
    Weissbaum, S.
    Manser, P.
    Fix, M.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2009, 185 (09) : 651 - 651
  • [19] Quantitative Assessment of Lipiodol Deposition after Chemoembolization: Comparison between Cone-Beam CT and Multidetector CT
    Chen, Rongxin
    Geschwind, Jean-Francois
    Wang, Zhijun
    Tacher, Vania
    Lin, MingDe
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2013, 24 (12) : 1837 - 1844
  • [20] Computed tomography of the abdomen with multidetector-array CT
    Schoepf, UJ
    Becker, C
    Brüning, R
    Hong, C
    Rust, GF
    Helmberger, T
    Leimeister, P
    Stadie, A
    Niethammer, M
    Kingemann, B
    Reiser, MF
    RADIOLOGE, 1999, 39 (08): : 652 - 661