Determination of gallium and indium by solution cathode glow discharge as an excitation source for atomic emission spectrometry

被引:14
|
作者
Yu, Jie [1 ]
Kang, Yuejing [1 ]
Lu, Quanfang [1 ,2 ]
Luo, Hui [1 ]
Cui, Lijuan [1 ]
Lu, Zhuanhong [1 ]
Li, Juanlong [1 ]
机构
[1] Northwest Normal Univ, Coll Chem & Chem Engn, Key Lab Bioelectrochem & Environm Anal Gansu Prov, Lanzhou 730070, Peoples R China
[2] Northwest Normal Univ, Editorial Dept Univ Journal, Lanzhou 730070, Peoples R China
基金
中国国家自然科学基金;
关键词
Atomic emission spectrometry (AES); Determination; Gallium; Indium; Solution cathode glow discharge (SCGD); DIELECTRIC-BARRIER DISCHARGE; FLOWING ELECTROLYTE-SOLUTIONS; ATMOSPHERIC-PRESSURE; LIQUID CATHODE; WATER SAMPLES; ANALYTICAL PERFORMANCE; RAPID-DETERMINATION; ELEMENTAL ANALYSIS; HEAVY-METALS; ANODE;
D O I
10.1016/j.sab.2020.105968
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
In this paper, gallium (Ga) and indium (In) in water samples were determined by atomic emission spectrometry (AES) with solution cathode glow discharge (SCGD) as an excitation source. The optimal working conditions for the detection of Ga and In by proposed method are 660 V discharge voltage, 3.0 mL min(-1) solution flow rate and pH = 1.0 HNO3 as supporting electrolyte. Under these conditions, it is found that there are no obvious inter-ferences from most of the foreign ions when they exist in 20-fold excess over the In. But for Ga, the interferences from foreign ions are severe. The linear correlation coefficient (R-2) is better than 0.99 with a range of 1-40 mu g mL(-1) for Ga and 0.1-30 mu g mL(-1) for In. The limits of detection (LODs) for Ga and In are 0.10 and 0.016 mu g mL(-1), respectively. The relative standard deviations (RSDs) are 2.9% for Ga and 3.9% for In, and the power is below 55 W. To validate the utility and accuracy of the proposed system, two water samples spiked with various concentration (0.3-20 mu g mL(-1)) of Ga and In are directly determined, and the measured results are only in good agreement with spiked values of In. After diluted 10 times for water samples, the spiked values (0.3-20 mu g mL(-1)) of Ga agree well with the measured value, suggesting that the matrix can interfere with the determination of Ga. The satisfactory recoveries between 90% and 109% of In and Ga are achieved.
引用
收藏
页数:8
相关论文
共 50 条
  • [11] Determination of Strontium in Strontium-Rich Mineral Water Using Solution Cathode Glow Discharge-Atomic Emission Spectrometry
    Zheng Pei-chao
    Luo Yuan-jiang
    Wang Jin-mei
    Hu Qiang
    Yang Yang
    Mao Xue-feng
    Lai Chun-hong
    Feng Chu-hui
    He Yu-tong
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2022, 42 (01) : 272 - 276
  • [12] Determination of Trace Metals in Lutetium-Yttrium Orthosilicate by Solution-Cathode Glow Discharge-Atomic Emission Spectrometry
    Gai Rong-Yin
    Wang Zheng
    He Yan-Feng
    Li Qing
    Zou Hui-Jun
    Zhang Guo-Xia
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2014, 42 (11) : 1617 - 1622
  • [13] Ultratrace Pb determination in seawater by solution-cathode glow discharge-atomic emission spectrometry coupled with hydride generation
    Cheng, Junqi
    Li, Qing
    Zhao, Mingyue
    Wang, Zheng
    ANALYTICA CHIMICA ACTA, 2019, 1077 : 107 - 115
  • [14] Determination of calcium and zinc in gluconates oral solution and blood samples by liquid cathode glow discharge-atomic emission spectrometry
    Yu, Jie
    Zhang, Xiaomin
    Lu, Quanfang
    Wang, Xing
    Sun, Duixiong
    Wang, Yongqiang
    Yang, Wu
    TALANTA, 2017, 175 : 150 - 157
  • [15] Solution Cathode Glow Discharge Atomic Emission Spectrometry in Hydrogen-Helium Atmosphere for On-line Determination of Chromium in Sewage
    Chen, Xiang-YU
    Cai, Zhao-Qing
    Pan, Yu-Bai
    Wang, Zheng
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2022, 50 (08) : 1252 - 1259
  • [16] Evaluation of solution-cathode glow discharge atomic emission spectrometry for the analysis of nanoparticle containing solutions
    Hazel, Nicholas
    Orejas, Jaime
    Ray, Steven J.
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2021, 176 (176)
  • [17] Elemental Analysis of Mineral Water by Solution-Cathode Glow Discharge-Atomic Emission Spectrometry
    Zheng, Peichao
    Gong, Yaming
    Wang, Jinmei
    Zeng, Xiaobo
    ANALYTICAL LETTERS, 2017, 50 (09) : 1512 - 1520
  • [18] Analytical Performance of Hollow Anode-Solution Cathode Glow Discharge-Atomic Emission Spectrometry
    Zheng, Peichao
    Li, Weiqi
    Wang, Jinmei
    Wang, Ningshen
    Zhong, Chao
    Luo, Yuanjiang
    Wang, Xiaofa
    Mao, Xuefeng
    Lai, Chunhong
    ANALYTICAL LETTERS, 2020, 53 (05) : 693 - 704
  • [19] Determination of silicate by hollow cathode glow discharge-atomic emission spectrometry with hydride generation technique
    Fujiwara, K
    Wagner, EP
    Smith, BW
    Winefordner, JD
    ANALYTICAL LETTERS, 1996, 29 (11) : 1985 - 1992
  • [20] Highly Sensitive Determination of Cadmium by Atomic Emission Spectrometry with Miniaturized Solution Anode Glow Discharge
    Yu, Jie
    Wang, Xuyi
    Cui, Lijuan
    Hao, Xiaoxia
    Lu, Quanfang
    JOURNAL OF ANALYTICAL CHEMISTRY, 2024, 79 (09) : 1229 - 1241