Neutron/Gamma-ray Pulse Shape Discrimination with a CLYC Scintillator

被引:4
|
作者
Nie, Qiyuan [1 ,2 ]
Zhu, Jinfu [2 ]
Ma, Hao [2 ]
Zeng, Zhi [2 ]
Cheng, Jianping [2 ,3 ]
Li, Junli [2 ]
Zhang, Hui [2 ]
机构
[1] Tsinghua Univ, Minist Educ, Key Lab Particle & Radiat Imaging, Beijing 100084, Peoples R China
[2] Tsinghua Univ, Dept Engn Phys, Beijing 100084, Peoples R China
[3] Beijing Normal Univ, Coll Nucl Sci & Technol, Beijing 100875, Peoples R China
关键词
CLYC scintillator; pulse shape discrimination; Fisher linear discriminant; CS2LIYCL6-CE3+; LUMINESCENCE;
D O I
10.1109/nss/mic42101.2019.9059774
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
Cs2LiYCl6:Ce3+ (CLYC) is a promising Li-based inorganic scintillator for its dual-mode detector capability to perform neutron detection and gamma spectroscopy. Pulse Shape Discrimination (PSD) between neutron and gamma-ray signals can achieve superior results owing to the fundamentally unique scintillation mechanisms within CLYC crystal. In this work, an 1x1 inch CLYC scintillator enriched with 95% Li-6 was used for simultaneous neutron and gamma-ray detection. Difference on pulse rise time between neutron and gamma-ray was found and its effects on traditional charge integration method were analyzed. A pulse shape discrimination (PSD) method based on Fisher linear discriminant was used to discriminate neutron and gamma-ray signals with different parts of waveform. The results show that positive effect on neutron and gamma-ray PSD appears when the integral window is well set based on the pulse rise time. Fisher classification achieves better figure-of-merit (FOM) of 4.6 +/- 0.1 compared to charge integration method. This method also shows reliable screening ability only using the short recording waveforms whose distinctions are mainly from the rising part. When the length of waveforms is 80 ns, the FOM is 2.3 +/- 0.1. This benefits the applications of CLYC in high count rate situations with short recording waveforms.
引用
收藏
页数:6
相关论文
共 50 条
  • [21] Online neutron/gamma discrimination and source detection using CLYC(Ce) scintillator: A sequential approach
    Qi, Sheng
    Li, Jinglun
    Wang, Shanqiang
    Chen, Ye
    Xiao, Wuyun
    Ai, Xianyun
    Zhao, Wei
    Zhang, Kun
    Chen, Wenzhuo
    Li, Jinglun (ljlmxx@126.com), 1600, Elsevier B.V. (1014):
  • [22] Plastic scintillator for pulse shape neutrons and gamma quanta discrimination
    Zhmurin, P. N.
    Lebedev, V. N.
    Adadurov, A. F.
    Pereymak, V. N.
    Gurkalenko, Yu A.
    RADIATION MEASUREMENTS, 2014, 62 : 1 - 5
  • [23] PULSE SHAPE-DISCRIMINATION FOR BACKGROUND REJECTION IN GERMANIUM GAMMA-RAY DETECTORS
    FEFFER, PT
    SMITH, DM
    CAMPBELL, RD
    PRIMBSCH, JH
    LIN, RP
    LUKE, PN
    MADDEN, NW
    PEHL, RH
    MATTESON, JL
    EUV, X-RAY, AND GAMMA-RAY INSTRUMENTATION FOR ASTRONOMY AND ATOMIC PHYSICS, 1989, 1159 : 287 - 294
  • [24] Neutron and Gamma Ray Discrimination for CLYC Using Normalized Cross Correlation Analysis
    Chandhran, Premkumar
    Holbert, Keith E.
    Johnson, Erik B.
    Whitney, Chad
    Vogel, Sam M.
    2014 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2014,
  • [25] Separation of gamma-ray and neutron events with CsI(Tl) pulse shape analysis
    Ashida, Y.
    Nagata, H.
    Koshio, Y.
    Nakaya, T.
    Wendell, R.
    PROGRESS OF THEORETICAL AND EXPERIMENTAL PHYSICS, 2018, 2018 (04):
  • [26] Simultaneous Detection of Energy and Flux for Neutron and Gamma by CLYC Scintillator
    Song, Ruiqiang
    Yan, Xiaoyu
    Han, Jifeng
    Luo, Xiaobing
    Ren, Feixu
    Zhang, Yirong
    Han, Zheng
    Wen, Chun
    Zhang, Xin
    Chen, Lei
    Lin, Weiping
    Qu, Guofeng
    Liu, Xingquan
    Leng, Qiangzhong
    Zhu, Jiajun
    Qian, Sen
    Wang, Zhigang
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2023, 70 (08) : 2148 - 2156
  • [27] Reference-Pulses Neutron/Gamma-Ray Pulse Shape Discrimination in Liquid Scintillators for Deposited Neutron Energies from 200 keV
    Ambers, Scott D.
    Huang, Lu
    Flaska, Marek
    Pozzi, Sara A.
    2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 991 - 994
  • [28] Calibration of EJ-276 plastic scintillator for neutron-gamma pulse shape discrimination experiments
    Ryabeva, E. V.
    Urupa, I. V.
    Lupar, E. E.
    Kadilin, V. V.
    Skotnikova, A. V.
    Kokorev, Y. A.
    Ibragimov, R. F.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2021, 1010
  • [29] X- and gamma-ray detection with a silicon drift detector coupled to a CsI(Tl) scintillator operated with pulse shape discrimination technique
    Marisaldi, M
    Labanti, C
    Soltau, H
    Fiorini, C
    Longoni, A
    Perotti, F
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2005, 52 (05) : 1842 - 1848
  • [30] Pulse Shape Discrimination for CLYC Based Handheld Instruments
    Gueorguiev, A.
    Glodo, J.
    Tower, J.
    Hawrami, R.
    Shirwadkar, U.
    O'Dougherty, P.
    Shah, K. S.
    2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC), 2012, : 32 - 37